Effects of Inflow and Radiofrequency Spoiling on the Arterial Input Function in Dynamic Contrast-Enhanced MRI: A Combined Phantom and Simulation Study

被引:35
|
作者
Garpebring, Anders [1 ]
Wirestam, Ronnie [2 ]
Ostlund, Nils [3 ]
Karlsson, Mikael [1 ]
机构
[1] Umea Univ, Dept Radiat Sci, Div Radiat Phys, SE-90187 Umea, Sweden
[2] Lund Univ, Dept Med Radiat Phys, Lund, Sweden
[3] Umea Univ Hosp, Dept Radiat Sci, Div Biomed Engn & Informat, S-90185 Umea, Sweden
基金
瑞典研究理事会;
关键词
dynamic contrast-enhanced MRI; arterial input function; blood flow effects; RF spoiling; HIGH-TEMPORAL-RESOLUTION; FAST GRADIENT-ECHO; BLOOD-VOLUME; IMPROVED ACCURACY; T-1; MEASUREMENT; FLOWING BLOOD; PERMEABILITY; PERFUSION; T1; QUANTIFICATION;
D O I
10.1002/mrm.22760
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The arterial input function is crucial in pharmacokinetic analysis of dynamic contrast-enhanced MRI data. Among other artifacts in arterial input function quantification, the blood inflow effect and nonideal radiofrequency spoiling can induce large measurement errors with subsequent reduction of accuracy in the pharmacokinetic parameters. These errors were investigated for a 3D spoiled gradient-echo sequence using a pulsatile flow phantom and a total of 144 typical imaging settings. In the presence of large inflow effects, results showed poor average accuracy and large spread between imaging settings, when the standard spoiled gradient-echo signal equation was used in the analysis. For example, one of the investigated inflow conditions resulted in a mean error of about 40% and a spread, given by the coefficient of variation, of 20% for K-trans. Minimizing inflow effects by appropriate slice placement, combined with compensation for nonideal radiofrequency spoiling, significantly improved the results, but they remained poorer than without flow (e. g., 3-4 times larger coefficient of variation for K-trans). It was concluded that the 3D spoiled gradient-echo sequence is not optimal for accurate arterial input function quantification and that correction for nonideal radiofrequency spoiling in combination with inflow minimizing slice placement should be used to reduce the errors. Magn Reson Med 65:1670-1679, 2011. (C)2011 Wiley-Liss, Inc.
引用
收藏
页码:1670 / 1679
页数:10
相关论文
共 50 条
  • [1] Arterial input function estimation compensating for inflow and partial voluming in dynamic contrast-enhanced MRI
    Tseng, Chih-Hsien
    Nagtegaal, Martijn A.
    van Osch, Matthias J. P.
    Jaspers, Jaap
    Romero, Alejandra Mendez
    Wielopolski, Piotr
    Smits, Marion
    Vos, Frans M.
    NMR IN BIOMEDICINE, 2024, 37 (12)
  • [2] Operator dependency of arterial input function in dynamic contrast-enhanced MRI
    Kleppesto, Magne
    Bjornerud, Atle
    Groote, Inge Rasmus
    Kim, Minjae
    Vardal, Jonas
    Larsson, Christopher
    MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2022, 35 (01) : 105 - 112
  • [3] Correction of arterial input function in dynamic contrast-enhanced MRI of the liver
    Wang, Hesheng
    Cao, Yue
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2012, 36 (02) : 411 - 421
  • [4] Operator dependency of arterial input function in dynamic contrast-enhanced MRI
    Magne Kleppestø
    Atle Bjørnerud
    Inge Rasmus Groote
    Minjae Kim
    Jonas Vardal
    Christopher Larsson
    Magnetic Resonance Materials in Physics, Biology and Medicine, 2022, 35 : 105 - 112
  • [5] Correcting saturation effects of the arterial input function in dynamic susceptibility contrast-enhanced MRI -: a Monte Carlo simulation
    Brunecker, Peter
    Villringer, Arno
    Schultze, Joerg
    Nolte, Christian H.
    Jungehuelsing, Gerhard Jan
    Endres, Matthias
    Steinbrink, Jens
    MAGNETIC RESONANCE IMAGING, 2007, 25 (09) : 1300 - 1311
  • [6] Effects of arterial input function selection on kinetic parameters in brain dynamic contrast-enhanced MRI
    Keil, Vera C.
    Maedler, Burkhard
    Gieseke, Juergen
    Fimmers, Rolf
    Hattingen, Elke
    Schild, Hans H.
    Hadizadeh, Dariusch R.
    MAGNETIC RESONANCE IMAGING, 2017, 40 : 83 - 90
  • [7] Effect of parallel radiofrequency transmission on arterial input function selection in dynamic contrast-enhanced 3 Tesla pelvic MRI
    Chafi, Hatim
    Elias, Saba N.
    Nguyen, Huyen T.
    Friel, Harry T.
    Knopp, Michael V.
    Guo, BeiBei
    Heymsfield, Steven B.
    Jia, Guang
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2016, 43 (01) : 229 - 235
  • [8] Arterial input function for quantitative dynamic contrast-enhanced MRI to diagnose prostate cancer
    Ziayee, Farid
    Mueller-Lutz, Anja
    Gross, Janina
    Ullrich, Tim
    Quentin, Michael
    Arsov, Christian
    Antoch, Gerald
    Wittsack, Hans-Joerg
    Schimmoeller, Lars
    DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY, 2022, 28 (02) : 108 - 114
  • [9] Selection of Fitting Model and Arterial Input Function for Repeatability in Dynamic Contrast-Enhanced Prostate MRI
    Peled, Sharon
    Vangel, Mark
    Kikinis, Ron
    Tempany, Clare M.
    Fennessy, Fiona M.
    Fedorov, Andrey
    ACADEMIC RADIOLOGY, 2019, 26 (09) : E241 - E251
  • [10] Blind Multi-Channel Estimation of Arterial Input Function in Dynamic Contrast-Enhanced MRI
    Jirik, R.
    Bartos, M.
    Standara, M.
    Taxt, T.
    ANALYSIS OF BIOMEDICAL SIGNALS AND IMAGES, 2010, : 373 - 377