A comparison of transthoracic impedance using standard defibrillation paddles and self-adhesive defibrillation pads

被引:20
|
作者
Deakin, CD [1 ]
McLaren, RM [1 ]
Petley, GW [1 ]
Clewlow, F [1 ]
Dalrymple-Hay, MJR [1 ]
机构
[1] Southampton Univ Hosp, NHS Trust, Southampton Gen Hosp, Dept Anaesthesia, Southampton SO16 6YD, Hants, England
关键词
defibrillation; transthoracic impedance; defibrillation paddles;
D O I
10.1016/S0300-9572(98)00117-8
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
The success of defibrillation is related to transmyocardial current. This current is inversely proportional to transthoracic impedance (TTI). A similar TTI between different pads and paddles is important to deliver a consistent therapeutic dose to all subjects. Failure to do so may result in either insufficient or excessive transmyocardial current leading to a risk of failed defibrillation or tissue damage respectively. Several different types of defibrillation paddles and self-adhesive defibrillation pads are currently available for clinical use but the TTI achieved with each type has not been established. We measured TTI using two types of commonly used paddles and self-adhesive pads to establish whether any significant differences exist between the products. TTI in 40 adult males was measured using defibrillation paddles and self-adhesive defibrillation pads placed in the antero-apical position. Measurements were made using a 30 kHz low amplitude AC current and taken at end-expiration. Mean TTI (Omega) +/- S.D. was 68.2 +/- 16.1 (Hewlett Packard paddles; A), 62.8 +/- 13.2 (Hewlett Packard pads; B), 64.6 +/- 14.3 (PhysioControl paddles, C) and 95.6 +/- 22.3 (PhysioControl pads; D). Significant differences existed between all groups (P < 0.05) except between B and C. Differences in TTI between A, B and C were small and probably of no clinical significance. TTI in group D is significantly larger. Although transmyocardial current is related to TTI, the relationship is complex and differences in TTI alone cannot predict the outcome from defibrillation. (C) 1998 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:43 / 46
页数:4
相关论文
共 50 条
  • [41] IMPEDANCE-BASED TRANSTHORACIC DEFIBRILLATION - A NEW APPROACH TO ENERGY SELECTION
    KERBER, RE
    KIENZLE, MG
    OLSHANSKY, B
    CARLSON, MD
    WALDO, AL
    WILBER, D
    ASCHOFF, AM
    BIRGER, S
    FUGATT, L
    WALSH, S
    ROCKWELL, M
    CHARBONNIER, F
    CIRCULATION, 1990, 82 (04) : 183 - 183
  • [42] PEDIATRIC DEFIBRILLATION - CURRENT FLOW IS IMPROVED BY USING ADULT ELECTRODE PADDLES
    ATKINS, DL
    KERBER, RE
    PEDIATRICS, 1994, 94 (01) : 90 - 93
  • [43] CONTACT URTICARIA AND DERMATITIS FROM SELF-ADHESIVE PADS
    STERRY, W
    SCHMOLL, M
    CONTACT DERMATITIS, 1985, 13 (04) : 284 - 285
  • [44] Problems with the use of self-adhesive electrode pads in neonates
    Cornwell, L
    Mukherjee, R
    Kelsall, AWR
    RESUSCITATION, 2006, 68 (03) : 425 - 428
  • [45] A MULTIPURPOSE, SELF-ADHESIVE PATCH ELECTRODE CAPABLE OF EXTERNAL PACING, CARDIOVERSION DEFIBRILLATION, AND 12-LEAD ELECTROCARDIOGRAM
    WINDLE, JR
    EASLEY, AR
    STRATBUCKER, RA
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (01): : 235 - 241
  • [46] Impedance-gradient electrode reduces skin irritation induced by transthoracic defibrillation
    Meyer, PF
    Gadsby, PD
    Van Sickle, D
    Schoenlein, WE
    Foster, KS
    Graber, GP
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2005, 43 (02) : 225 - 229
  • [47] Effects of positive end-expiratory pressure on transthoracic impedance - implications for defibrillation
    Deakin, CD
    McLaren, RM
    Petley, GW
    Clewlow, F
    Dalrymple-Hay, MJR
    RESUSCITATION, 1998, 37 (01) : 9 - 12
  • [48] High perimeter impedance defibrillation electrodes reduce skin burns in transthoracic cardioversion
    Garcia, LA
    Pagan-Carlo, LA
    Stone, MS
    Kerber, RE
    AMERICAN JOURNAL OF CARDIOLOGY, 1998, 82 (09): : 1125 - +
  • [49] Impedance-gradient electrode reduces skin irritation induced by transthoracic defibrillation
    P. F. Meyer
    P. D. Gadsby
    D. Van Sickle
    W. E. Schoenlein
    K. S. Foster
    G. P. Graber
    Medical and Biological Engineering and Computing, 2005, 43 : 225 - 229
  • [50] Effects of transthoracic impedance and peak current flow on defibrillation success in a prehospital setting
    Heavens, JP
    Cleland, MJ
    Maloney, JP
    Rowe, BH
    ANNALS OF EMERGENCY MEDICINE, 1998, 32 (02) : 191 - 199