Zirconia Cantilever Fixed Dental Prostheses Supported by One or Two Implants: An In Vitro Study on Mechanical Stability and Technical Outcomes

被引:3
|
作者
Karasan, Duygu [1 ]
Canay, Senay [2 ]
Sailer, Irena [1 ]
Att, Wael [3 ]
机构
[1] Univ Geneva, Univ Clin Dent Med, Div Fixed Prosthodont & Biomat, Michel Servet 1, CH-1211 Geneva, Switzerland
[2] Hacettepe Univ, Dept Prosthodont, Fac Dent, Ankara, Turkey
[3] Tufts Univ, Sch Dent Med, Dept Prosthodont, Boston, MA 02111 USA
关键词
dental implants; fixed dental prosthesis; cantilever; zirconia; highly translucent zirconia; OCCLUSAL CONTACT AREA; MAXIMAL BITE FORCE; COMPLICATION RATES; FRACTURE-TOUGHNESS; ANTERIOR MAXILLA; PARTIAL DENTURES; AESTHETIC ZONE; SINGLE CROWNS; RESTORATIONS; SURVIVAL;
D O I
10.11607/jomi.8953
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: To evaluate the mechanical stability of highly translucent zirconia (Zr) cantilevered fixed dental prostheses (cFDPs) and to investigate the influence of the number of implants (one versus two) supporting cFDPs with different restorative materials on their mechanical stability and load-bearing capacity. Materials and Methods: Thirty-two specimens consisting of implant-supported prostheses embedded in resin blocks were fabricated. Sixteen specimens received one implant (bone-level implant, 4.1-mm diameter, 13-mm length; Straumann) to support two-unit cement-retained cFDPs with one extension unit and the other 16 received two implants (bone-level implant, 4.1-mm diameter, 13-mm length; Straumann) positioned corresponding to the missing maxillary central incisors to support three-unit cement-retained cFDPs with one extension unit. Two different prosthetic materials, chromium-cobalt (Cr-Co; Wirobond C+, Bego) and highly translucent Zr (Lava Plus, 3M ESPE) were selected to fabricate the two- and three-unit cFDPs. Standardized two-and three-unit Cr-Co frameworks (CC-I, n = 8; CC-II, n = 8) and highly translucent Zr frameworks (Zr-I, n = 8; Zr-II, n = 8) with a 6-mm cantilever extension were fabricated using CAD/CAM (EOS M 290). Following thermomechanical fatigue loading, the specimens were tested for fracture resistance under static loading. The influence of restoration material and number of supporting implants on fracture resistance were tested using two-way analysis of variance (ANOVA). The level of statistical significance was set below 5% (alpha<.05). Results: All specimens survived aging. The mean (+/- standard deviation) fracture resistance values were 416.25 (+/- 42.71) N for Zr-I, 548.75 (+/- 75.41) N for Zr-II, 601.0 (+/- 41.51) N for CC-I, and 664.5 (+/- 37.59) N for CC-II. CC and Zr group specimens showed significantly different fracture resistance results (P<.001). The number of implants significantly influenced the fracture resistance of Zr groups (P=.001), whereas the influence was not significant for CC groups (P=.089). Conclusion: Within the limitations of this in vitro study, highly translucent zirconia cFDP frameworks demonstrated the potential to withstand reported physiologic occlusal forces applied in the anterior region. The increase in the number of implants supporting zirconia cFDPs significantly contributed to achieving higher fracture resistance values.
引用
收藏
页码:748 / 755
页数:8
相关论文
共 50 条
  • [1] Stability of Cantilever Fixed Dental Prostheses on Zirconia Implants
    Rohr, Nadja
    Nuesch, Reto
    Greune, Rebecca
    Mainetti, Gino
    Karlin, Sabrina
    Zaugg, Lucia K.
    Zitzmann, Nicola U.
    MATERIALS, 2022, 15 (10)
  • [2] Fracture behaviour of zirconia ceramic cantilever fixed dental prostheses in vitro
    Gabbert, Olaf
    Ohlmann, Brigitte
    Schmitter, Marc
    Gilde, Herbert
    Ruef, Thomas
    Rammelsberg, Peter
    ACTA ODONTOLOGICA SCANDINAVICA, 2008, 66 (04) : 200 - 206
  • [3] Effect of implant type on the stability of cantilever fixed dental prostheses: An in vitro study
    Rohr, Nadja
    Karakas-Stupar, Irina
    Karlin, Sabrina
    Zitzmann, Nicola U.
    Zaugg, Lucia K.
    CLINICAL ORAL IMPLANTS RESEARCH, 2024, 35 (04) : 419 - 426
  • [4] Fracture Resistance of Posterior Tooth-Supported Cantilever Fixed Dental Prostheses of Different Zirconia Generations and Framework Thicknesses: An In Vitro Study
    Klotz, Anna-Luisa
    Halfmann, Janina
    Rues, Stefan
    Boemicke, Wolfgang
    Rammelsberg, Peter
    Zenthoefer, Andreas
    MATERIALS, 2024, 17 (01)
  • [5] Survival and complications of implant-supported cantilever fixed dental prostheses with zirconia and metal frameworks: A retrospective cohort study
    Horsch, Laura
    Kronsteiner, Dorothea
    Rammelsberg, Peter
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2022, 24 (05) : 621 - 629
  • [6] Immediate and delayed loading of fixed dental prostheses supported by single or two splinted implants: A histomorphometric study in dogs
    Cesaretti, G.
    Lang, N. P.
    Vigano, P.
    Bengazi, F.
    Alccayhuaman, K. A. Apaza
    Botticelli, D.
    JOURNAL OF ORAL REHABILITATION, 2018, 45 (04) : 308 - 316
  • [7] Mechanical Stability of Restorations Supported by Titanium Base, Zirconia, and Polyetherketoneketone Abutments on One- and Two-Piece Zirconia Implants
    Cantarella, Josef
    Pitta, Joao
    Mojon, Philippe
    Hicklin, Stefan P.
    Fehmer, Vincent
    Sailer, Irena
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2021, 36 (02) : 313 - 321
  • [8] Clinical and radiographic outcomes of implant-supported zirconia fixed dental prostheses with cantilever extension: A proof-of-principle study with a follow-up of at least 1 year
    Roccuzzo, Andrea
    Morandini, Michele
    Stahli, Alexandra
    Imber, Jean-Claude
    Sculean, Anton
    Salvi, Giovanni E.
    CLINICAL ORAL IMPLANTS RESEARCH, 2023, 34 (10) : 1073 - 1082
  • [9] Mechanical Stability of Zirconia Implant Abutments Supporting Cantilevered Fixed Dental Prostheses After Fatigue Loading
    Karasan, Duygu
    Att, Wael
    Sailer, Irena
    Canay, Senay
    INTERNATIONAL JOURNAL OF PROSTHODONTICS, 2021, 34 (05) : 615 - 625
  • [10] Failure Load and Fatigue Behavior of Monolithic and Bi-Layer Zirconia Fixed Dental Prostheses Bonded to One-Piece Zirconia Implants
    Spitznagel, Frank A.
    Hoppe, Johanna S.
    Bonfante, Estevam A.
    Campos, Tiago M. B.
    Langner, Robert
    Gierthmuehlen, Petra C.
    MATERIALS, 2022, 15 (23)