Variability in low-flow oxygen delivery by nasal cannula evaluated in neonatal and infant airway replicas

被引:5
|
作者
Sabz, Mozhgan [1 ]
Tavernini, Scott [1 ]
Pillay, Kineshta [1 ]
Christianson, Cole [1 ]
Noga, Michelle [2 ]
Finlay, Warren H. [1 ]
Rouhani, Hossein [1 ]
Martin, Andrew R. [1 ]
机构
[1] Univ Alberta, Dept Mech Engn, Edmonton, AB, Canada
[2] Univ Alberta, Dept Radiol & Diagnost Imaging, Edmonton, AB, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
Low-flow oxygen delivery; Nasal cannula; In vitro testing; Neonates; Infants; Airway replicas; Fraction of inspired oxygen; Oxygen concentration; Oxygen therapy; Supplemental oxygen; OBSTRUCTION; CHILDREN; HEALTHY;
D O I
10.1186/s12931-022-02260-x
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
Background: The nasal cannula is considered a trusted and effective means of administering low-flow oxygen and is widely used for neonates and infants requiring oxygen therapy, despite an understanding that oxygen concentrations delivered to patients are variable. Methods: In the present study, realistic nasal airway replicas derived from medical scans of children less than 3 months old were used to measure the fraction of oxygen inhaled (FiO(2)) through nasal cannulas during low-flow oxygen delivery. Parameters influencing variability in FiO(2) were evaluated, as was the hypothesis that measured FiO(2) values could be predicted using a simple, flow-weighted calculation that assumes ideal mixing of oxygen with entrained room air. Tidal breathing through neonatal and infant nasal airway replicas was controlled using a lung simulator. Parameters for nasal cannula oxygen flow rate, nasal airway geometry, tidal volume, respiratory rate, inhalation/exhalation, or I:E ratio (t(i)/t(e)), breath waveform, and cannula prong insertion position were varied to determine their effect on measured FiO(2). In total, FiO(2) was measured for 384 different parameter combinations, with each combination repeated in triplicate. Analysis of variance (ANOVA) was used to assess the influence of parameters on measured FiO(2). Results: Measured FiO(2) was not appreciably affected by the breath waveform shape, the replica geometry, or the cannula position but was significantly influenced by the tidal volume, the inhalation time, and the nasal cannula flow rate. Conclusions: The flow-weighted calculation overpredicted FiO(2) for measured values above 60%, but an empirical correction to the calculation provided good agreement with measured FiO(2) across the full range of experimental data.
引用
收藏
页数:10
相关论文
共 50 条
  • [21] High flow nasal cannula: Influence of gas type and flow rate on airway pressure and CO2 clearance in adult nasal airway replicas
    Moore, C. P.
    Katz, I. M.
    Pichelin, M.
    Caillibotte, G.
    Finlay, W. H.
    Martin, A. R.
    CLINICAL BIOMECHANICS, 2019, 65 : 73 - 80
  • [22] High flow nasal cannula: influence of gas type and flow rate on positive airway pressure and CO2 clearance in adult nasal airway replicas
    Moore, Charles
    Katz, Ira
    Pichelin, Marine
    Caillibotte, Georges
    Finlay, Warren
    Martin, Andrew
    EUROPEAN RESPIRATORY JOURNAL, 2018, 52
  • [23] Characteristics of Flow in the Upper Airway during High Flow Nasal Cannula Oxygen Therapy
    Kacinski, Robert
    Strasser, Wayne
    Leonard, Scott
    8TH THERMAL AND FLUIDS ENGINEERING CONFERENCE, 2023, : 193 - 201
  • [24] A Randomized Trial of Low-Flow Oxygen versus Nasal Continuous Positive Airway Pressure in Preterm Infants
    Heiring, Christian
    Steensberg, Jesper
    Bjerager, Mia
    Greisen, Gorm
    NEONATOLOGY, 2015, 108 (04) : 259 - 265
  • [25] Performance of Different Low-Flow Oxygen Delivery Systems
    Duprez, Frederic
    Dupriez, Florence
    De Greef, Julien
    Gabriel, Julie
    Bruyneel, Arnaud
    Reychler, Gregory
    De Terwangne, Christophe
    Poncin, William
    RESPIRATORY CARE, 2022, 67 (03) : 322 - 330
  • [26] Performance of different low-flow oxygen delivery systems
    Poncin, William
    Reychler, Gregory
    De Terwangne, Christophe
    Marnette, Matthieu
    Duprez, Frederic
    EUROPEAN RESPIRATORY JOURNAL, 2021, 58
  • [27] Humidity and Inspired Oxygen Concentration During High-Flow Nasal Cannula Therapy in Neonatal and Infant Lung Models
    Chikata, Yusuke
    Ohnishi, Saki
    Nishimura, Masaji
    RESPIRATORY CARE, 2017, 62 (05) : 532 - 537
  • [28] Heated humidified high-flow nasal cannula versus low-flow nasal cannula as weaning mode from nasal CPAP in infants ≤28 weeks of gestation
    Jose Ramon Fernandez-Alvarez
    Rashmi Shreyans Gandhi
    Philip Amess
    Liam Mahoney
    Ryan Watkins
    Heike Rabe
    European Journal of Pediatrics, 2014, 173 : 93 - 98
  • [29] Aerosol Delivery Efficiency With High-Flow Nasal Cannula Therapy in Neonatal, Pediatric, and Adult Nasal Upper-Airway and Lung Models
    DiBlasi, Robert M.
    Engberg, Rebecca J.
    Poli, Jonathan
    Carlin, Kristen E.
    Kontoudios, Niko
    Longest, Worth
    Kajimoto, Masaki
    RESPIRATORY CARE, 2024, 69 (09) : 1146 - 1160
  • [30] Correlation of high flow nasal cannula outlet area with gas clearance and pressure in adult upper airway replicas
    Moore, Charles P.
    Katz, Ira M.
    Caillibotte, Georges
    Finlay, Warren H.
    Martin, Andrew R.
    CLINICAL BIOMECHANICS, 2019, 66 : 66 - 73