Determination of the modulation transfer function of digitally reconstructed radiographs in radiotherapy treatment planning using a point phantom

被引:5
|
作者
Kirwin, SL [1 ]
Langmack, KA [1 ]
机构
[1] City Hosp Nottingham, Nottingham NG5 1PB, England
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2005年 / 50卷 / 20期
关键词
D O I
10.1088/0031-9155/50/20/N01
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Digitally reconstructed radiographs (DRRs) have become an important tool in radiotherapy treatment planning and treatment verification but their use can be limited by poor image quality. A point phantom comprised of a 1 mm diameter tungsten carbide ball bearing in a wax block has been manufactured as a means of determining a quantitative measure of DRR spatial resolution by calculating the modulation transfer function (MTF) of the image. This method is an alternative to the more often used bar pattern method. The effect of varying CT slice thickness upon DRR spatial resolution has been assessed and a power law relationship between the two variables has been established.
引用
收藏
页码:N251 / N255
页数:5
相关论文
共 49 条
  • [41] Dosimetric Comparison of IMRT with 3D-CRT Regarding Their Contribution to the Treatment Plan Optimization Using Rando Phantom with a Realistic Lung Cancer Radiotherapy Treatment Planning
    Benkahila K.
    Kharfi F.
    Boulakhssaim F.
    Khoudri S.
    Iranian Journal of Medical Physics, 2021, 18 (03) : 154 - 163
  • [42] Performance evaluation of Monaco radiotherapy treatment planning system using CIRS Thorax Phantom: Dosimetric assessment of flattened and non-flattened photon beams
    Saini, Amit
    Verma, Teerthraj
    Pandey, V. P.
    Singh, Avtar
    Kumar, Pankaj
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2023, 19 (03) : 793 - 800
  • [43] Evaluation of several automatic PET-segmentation algorithms for radiotherapy treatment planning in H&N using a printed sub-resolution sandwich phantom
    Berthon, B.
    Holmes, R. B.
    Marshall, C.
    Jayaprakasam, V. S.
    Evans, M.
    Spezi, E.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2013, 40 : S236 - S236
  • [44] Optimization of image reconstruction technique for respiratory-gated lung stereotactic body radiotherapy treatment planning using four-dimensional CT: a phantom study
    Yasue, Kenji
    Fuse, Hiraku
    Takaoka, Minori
    Miyakawa, Shin
    Koori, Norikazu
    Takahashi, Masato
    Shinoda, Kazuya
    Ikoma, Hideaki
    Fujisaki, Tatsuya
    Abe, Shinji
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2025, 18 (01) : 27 - 35
  • [45] Impact of treatment planning using a structure block function on the target and organ doses related to patient movement in cervical esophageal cancer: A phantom study
    Shimizu, Hidetoshi
    Sasaki, Koji
    Ito, Makoto
    Aoyama, Takahiro
    Tachibana, Hiroyuki
    Tomita, Natsuo
    Makita, Chiyoko
    Tanaka, Hiroshi
    Koide, Yutaro
    Iwata, Tohru
    Kodaira, Takeshi
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (05): : 75 - 83
  • [46] A simple approach to measure computed tomography (CT) modulation transfer function (MTF) and noise-power spectrum (NPS) using the American College of Radiology (ACR) accreditation phantom
    Friedman, Saul N.
    Fung, George S. K.
    Siewerdsen, Jeffrey H.
    Tsui, Benjamin M. W.
    MEDICAL PHYSICS, 2013, 40 (05)
  • [47] A method for cranial target delineation in radiotherapy treatment planning aided by single-voxel magnetic resonance spectroscopy: evaluation using a custom-designed gel-based phantom and simulations
    Zeinali-Rafsanjani, Banafsheh
    Mosleh-Shirazi, Mohammad Amin
    Faghihi, Reza
    Saeedi-Moghadam, Mahdi
    Lotfi, Mehrzad
    Jalli, Reza
    BRITISH JOURNAL OF RADIOLOGY, 2019, 92 (1104):
  • [48] MODELING OF SALIVARY PRODUCTION RECOVERY AFTER RADIOTHERAPY USING MIXED MODELS: DETERMINATION OF OPTIMAL DOSE CONSTRAINT FOR IMRT PLANNING AND CONSTRUCTION OF CONVENIENT TOOLS TO PREDICT SALIVARY FUNCTION
    Ortholan, Cecile
    Chamorey, Emmanuel
    Benezery, Karen
    Thariat, Juliette
    Dassonville, Olivier
    Poissonnet, Gilles
    Bozec, Alexandre
    Follana, Philippe
    Peyrade, Frederique
    Sudaka, Anne
    Gerard, Jean Pierre
    Bensadoun, Rene Jean
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 73 (01): : 178 - 186
  • [49] Optical performance analysis using a point spread function and modulation transfer function for W-band PMMW imaging system based on quasi optics focused array of dielectric rod waveguide antenna
    Singh M.K.
    Kim W.-G.
    Tiwary U.S.
    Lee S.-J.
    Cho D.-R.
    Kim Y.-H.
    Progress In Electromagnetics Research B, 2010, (23): : 273 - 291