Speech recognition in noise as a function of the number of spectral channels: Comparison of acoustic hearing and cochlear implants

被引:789
|
作者
Friesen, LM [1 ]
Shannon, RV [1 ]
Baskent, D [1 ]
Wang, X [1 ]
机构
[1] House Ear Res Inst, Dept Auditory Implants & Percept, Los Angeles, CA 90057 USA
来源
关键词
D O I
10.1121/1.1381538
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Speech recognition was measured as a function of spectral resolution (number of spectral channels) and speech-to-noise ratio in normal-hearing (NH) and cochlear-implant (CI) listeners. Vowel, consonant, word, and sentence recognition were measured in five normal -hearing listeners, ten listeners with the Nucleus-22 cochlear implant, and nine listeners with the Advanced Bionics Clarion cochlear implant. Recognition was measured as a function of the number of spectral channels (noise bands or electrodes) at signal-to-noise ratios of + 15, + 10, +5, 0 dB, and in quiet. Performance with three different speech processing strategies (SPEAK, CIS, and SAS) was similar across all conditions, and improved as the number of electrodes increased (up to seven or eight) for all conditions. For all noise levels, vowel and consonant recognition with the SPEAK speech processor did not improve with more than seven electrodes, while for normal-hearing listeners, performance continued to increase up to at least 20 channels. Speech recognition on more difficult speech materials (word and sentence recognition) showed a marginally significant increase in Nucleus-22 listeners from seven to ten electrodes. The average implant score on all processing strategies was poorer than scores of NH listeners with similar processing. However, the best Cl scores were similar to the normal-hearing scores for that condition (up to seven channels). CI listeners with the highest performance level increased in performance as the number of electrodes increased up to seven, while CI listeners with low levels of speech recognition did not increase in performance as the number of electrodes was increased beyond four. These results quantity the effect of number of spectral channels on speech recognition in noise and demonstrate that most Cl subjects are not able to fully utilize the spectral information provided by the number of electrodes used in their implant. (C) 2001 Acoustical Society, of America.
引用
收藏
页码:1150 / 1163
页数:14
相关论文
共 50 条
  • [41] Effects of stimulation rate on speech recognition with cochlear implants
    Friesen, LM
    Shannon, RV
    Cruz, RJ
    AUDIOLOGY AND NEURO-OTOLOGY, 2005, 10 (03) : 169 - 184
  • [42] Speech recognition performance of older children with cochlear implants
    Osberger, MJ
    Fisher, L
    Zimmerman-Phillips, S
    Geier, L
    Barker, MJ
    AMERICAN JOURNAL OF OTOLOGY, 1998, 19 (02): : 152 - 157
  • [43] Neurophysiological parameters for speech recognition in patients with cochlear implants
    Ernst, Robert
    Linxweiler, Maximilian
    Rink, Katharina Anna
    Rothe, Heike
    Lecomte, Gregory
    Bozzato, Alessandro
    Hecker, Dietmar
    LARYNGO-RHINO-OTOLOGIE, 2022, 101 (03) : 216 - 227
  • [44] A meta-analysis to compare speech recognition in noise with bilateral cochlear implants and bimodal stimulation
    Schafer, Erin C.
    Amlani, Amyn M.
    Paiva, Daniele
    Nozari, Ladan
    Verret, Sybil
    INTERNATIONAL JOURNAL OF AUDIOLOGY, 2011, 50 (12) : 871 - 880
  • [45] Cochlear function and speech recognition in the elderly
    Quaranta, A
    Sallustio, V
    Scaringi, A
    AUDIOLOGY, 2001, 40 (06): : 301 - 307
  • [46] Evaluating Speech Intelligibility for Cochlear Implants Using Automatic Speech Recognition
    Zhou, Hengzhi
    Shi, Mingyue
    Meng, Qinglin
    2024 IEEE 14TH INTERNATIONAL SYMPOSIUM ON CHINESE SPOKEN LANGUAGE PROCESSING, ISCSLP 2024, 2024, : 1 - 5
  • [47] Training of Speech Perception in Noise in Pre-Lingual Hearing Impaired Adults With Cochlear Implants Compared With Normal Hearing Adults
    Bugannim, Yossi
    Roth, Daphne Ari-Even
    Zechoval, Doreen
    Kishon-Rabin, Liat
    OTOLOGY & NEUROTOLOGY, 2019, 40 (03) : E316 - E325
  • [48] Speech recognition by normal-hearing and cochlear implant listeners as a function of intensity resolution
    Loizou, PC
    Dorman, M
    Poroy, O
    Spahr, T
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2000, 108 (05): : 2377 - 2387
  • [49] The number of spectral channels required for speech recognition depends on the difficulty of the listening situation
    Shannon, RV
    Fu, QJ
    Galvin, J
    ACTA OTO-LARYNGOLOGICA, 2004, 124 : 50 - 54
  • [50] Comparison of speech recognition in cochlear implant patients with and without residual hearing: A review of indications
    Sargsyan, Gayane
    Kanaan, Natalie
    Lenarz, Thomas
    Lesinski-Schiedat, Anke
    COCHLEAR IMPLANTS INTERNATIONAL, 2021, 22 (05) : 257 - 264