摘要
Objectives: Development and validation of a direct enzymatic HbA1c assay that utilizes a single channel on chemistry auto-analyzers without the need to run separate glycated hemoglobin and total hemoglobin assays. Design and methods: An enzyme based single channel assay was developed to measure %HbA1c in human whole blood samples. The performance characteristics of the Diazyme Direct Enzymatic HbA1c Assay were evaluated on the Hitachi 917 auto-analyzer using whole blood samples, appropriate controls and a reference lot of manufactured reagents. Accuracy studies were completed by comparing the Direct Enzymatic Assay to existing HPLC and immunoassay methods. Interference testing was performed to determine the effect of total hemoglobin, glycated serum proteins, chemical substances and hemoglobin variants in patient samples. Results: The Direct Enzymatic HbA1c Assay showed within run precision and total precision results of <= 2% CV for both normal and abnormal level samples. Method comparison studies showed that there was a good correlation between the Direct Enzymatic HbA1c and the HPLC (R-2=0.98) or the immunoassay (R-2=0.97) methods. The assay measured within the range of 4-16% HbA1c and showed excellent performance with variant hemoglobin in samples. Conclusions: Diazyme Direct Enzymatic HbA1c Assay is accurate and precise when compared to currently marketed medical devices. The assay is designed to report %HbA1c values directly without need for a separate measurement of total hemoglobin and is not adversely affected by interferences from common hemoglobin variants in samples. It is a cost effective, user-friendly method and is adaptable to most general chemistry analyzers. (c) 2008 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:576 / 583
页数:8
相关论文
共 50 条
  • [1] Performance characteristics of the enzymatic Abbott Architect HbA1c whole blood assay
    Ajubi, Nasser E.
    Kross-Brion, Jesmine
    Newton, Cora J.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2014, 52 (12) : E283 - E285
  • [2] Comparing a Direct Whole Blood Enzymatic Assay With High-Performance Liquid Chromatography for HbA1c Measurement
    Zechmeister, Bozena
    Fischer, Andreas
    Eidizadeh, Abass
    JOURNAL OF DIABETES SCIENCE AND TECHNOLOGY, 2024, 18 (01): : 244 - 246
  • [3] Blood cell HbA1c measured by enzymatic assay show higher than whole blood HbA1c in patients with iron deficiency anemia
    Koga, Masafumi
    Ishibashi, Midori
    DIABETOLOGY INTERNATIONAL, 2025, 16 (01) : 86 - 91
  • [4] Direct assay for quantitative measurement of percent HbA1c in anticoagulated whole blood on the Abbott AxSYM® analyzer
    Lawlor, M
    Geekie, C
    Detzner, K
    Milne, K
    CLINICA CHIMICA ACTA, 2005, 355 : S345 - S345
  • [5] Adaptation of the Diazyme Direct Enzymatic HbA1c Assay for a microplate reader at room temperature
    Penttila, Ilkka
    Penttila, Karri
    Halonen, Toivo
    Pulkki, Kari
    Torronen, Jukka
    Rauramaa, Rainer
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2011, 49 (07) : 1221 - 1223
  • [6] THE DIAZYME DIRECT ENZYMATIC HBA1C ASSAY MODIFIED FOR A MICROPLATE READER AT ROOM TEMPERATURE
    Penttila, I.
    Penttila, K.
    Halonen, T.
    Pulkki, K.
    Tarranen, J.
    Rauramaa, R.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2011, 49 : S396 - S396
  • [7] Evaluation of two methods to whole blood HbA1c
    Wang, J. Z.
    Jiang, H.
    Tian, Y. P.
    CLINICAL CHEMISTRY, 2009, 55 (06) : A211 - A212
  • [8] Analytical performance evaluation of an assay for the direct measurement of HbA1c in whole blood on RX series analysers of different throughput
    Mcgivern, P.
    Donnelly, E.
    Sheehy, N.
    Rodriguez, M.
    Lenehan, E.
    Fitzgerald, S.
    CLINICA CHIMICA ACTA, 2019, 493 : S290 - S291
  • [9] Luminol chemiluminescence biosensor for glycated hemoglobin (HbA1c) in human blood samples
    Ahn, Kwang-Soo
    Lee, Jung Hoon
    Park, Jong-Myeon
    Choi, Han Nim
    Lee, Won-Yong
    BIOSENSORS & BIOELECTRONICS, 2016, 75 : 82 - 87
  • [10] Measurement of HbA1c from stored whole blood samples in the Atherosclerosis Risk in Communities study
    Selvin, Elizabeth
    Coresh, Josef
    Zhu, Hong
    Folsom, Aaron
    Steffes, Michael W.
    JOURNAL OF DIABETES, 2010, 2 (02) : 118 - 124