The Measurement of the Air-Kerma Rate in Air and a Solid Phantom with Ionization Chambers for a 192Ir HDR Brachytherapy Source

被引:0
|
作者
Zeng, Jing [1 ]
Qu, Pengpeng [1 ]
Pang, Qingsong [2 ]
Wang, Ping [2 ]
机构
[1] Nankai Univ, Tianjin Cent Hosp Gynecol & Obstet, Dept Gynecol Oncol, Affiliated Hosp, Tianjin 300100, Peoples R China
[2] Tianjin Med Univ Canc Inst & Hosp, Natl Clin Res Ctr Canc, Tianjins Clin Res Ctr Canc, Key Lab Canc Prevent & Therapy, Tianjin 300060, Peoples R China
来源
关键词
brachytherapy; Ir-192; source; calibration; ionization chamber; TREATMENT PLANNING SYSTEMS; PROTOCOL;
D O I
10.2147/CMAR.S275378
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Introduction: This study aims to measure the air-kerma rate of 192-Ir-HDR-afterloading source with an ionization chamber in air and a solid cylindrical phantom separately and to compare the dose calibration by the American Association of Physicists in Medicine (AAPM) Task Group TG-43U1 formalism with the Abacus treatment planning system (TPS). Materials and Methods: The air-kerma rate of Ir-192 source was measured by an ionization chamber in air and a solid cylindrical phantom separately. For the interesting point position P (8cm, 90 degrees), the values of the dose were calculated with the TG-43U1 formula and compared with data from the Abacus TPS with single and multiple dwell positions, respectively. Results: The air-kerma rate percentage deviations between the detector measurements in air and the source certificate were -1.28%, -0.91%, -0.71%, and 0.33% at the distances of 25cm, 50cm, 75cm, and 100cm, respectively. For the measurement in solid cylindrical phantom, the percent deviation from the air-kerma rate certificate was 1.85%. The percentage deviations of the dose calibration between Abacus TPS and TG-43U1 formalism at P (8cm, 90 degrees) were -2.30%, 1.76%, and 2.10% with different distances (between the dwell positions) of 0cm, 0.5cm, and 1cm, respectively. Conclusion: The in-air technique was a new attempt for clinic routine measurement. Further studies are still necessary. As a treatment planning system, the Abacus TPS should apply the AAPM TG-43U1 formulism for the development required in the future.
引用
收藏
页码:10821 / 10828
页数:8
相关论文
共 50 条
  • [31] Design of a phantom for the quality control of high dose rate 192Ir source used in brachytherapy
    Ochoa, Ricardo
    Gomez, Fredy
    Ferreira, Ivaldo H.
    Gutt, Federico
    de Almeida, Carlos E.
    RADIOTHERAPY AND ONCOLOGY, 2007, 82 (02) : 222 - 228
  • [32] Calibration of MOSFET detectors for absolute dosimetry with an 192IR HDR brachytherapy source
    Zilio, V
    Joneja, O.
    Popowski, Y.
    Rosenfeld, A.
    Chawla, R.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2006, 182 : 175 - 175
  • [33] Energy dependence of a radiophotoluminescent glass dosimeter for HDR 192Ir brachytherapy source
    Hashimoto, Shimpei
    Nakajima, Yujiro
    Kadoya, Noriyuki
    Abe, Kota
    Karasawa, Katsuyuki
    MEDICAL PHYSICS, 2019, 46 (02) : 964 - 972
  • [34] Application of a pelvic phantom in brachytherapy dosimetry for high-dose-rate (HDR) 192Ir source based on Monte Carlo simulations
    Ahn, Woo Sang
    Choi, Wonsik
    Shin, Seong Soo
    Jung, Jinhong
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2014, 65 (04) : 557 - 564
  • [35] Application of a pelvic phantom in brachytherapy dosimetry for high-dose-rate (HDR) 192Ir source based on Monte Carlo simulations
    Woo Sang Ahn
    Wonsik Choi
    Seong Soo Shin
    Jinhong Jung
    Journal of the Korean Physical Society, 2014, 65 : 557 - 564
  • [36] Audit on source strength determination for HDR and PDR 192Ir brachytherapy in Sweden
    Tedgren, Asa Carlsson
    Grindborg, Jan-Erik
    RADIOTHERAPY AND ONCOLOGY, 2008, 86 (01) : 126 - 130
  • [37] APMP Key comparison report of reference air kerma rate for HDR 192Ir brachytherapy sources (BIPM KCDB: APMP.RI(I)-K8)
    Ishii, J.
    Kurosawa, T.
    Kato, M.
    Toroi, P.
    Chu, W-H
    Yi, C-Y
    Kim, Y. H.
    Seong, Y. M.
    Ngcezu, S. A.
    Mainegra-Hing, E.
    Downtonf, B.
    Dolah, M. T.
    Budiantari, C. T.
    METROLOGIA, 2021, 58 (1A)
  • [38] Air-kerma strength determination of a 169Yb high dose rate brachytherapy source
    VanDamme, J. J.
    Culberson, W. S.
    DeWerd, L. A.
    Micka, J. A.
    MEDICAL PHYSICS, 2008, 35 (09) : 3935 - 3942
  • [39] Air-kerma strength determination of a 169Ytterbium high dose rate brachytherapy source
    VanDamme, J.
    DeWerd, L.
    Micka, J.
    Culberson, W.
    Davis, S.
    MEDICAL PHYSICS, 2006, 33 (06) : 1984 - 1984
  • [40] Second EUROMET comparison of air kerma rate and activity measurements of 192 Ir brachytherapy wires
    Reher, D.F.G., 1600, Publ by Elsevier Science Publishers B.V., Amsterdam, Netherlands (A339): : 1 - 2