Spencer-Attix water/medium stopping-power ratios for the dosimetry of proton pencil beams

被引:22
|
作者
Goma, C. [1 ,2 ]
Andreo, P. [3 ]
Sempau, J. [4 ]
机构
[1] Paul Scherrer Inst, Ctr Proton Therapy, Villigen, Switzerland
[2] Swiss Fed Inst Technol Zurich, Dept Phys, Zurich, Switzerland
[3] Stockholm Univ, Karolinska Univ Hosp, S-10691 Stockholm, Sweden
[4] Univ Politecn Cataluna, Inst Tecn Energet, Barcelona, Spain
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2013年 / 58卷 / 08期
关键词
QUALITY CORRECTION FACTORS; MONTE-CARLO; IONIZATION CHAMBERS; ENERGY; PERTURBATION; SYSTEM; PHOTON;
D O I
10.1088/0031-9155/58/8/2509
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper uses Monte Carlo simulations to calculate the Spencer-Attix water/medium stopping-power ratios (s(w, med)) for the dosimetry of scanned proton pencil beams. It includes proton energies from 30 to 350 MeV and typical detection materials such as air (ionization chambers), radiochromic film, gadolinium oxysulfide (scintillating screens), silicon and lithium fluoride. Track-ends and particles heavier than protons were found to have a negligible effect on the water/air stopping-power ratios (s(w, air)), whereas the mean excitation energy values were found to carry the largest source of uncertainty. The initial energy spread of the beam was found to have a minor influence on the s(w, air) values in depth. The water/medium stopping-power ratios as a function of depth in water were found to be quite constant for air and radiochromic film-within 2.5%. Also, the s(w, med) values were found to have no clinically relevant dependence on the radial distance-except for the case of gadolinium oxysulfide and proton radiography beams. In conclusion, the most suitable detection materials for depth-dose measurements in water were found to be air and radiochromic film active layer, although a small correction is still needed to compensate for the different s(w, med) values between the plateau and the Bragg peak region. Also, all the detection materials studied in this work-except for gadolinium oxysulfide-were found to be suitable for lateral dose profiles and field-specific dose distribution measurements in water.
引用
收藏
页码:2509 / 2522
页数:14
相关论文
共 33 条
  • [1] Spencer-Attix stopping power ratios for flattening filter and flattening filter free photon beams
    Delbaere, A.
    Younes, T.
    Simon, L.
    Khamphan, C.
    Vieillevigne, L.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S1307 - S1307
  • [2] Monte Carlo Calculation of the Spencer-attix Graphite-to-Air Stopping Power Ratios for Photon Beams
    Yi, Chul-Young
    Hah, Suck-Ho
    Chun, Kook Jin
    Kim, Hyun-Moon
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2010, 56 (03) : 926 - 929
  • [3] Monte Carlo calculation of Spencer-Attix and Bragg-Gray stopping-power ratios of tissue-to-air for ISO reference beta sources - an EGSnrc study
    Selvam, T. Palani
    Shrivastava, V
    Bakshi, A. K.
    JOURNAL OF INSTRUMENTATION, 2021, 16 (03)
  • [4] Monte Carlo-based Spencer-Attix and Bragg-Gray tissue-to-air stopping power ratios for ISO beta sources
    Selvam, T. Palani
    Vandana, S.
    Bakshi, A. K.
    Babu, D. A. R.
    RADIATION PROTECTION DOSIMETRY, 2016, 168 (02) : 184 - 189
  • [5] Monte Carlo calculated stopping-power ratios, water/air, for clinical proton dosimetry (50-250 MeV)
    Medin, J
    Andreo, P
    PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (01): : 89 - 105
  • [6] STOPPING-POWER RATIOS AND THEIR UNCERTAINTIES FOR CLINICAL ELECTRON-BEAM DOSIMETRY
    ANDREO, P
    FRANSSON, A
    PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (12): : 1847 - 1861
  • [7] Beam characteristics and stopping-power ratios of small radiosurgery photon beams
    Ding, George X.
    Ding, Frances
    PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (17): : 5509 - 5521
  • [8] Prediction of stopping-power ratios in flattening-filter free beams
    Ceberg, Crister
    Johnsson, Stefan
    Lind, Marten
    Knoos, Tommy
    MEDICAL PHYSICS, 2010, 37 (03) : 1164 - 1168
  • [9] Analytical expressions for water-to-air stopping-power ratios relevant for accurate dosimetry in particle therapy
    Luhr, Armin
    Hansen, David C.
    Jaekel, Oliver
    Sobolevsky, Nikolai
    Bassler, Niels
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (08): : 2515 - 2533
  • [10] Monte Carlo based water/medium stopping-power ratios for various ICRP and ICRU tissues
    Fernandez-Varea, Jose M.
    Carrasco, Pablo
    Panettieri, Vanessa
    Brualla, Lorenzo
    PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (21): : 6475 - 6483