THE EFFECT OF ASPHERICITY OF ACETABULAR BEARING SURFACE ON CONTACT MECHANICS OF UHMWPE TOTAL HIP IMPLANTS BY FINITE ELEMENT ANALYSIS

被引:1
|
作者
Zhang, Xuan [1 ]
Wang, Ling [1 ]
Peng, Xifeng [1 ]
Li, Dichen [1 ]
He, Jiankang [1 ]
Liu, Yaxiong [1 ]
Lian, Qin [1 ]
Jin, Zhongmin [1 ,2 ]
Liu, Chaozong [3 ]
机构
[1] Xi An Jiao Tong Univ, State Key Lab Mfg Syst Engn, Xian 710049, Peoples R China
[2] Univ Leeds, Sch Mech Engn, Inst Med & Biol Engn, Leeds LS2 9JT, W Yorkshire, England
[3] UCL, Royal Natl Orthopaed Hosp, John Scales Ctr Biomed Engn, London HA7 4LP, England
基金
中国国家自然科学基金;
关键词
Total hip arthroplasty; asphericity; contact mechanics; finite element analysis; edge loading; MOLECULAR-WEIGHT POLYETHYLENE; METAL-ON-METAL; WEAR DEBRIS; REPLACEMENTS; ARTHROPLASTY; CUPS; PROSTHESES; RATES;
D O I
10.1142/S0219519417500117
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Asphericity and out-of-roundness are generally used to evaluate the manufacturing quality of ultra-high molecular weight polyethylene (UHMWPE) cup inner surfaces, which can potentially affect initial clinical wear and contribute to osteolysis of total hip arthroplasty. This study measured the location and magnitude of asphericity and the out-of-roundness value for four UHMWPE cups in a single set, and then investigated the effects of the asphericity on the contact mechanics of UHMWPE cups. A co-ordinate measuring machine (CMM) was used for the surface measurement and finite element analysis (FEA) was adopted for contact mechanics study. The results demonstrated that the asphericity varied between cups with the maximum value as 0.088 +/- 0.004 mm. Although such a value met the ISO specification, large difference of volume appeared for the asphericity above 0.060 mm. Actual surface profile accounting for the asphericity was found to affect the value of contact pressure and contact area by around 12%. The inferior asphericity resulted in a nonsmoothly distributed contact pressure, which had a negative effect on the contact mechanics of UHMWPE cups and the edge loading was predicted to occur for the sample with a large asphericity. In conclusion, the asphericity of UHMWPE cup could affect the contact mechanics of the articular bearings and may subsequently contribute to initial wear during bedding-in phase.
引用
收藏
页数:14
相关论文
共 50 条
  • [31] Frictional heating of total hip implants. Part 2: finite element study
    Bergmann, G
    Graichen, F
    Rohlmann, A
    Verdonschot, N
    van Lenthe, GH
    JOURNAL OF BIOMECHANICS, 2001, 34 (04) : 429 - 435
  • [32] Finite Element Analysis of Hybird Ceramic Ball Bearing Contact
    Guo, B.
    Han, Y. Q.
    Lei, W. J.
    Wei, H. B.
    Li, F. Z.
    ADVANCED MATERIALS AND COMPUTER SCIENCE, PTS 1-3, 2011, 474-476 : 2064 - +
  • [33] Contact stresses, pressure and area in a fixed-bearing total ankle replacement: a finite element analysis
    Nicolo Martinelli
    Silvia Baretta
    Jenny Pagano
    Alberto Bianchi
    Tomaso Villa
    Gloria Casaroli
    Fabio Galbusera
    BMC Musculoskeletal Disorders, 18
  • [34] Contact stresses, pressure and area in a fixed-bearing total ankle replacement: a finite element analysis
    Martinelli, Nicolo
    Baretta, Silvia
    Pagano, Jenny
    Bianchi, Alberto
    Villa, Tomaso
    Casaroli, Gloria
    Galbusera, Fabio
    BMC MUSCULOSKELETAL DISORDERS, 2017, 18
  • [35] Effect of Surface Roughness on Friction of CoCrMo-on-UHMWPE Bearing in Total Hip Arthroplasty Under Lubrication of Bovine Serum Albumin
    Cong-Truyen Duong
    Cong-Dat Le
    Duc-Nam Nguyen
    7TH INTERNATIONAL CONFERENCE ON THE DEVELOPMENT OF BIOMEDICAL ENGINEERING IN VIETNAM (BME7): TRANSLATIONAL HEALTH SCIENCE AND TECHNOLOGY FOR DEVELOPING COUNTRIES, 2020, 69 : 247 - 251
  • [36] Effect of acetabular cup position on the contact mechanics of ceramic-on-ceramic hip joint replacements
    Mak, MM
    Jin, ZM
    BIOCERAMICS 16, 2004, 254-2 : 639 - 642
  • [37] Acetabular Bone Defect in Total Hip Arthroplasty for Crowe II or III Developmental Dysplasia of the Hip: A Finite Element Study
    Du, Yinqiao
    Fu, Jun
    Sun, Jingyang
    Zhang, Guoqiang
    Chen, Jiying
    Ni, Ming
    Zhou, Yonggang
    BIOMED RESEARCH INTERNATIONAL, 2020, 2020
  • [38] A finite element analysis of factors influencing total hip dislocation
    Scifert, CF
    Brown, TD
    Pedersen, DR
    Callaghan, JJ
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1998, (355) : 152 - 162
  • [39] Selection of Contact Bearing Couple Materials for Hip Prosthesis Using Finite Element Analysis under Static Conditions
    Arirajan, K. A.
    Chockalingam, K.
    Vignesh, C.
    ADVANCES IN MECHANICAL DESIGN, MATERIALS AND MANUFACTURE, 2018, 1943
  • [40] Finite Element Analysis of the Impingement on the Acetabular Liner Rim due to Wear of the Acetabular Liner Surface
    Saputra, Eko
    Anwar, Iwan Budiwan
    Ismail, Rifky
    Jamari, J.
    van der Heide, Emile
    3RD INTERNATIONAL CONFERENCE ON ADVANCED MATERIALS SCIENCE AND TECHNOLOGY (ICAMST 2015), 2016, 1725