Validation of the fully automated Pangao PG-800B68 upper-arm device according to the European Society of Hypertension International Protocol revision 2010

被引:0
|
作者
Yao, Rui [1 ]
Du, You-You [1 ]
Li, Ya-Peng [1 ]
Zhang, Yan-Zhou [1 ]
Chen, Qing-Hua [1 ]
Zhao, Luo-Sha [1 ]
Li, Ling [1 ]
机构
[1] Zhengzhou Univ, Affiliated Hosp 1, Dept Cardiol, Zhengzhou 450052, Peoples R China
关键词
blood pressure measurement; European Society of Hypertension; hypertension; validation; BLOOD-PRESSURE;
D O I
10.1097/MBP.0000000000000111
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
ObjectiveThis study aimed to validate the fully automated Pangao PG-800B68 upper-arm device according to the European Society of Hypertension International Protocol (ESH-IP) revision 2010.Materials and methodsSequential measurements of systolic blood pressure (SBP) and diastolic blood pressure (DBP) were obtained in 33 participants using the standard mercury sphygmomanometer and the test device. The ESH-IP revision 2010 was followed precisely and 99 pairings of test device and reference blood pressure measurements were obtained for analysis.ResultsThe number of blood pressure differences within 5, 10, and 15mmHg was 76, 95, and 98, respectively, for SBP and 79, 96, and 99, respectively, for DBP. The device achieved all the required criteria of the ESH-IP revision 2010 with a device-observer difference of -0.034.64mmHg for SBP and -0.64 +/- 4.50mmHg for DBP.ConclusionAccording to the validation results on the basis of the ESH-IP revision 2010, the Pangao PG-800B68 can be recommended for clinical use and self-measurement in adults.
引用
收藏
页码:221 / 224
页数:4
相关论文
共 50 条
  • [31] Validation of the A&D UM-211 device for office blood pressure measurement according to the European Society of Hypertension International Protocol revision 2010
    Fania, Claudio
    Albertini, Federica
    Palatini, Paolo
    BLOOD PRESSURE MONITORING, 2017, 22 (05) : 302 - 305
  • [32] Validation of the TONOPORT VI ambulatory blood pressure monitor in adults according to the European Society of Hypertension International Protocol revision 2010
    M Abou-Dakn
    C Döhmen
    S Wenzel
    Journal of Human Hypertension, 2017, 31 : 89 - 92
  • [33] Erratum to: Validation of the custo screen pediatric blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010
    Beate Beime
    Cornelia Deutsch
    Ralf Krüger
    Andreas Wolf
    Peter Müller
    Gertrud Hammel
    Peter Bramlage
    European Journal of Pediatrics, 2017, 176 : 835 - 836
  • [34] Validation of the TONOPORT VI ambulatory blood pressure monitor in adults according to the European Society of Hypertension International Protocol revision 2010
    Abou-Dakn, M.
    Doehmen, C.
    Wenzel, S.
    JOURNAL OF HUMAN HYPERTENSION, 2017, 31 (02) : 89 - 92
  • [35] Validation of the Globalcare GCE603 automated blood pressure monitor for self-measurement according to the European Society of Hypertension International Protocol revision 2010
    Song, Cheng
    Yu, Yang
    Lu, Bao-Chuan
    Yan, Xi-Ling
    BLOOD PRESSURE MONITORING, 2020, 25 (05) : 291 - 294
  • [36] Validation of the Andon KD-5917 automatic upper arm blood pressure monitor, for clinic use and self-measurement, according to the European Society of Hypertension International Protocol revision 2010
    Guo, Wan-Gang
    Li, Bing-Ling
    He, Yong
    Xue, Yu-Sheng
    Wang, Hai-Yan
    Zheng, Qiang-Sun
    Xiang, Ding-Cheng
    BLOOD PRESSURE MONITORING, 2014, 19 (04) : 242 - 245
  • [37] Validation of the A&D BP UA-651 device for home blood pressure measurement according to the European Society of Hypertension International Protocol revision 2010
    Benetti, Elisabetta
    Fania, Claudio
    Palatini, Paolo
    BLOOD PRESSURE MONITORING, 2014, 19 (01) : 50 - 53
  • [38] Validation of the custo screen pediatric blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010 (vol 176, pg 573, 2017)
    Beime, Beate
    Deutsch, Cornelia
    Kruger, Ralf
    Wolf, Andreas
    Mueller, Peter
    Hammel, Gertrud
    Bramlage, Peter
    EUROPEAN JOURNAL OF PEDIATRICS, 2017, 176 (06) : 835 - 836
  • [39] Validation of the Somnotouch-NIBP noninvasive continuous blood pressure monitor according to the European Society of Hypertension International Protocol revision 2010
    Bilo, Grzegorz
    Zorzi, Cristina
    Munera, Juan E. Ochoa
    Torlasco, Camilla
    Giuli, Valentina
    Parati, Gianfranco
    BLOOD PRESSURE MONITORING, 2015, 20 (05) : 291 - 294
  • [40] European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults (vol 15, pg 23, 2010)
    O'Brien, E.
    Atkins, N.
    Stergiou, G.
    Karpettas, N.
    Parati, G.
    Asmar, R.
    BLOOD PRESSURE MONITORING, 2010, 15 (03) : 171 - 172