The Measurement of the Air-Kerma Rate in Air and a Solid Phantom with Ionization Chambers for a 192Ir HDR Brachytherapy Source

被引:0
|
作者
Zeng, Jing [1 ]
Qu, Pengpeng [1 ]
Pang, Qingsong [2 ]
Wang, Ping [2 ]
机构
[1] Nankai Univ, Tianjin Cent Hosp Gynecol & Obstet, Dept Gynecol Oncol, Affiliated Hosp, Tianjin 300100, Peoples R China
[2] Tianjin Med Univ Canc Inst & Hosp, Natl Clin Res Ctr Canc, Tianjins Clin Res Ctr Canc, Key Lab Canc Prevent & Therapy, Tianjin 300060, Peoples R China
来源
关键词
brachytherapy; Ir-192; source; calibration; ionization chamber; TREATMENT PLANNING SYSTEMS; PROTOCOL;
D O I
10.2147/CMAR.S275378
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Introduction: This study aims to measure the air-kerma rate of 192-Ir-HDR-afterloading source with an ionization chamber in air and a solid cylindrical phantom separately and to compare the dose calibration by the American Association of Physicists in Medicine (AAPM) Task Group TG-43U1 formalism with the Abacus treatment planning system (TPS). Materials and Methods: The air-kerma rate of Ir-192 source was measured by an ionization chamber in air and a solid cylindrical phantom separately. For the interesting point position P (8cm, 90 degrees), the values of the dose were calculated with the TG-43U1 formula and compared with data from the Abacus TPS with single and multiple dwell positions, respectively. Results: The air-kerma rate percentage deviations between the detector measurements in air and the source certificate were -1.28%, -0.91%, -0.71%, and 0.33% at the distances of 25cm, 50cm, 75cm, and 100cm, respectively. For the measurement in solid cylindrical phantom, the percent deviation from the air-kerma rate certificate was 1.85%. The percentage deviations of the dose calibration between Abacus TPS and TG-43U1 formalism at P (8cm, 90 degrees) were -2.30%, 1.76%, and 2.10% with different distances (between the dwell positions) of 0cm, 0.5cm, and 1cm, respectively. Conclusion: The in-air technique was a new attempt for clinic routine measurement. Further studies are still necessary. As a treatment planning system, the Abacus TPS should apply the AAPM TG-43U1 formulism for the development required in the future.
引用
收藏
页码:10821 / 10828
页数:8
相关论文
共 50 条
  • [21] The Application of Elliptic Cylindrical Phantom in Brachytherapy Dosimetric Study of HDR 192Ir Source
    Ahn, Woo Sang
    Park, Sung Ho
    Jung, Sang Hoon
    Choi, Wonsik
    Do Ahn, Seung
    Shin, Seong Soo
    SNA + MC 2013 - JOINT INTERNATIONAL CONFERENCE ON SUPERCOMPUTING IN NUCLEAR APPLICATIONS + MONTE CARLO, 2014,
  • [22] Air kerma standard for calibration of well-type chambers in Brazil using 192Ir HDR sources and its traceability
    Di Prinzio, Renato
    deAlmeida, Carlos Eduardo
    MEDICAL PHYSICS, 2009, 36 (03) : 953 - 960
  • [23] Comparison of air kerma standards of LNE-LNHB and NPL for 192Ir HDR brachytherapy sources:: EUROMET project no 814
    Douysset, Guilhem
    Sander, Thorsten
    Gouriou, Jean
    Nutbrown, Rebecca
    PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (06): : N85 - N97
  • [24] CALIBRATION OF LOW ACTIVITY IR-192 BRACHYTHERAPY SOURCES IN TERMS OF REFERENCE AIR KERMA RATE WITH LARGE VOLUME SPHERICAL IONIZATION CHAMBERS
    VERHAEGEN, F
    VANDIJK, E
    THIERENS, H
    AALBERS, A
    SEUNTJENS, J
    PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (11): : 2071 - 2082
  • [25] Calibration of 192Ir High Dose Rate Brachytherapy Source Using Water Phantom
    Yu, L.
    Yang, B.
    Liu, X.
    Pang, T.
    Liu, N.
    Li, B.
    Dong, T.
    Wang, B.
    Wang, Z.
    Qiu, J.
    MEDICAL PHYSICS, 2018, 45 (06) : E212 - E212
  • [26] Gafchromic film dosimetry of a new HDR 192Ir brachytherapy source
    Ayoobian, Navid
    Asl, Akbar Sarabi
    Poorbaygi, Hosein
    Javanshir, Mohammad Reza
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (02): : 194 - 205
  • [27] Determination of the dose rate around a HDR 192Ir brachytherapy source with the microDiamond and the microSilicon detector
    Rossi, Giulio
    Failing, Thomas
    Gainey, Mark
    Kollefrath, Michael
    Hensley, Frank
    Zink, Klemens
    Baltas, Dimos
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2023, 33 (04): : 463 - 478
  • [28] Monte Carlo aided room scatter studies in the primary air kerma strength standardization of a remote afterloading 192Ir HDR source
    Selvam, TP
    Rajan, KNG
    Nagarajan, PS
    Sethulakshmi, P
    Bhatt, BC
    PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (09): : 2299 - 2315
  • [29] Direct calibration of a reference standard against the air kerma strength primary standard, at 192Ir HDR energy
    Rajan, KNG
    Selvam, TP
    Bhatt, BC
    Vijayam, M
    Patki, VS
    Vinatha
    Pendse, AM
    Kannan, V
    PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (07): : 1047 - 1058
  • [30] Calibration system correction factor for measuring the reference air kerma rate (RAKR) applied to high dose rate 192Ir sources (HDR)
    da Silva, Douglas Batista
    da Silva, Ademir Xavier
    Alves, Carlos Frederico Estrada
    Magalhaes, Luis Alexandre Goncalves
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2024, 10 (03):