Analytical expressions for water-to-air stopping-power ratios relevant for accurate dosimetry in particle therapy

被引:22
|
作者
Luhr, Armin [1 ,2 ]
Hansen, David C. [2 ]
Jaekel, Oliver [3 ,4 ]
Sobolevsky, Nikolai [5 ]
Bassler, Niels [1 ,2 ]
机构
[1] Aarhus Univ Hosp, Dept Expt Clin Oncol, DK-8000 Aarhus, Denmark
[2] Univ Aarhus, Dept Phys & Astron, Aarhus, Denmark
[3] German Canc Res Ctr, Dept Med Phys Radiat Oncol, Heidelberg, Germany
[4] Univ Heidelberg Hosp, Heidelberg Ion Beam Therapy Ctr HIT, Heidelberg, Germany
[5] Russian Acad Sci, Inst Nucl Res, Dept Neutron Res, Moscow 117312, Russia
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2011年 / 56卷 / 08期
关键词
HEAVY-ION RADIOTHERAPY; MONTE-CARLO; RADIATION; TRANSPORT; SOLIDS; SHIELD; CODE;
D O I
10.1088/0031-9155/56/8/012
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In particle therapy, knowledge of the stopping-power ratio (STPR) of the ion beam for water and air is necessary for accurate ionization chamber dosimetry. Earlier work has investigated the STPR for pristine carbon ion beams, but here we expand the calculations to a range of ions (1 <= z <= 18) as well as spread-out Bragg peaks (SOBPs) and provide a theoretical in-depth study with a special focus on the parameter regime relevant for particle therapy. The Monte Carlo transport code SHIELD-HIT is used to calculate complete particle-fluence spectra which are required for determining the STPR according to the recommendations of the International Atomic Energy Agency. The STPR at a depth d depends primarily on the average energy of the primary ions at d rather than on their charge z or absolute position in the medium. However, STPRs for different sets of stopping-power data for water and air recommended by the International Commission on Radiation Units and Measurements are compared, including also the recently revised data for water, yielding deviations up to 2% in the plateau region. In comparison, the influence of the secondary particle spectra on the STPR is about two orders of magnitude smaller in the whole region up till the practical range. The gained insights enable us to propose simple analytical expressions for the STPR for both pristine and SOBPs as a function of penetration depth depending parametrically on the practical range.
引用
收藏
页码:2515 / 2533
页数:19
相关论文
共 35 条
  • [1] Stopping power for particle therapy: The generic library libdEdx and clinically relevant stopping-power ratios for light ions
    Luhr, Armin
    Toftegaard, Jakob
    Kantemiris, Ioannis
    Hansen, David C.
    Bassler, Niels
    INTERNATIONAL JOURNAL OF RADIATION BIOLOGY, 2012, 88 (1-2) : 209 - 212
  • [2] Monte Carlo calculated stopping-power ratios, water/air, for clinical proton dosimetry (50-250 MeV)
    Medin, J
    Andreo, P
    PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (01): : 89 - 105
  • [3] STOPPING-POWER RATIOS AND THEIR UNCERTAINTIES FOR CLINICAL ELECTRON-BEAM DOSIMETRY
    ANDREO, P
    FRANSSON, A
    PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (12): : 1847 - 1861
  • [4] Monte Carlo simulations on the water-to-air stopping power ratio for carbon ion dosimetry
    Henkner, Katrin
    Bassler, Niels
    Sobolevsky, Nikolai
    Jaekel, Oliver
    MEDICAL PHYSICS, 2009, 36 (04) : 1230 - 1235
  • [5] Spencer-Attix water/medium stopping-power ratios for the dosimetry of proton pencil beams
    Goma, C.
    Andreo, P.
    Sempau, J.
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (08): : 2509 - 2522
  • [6] Experimental study of the water-to-air stopping power ratio of monoenergetic carbon ion beams for particle therapy
    Sanchez-Parcerisa, D.
    Gemmel, A.
    Jaekel, O.
    Parodi, K.
    Rietzel, E.
    PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (11): : 3629 - 3641
  • [7] Stopping-power ratios for electron beams used in total skin electron therapy
    Ding, George X.
    MEDICAL PHYSICS, 2021, 48 (09) : 5472 - 5478
  • [8] STOPPING-POWER AND MASS ENERGY-ABSORPTION COEFFICIENT RATIOS FOR SOLID WATER
    HO, AK
    PALIWAL, BR
    MEDICAL PHYSICS, 1986, 13 (03) : 403 - 404
  • [9] THE WATER-TO-AIR STOPPiNG POWER RATIO FOR PROTON AND CARBON ION BEAMS
    Henkner, K.
    Sobolevsky, N.
    Jaekel, O.
    RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S71 - S72
  • [10] Investigation of the Water-to-Air Stopping Power Ratio for Carbon Ion Beam Dosimetry Based on Experimental Data and FLUKA Simulation
    Sanchez-Parcerisa, D.
    Gemmel, A.
    Parodi, K.
    Jakel, O.
    Rietzel, E.
    MEDICAL PHYSICS, 2011, 38 (06) : 3768 - +