Validation of the fully automated Pangao PG-800B68 upper-arm device according to the European Society of Hypertension International Protocol revision 2010

被引:0
|
作者
Yao, Rui [1 ]
Du, You-You [1 ]
Li, Ya-Peng [1 ]
Zhang, Yan-Zhou [1 ]
Chen, Qing-Hua [1 ]
Zhao, Luo-Sha [1 ]
Li, Ling [1 ]
机构
[1] Zhengzhou Univ, Affiliated Hosp 1, Dept Cardiol, Zhengzhou 450052, Peoples R China
关键词
blood pressure measurement; European Society of Hypertension; hypertension; validation; BLOOD-PRESSURE;
D O I
10.1097/MBP.0000000000000111
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
ObjectiveThis study aimed to validate the fully automated Pangao PG-800B68 upper-arm device according to the European Society of Hypertension International Protocol (ESH-IP) revision 2010.Materials and methodsSequential measurements of systolic blood pressure (SBP) and diastolic blood pressure (DBP) were obtained in 33 participants using the standard mercury sphygmomanometer and the test device. The ESH-IP revision 2010 was followed precisely and 99 pairings of test device and reference blood pressure measurements were obtained for analysis.ResultsThe number of blood pressure differences within 5, 10, and 15mmHg was 76, 95, and 98, respectively, for SBP and 79, 96, and 99, respectively, for DBP. The device achieved all the required criteria of the ESH-IP revision 2010 with a device-observer difference of -0.034.64mmHg for SBP and -0.64 +/- 4.50mmHg for DBP.ConclusionAccording to the validation results on the basis of the ESH-IP revision 2010, the Pangao PG-800B68 can be recommended for clinical use and self-measurement in adults.
引用
收藏
页码:221 / 224
页数:4
相关论文
共 50 条
  • [41] European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults
    O'Brien, Eoin
    Atkins, Neil
    Stergiou, George
    Karpettas, Nikos
    Parati, Gianfranco
    Asmar, Roland
    Imai, Yutaka
    Wang, Jiguang
    Mengden, Thomas
    Shennan, Andrew
    BLOOD PRESSURE MONITORING, 2010, 15 (01) : 23 - 38
  • [42] Validation of the Omron HEM-7201 upper arm blood pressure monitor, for self-measurement in a high-altitude environment, according to the European Society of Hypertension International Protocol revision 2010
    K Cho
    M Tian
    Y Lan
    X Zhao
    L L Yan
    Journal of Human Hypertension, 2013, 27 : 487 - 491
  • [43] Validation of the PHILIPS DL8760 upper arm blood pressure monitor, in oscillometry mode, for self-measurement in a general population, according to the European Society of Hypertension International Protocol revision 2010
    Chahine, Mirna N.
    Bou Harb, Souad
    Saad, Abdel Rahman
    Sarkis, Patrick
    Azaki, Alaa
    Harb, Aya
    Allouch, Ali
    Asmar, Roland
    BLOOD PRESSURE MONITORING, 2021, 26 (03) : 237 - 241
  • [44] Validation of the Omron HEM-7201 upper arm blood pressure monitor, for self-measurement in a high-altitude environment, according to the European Society of Hypertension International Protocol revision 2010
    Cho, K.
    Tian, M.
    Lan, Y.
    Zhao, X.
    Yan, L. L.
    JOURNAL OF HUMAN HYPERTENSION, 2013, 27 (08) : 487 - 491
  • [45] Validation of the A&D BP UB-542 wrist device for home blood pressure measurement according to the European Society of Hypertension International Protocol revision 2010
    Saladini, Francesca
    Benetti, Elisabetta
    Fania, Claudio
    Palatini, Paolo
    BLOOD PRESSURE MONITORING, 2013, 18 (04) : 219 - 222
  • [46] Validation of the A&D BP UB-543 wrist device for home blood pressure measurement according to the European Society of Hypertension International Protocol revision 2010
    Fania, Claudio
    Benetti, Elisabetta
    Palatini, Paolo
    BLOOD PRESSURE MONITORING, 2015, 20 (04) : 237 - 240
  • [47] QardioArm Upper Arm Blood Pressure Monitor Against Omron M3 Upper Arm Blood Pressure Monitor in Patients With Chronic Kidney Disease: A Validation Study According to the European Society of Hypertension International Protocol Revision 2010
    Mazoteras-Pardo, Victoria
    Becerro-De-Bengoa-Vallejo, Ricardo
    Losa-Iglesias, Marta Elena
    Lopez-Lopez, Daniel
    Rodriguez-Sanz, David
    Casado-Hernandez, Israel
    Calvo-Lobo, Cesar
    Palomo-Lopez, Patricia
    JOURNAL OF MEDICAL INTERNET RESEARCH, 2019, 21 (12)
  • [48] Validation of the Rossmax AC1000f upper-arm blood pressure monitor in adults according to the British Hypertension Society Protocol
    Huang, Jian-Feng
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2020, 25 (04) : 231 - 235
  • [49] Lack of validation of the Dixtal (DX 2020) upper arm blood pressure monitor, in oscillometric mode, for clinical use in an intensive care unit, according to the European Society of Hypertension-International Protocol revision 2010
    Gothardo, Ana C. L. O.
    Savioli, Amanda F.
    Santos, Dayanna S.
    Lamas, Jose L. T.
    BLOOD PRESSURE MONITORING, 2013, 18 (04) : 215 - 218
  • [50] Validation of the Beurer BM 44 upper arm blood pressure monitor for home measurement, according to the European Society of Hypertension International Protocol 2002
    Lueders, Stephan
    Krueger, Ralf
    Zemmrich, Claudia
    Forstner, Klaus
    Sturm, Claus-Dieter
    Bramlage, Peter
    BLOOD PRESSURE MONITORING, 2012, 17 (06) : 248 - 252