Monte Carlo calculation of scattered radiation from applicators in low energy clinical electron beams

被引:0
|
作者
Jabbari, Nasrollah
Hashemi-Malayeri, Bijan [1 ]
Farajollahi, Ali Reza
Kazemnejad, Anoshiravan
机构
[1] Tarbiat Modares Univ, Dept Med Phys, Tehran, Iran
[2] Tarbiat Modares Univ, Dept Biostat, Tehran, Iran
关键词
Monte Carlo; radiation therapy; medical linear accelerator; electron applicator; scattered radiation; treatment planning;
D O I
暂无
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
In radiotherapy with electron beams, scattered radiation from an electron applicator influences the dose distribution in the patient. The contribution of this radiation to the patient dose is significant, even in modern accelerators. In most of radiotherapy treatment planning systems, this component is not explicitly included. In addition, the scattered radiation produced by applicators varies based on the applicator design as well as the field size and distance from the applicators. The aim of this study was to calculate the amount of scattered dose contribution from applicators. We also tried to provide an extensive set of calculated data that could be used as input or benchmark data for advanced treatment planning systems that use Monte Carlo algorithms for dose distribution calculations. Electron beams produced by a NEPTUN 10PC medical linac were modeled using the BEAMnrc system. Central axis depth dose curves of the electron beams were measured and calculated, with and without the applicators in place, for different field sizes and energies. The scattered radiation from the applicators was determined by subtracting the central axis depth dose curves obtained without the applicators from that with the applicator. The results of this study indicated that the scattered radiation from the electron applicators of the NEPTUN 10PC is significant and cannot be neglected in advanced treatment planning systems. Furthermore, our results showed that the scattered radiation depends on the field size and decreases almost linearly with depth.
引用
收藏
页码:97 / 103
页数:7
相关论文
共 50 条
  • [1] Scattered radiation from applicators in clinical electron beams
    van Battum, LJ
    van der Zee, W
    Huizenga, H
    PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (15): : 2493 - 2507
  • [2] Improved Electron Monte Carlo Dose Calculation for Low Energy Electron Beams Using EMC
    Fix, M. K.
    Frei, D.
    Volken, W.
    Neuenschwander, H.
    Born, E. J.
    Manser, P.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [3] Monte Carlo dose calculation improvements for low energy electron beams using eMC
    Fix, Michael K.
    Frei, Daniel
    Volken, Werner
    Neuenschwander, Hans
    Born, Ernst J.
    Manser, Peter
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (16): : 4577 - 4588
  • [4] Modelling of electron contamination in clinical photon beams for Monte Carlo dose calculation
    Yang, J
    Li, JS
    Qin, L
    Xiong, W
    Ma, CM
    PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (12): : 2657 - 2673
  • [5] Evaluation of an electron Monte Carlo dose calculation algorithm for electron beams
    Hu, Ye Angela
    Song, Haijun
    Chen, Zhe
    Zhou, Sumin
    Yin, Fang-Fang
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2008, 9 (03): : 1 - 15
  • [6] Voxel size effects on Monte Carlo dose calculation for clinical radiotherapy electron beams
    Husson, F
    Guillerminet, C
    RADIOTHERAPY AND ONCOLOGY, 2005, 76 : S153 - S153
  • [7] Monte Carlo simulation and measurement of radiation leakage from applicators used in external electron radiotherapy
    Shimozato, Tomohiro
    Okudaira, Kuniyasu
    Fuse, Hiraku
    Tabushi, Katsuyoshi
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2013, 29 (04): : 388 - 396
  • [8] Monte Carlo commissioning of low energy electron beams using NXEGS software
    Both, JA
    Pawlicki, T
    MEDICAL PHYSICS, 2003, 30 (06) : 1517 - 1517
  • [9] Comparison of MCNP4C and EGSnrc Monte Carlo codes in depth-dose calculation of low energy clinical electron beams
    Jabbari, N.
    Hashemi-Malayeri, B.
    Farajollahi, A. R.
    Kazemnejad, A.
    Shafaei, A.
    Jabbari, S.
    JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2007, 40 (15) : 4519 - 4524
  • [10] MONTE-CARLO CALCULATION OF LOW-ENERGY ELECTRON-EMISSION FROM SURFACES
    ALLEN, TE
    KUNZ, RR
    MAYER, TM
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY B, 1988, 6 (06): : 2057 - 2060