Determination of the modulation transfer function of digitally reconstructed radiographs in radiotherapy treatment planning using a point phantom

被引:5
|
作者
Kirwin, SL [1 ]
Langmack, KA [1 ]
机构
[1] City Hosp Nottingham, Nottingham NG5 1PB, England
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2005年 / 50卷 / 20期
关键词
D O I
10.1088/0031-9155/50/20/N01
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Digitally reconstructed radiographs (DRRs) have become an important tool in radiotherapy treatment planning and treatment verification but their use can be limited by poor image quality. A point phantom comprised of a 1 mm diameter tungsten carbide ball bearing in a wax block has been manufactured as a means of determining a quantitative measure of DRR spatial resolution by calculating the modulation transfer function (MTF) of the image. This method is an alternative to the more often used bar pattern method. The effect of varying CT slice thickness upon DRR spatial resolution has been assessed and a power law relationship between the two variables has been established.
引用
收藏
页码:N251 / N255
页数:5
相关论文
共 49 条
  • [21] The determination of geometrical modulation transfer function of the eye using a ray tracing method
    Xie, J
    Liu, D
    Loshin, DS
    Jiang, BC
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1997, 38 (04) : 4719 - 4719
  • [22] Determination of the modulation transfer function using the edge method: Influence of scattered radiation
    Neitzel, U
    Buhr, E
    Hilgers, G
    Granfors, PR
    MEDICAL PHYSICS, 2004, 31 (12) : 3485 - 3491
  • [23] Comparing the results of 3D treatment planning and practical dosimetry in craniospinal radiotherapy using Rando phantom
    Baghani, H. R.
    Aghamiri, S. M. R.
    Gharaati, H.
    Mahdavi, S. R.
    Daghigh, S. M. Hosseini
    IRANIAN JOURNAL OF RADIATION RESEARCH, 2011, 9 (03): : 151 - 158
  • [24] A fast method for measurement of modulation transfer function (MTF) and detective quantum efficiency (DQE) in presence of phantom scatter in image guided radiotherapy
    Gopal, A.
    MEDICAL PHYSICS, 2007, 34 (06) : 2619 - 2619
  • [25] Using An Open Source Interior Point Optimization Solver for Intensity Modulated Radiotherapy Treatment Planning
    Tiwari, P.
    Chen, Y.
    Xie, Y.
    Deasy, J.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [26] MEASUREMENT OF THE LANDSAT THEMATIC MAPPER MODULATION TRANSFER-FUNCTION USING AN ARRAY OF POINT SOURCES
    RAUCHMILLER, RF
    SCHOWENGERDT, RA
    OPTICAL ENGINEERING, 1988, 27 (04) : 334 - 343
  • [27] Elastic Bar Transfer Function Determination Using Two-Point Strain Measurements
    Kaczmarek, Zdzislaw
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2009, 58 (01) : 147 - 151
  • [28] CHARACTERIZATION OF THE POINT SPREAD FUNCTION AND MODULATION TRANSFER-FUNCTION OF SCATTERED RADIATION USING A DIGITAL IMAGING-SYSTEM
    BOONE, JM
    ARNOLD, BA
    SEIBERT, JA
    MEDICAL PHYSICS, 1986, 13 (02) : 254 - 256
  • [29] Assessment and improvement of the quality of radiotherapy treatment planning CT images using a clinically validated phantom based method and a multicentre intercomparison
    Davis, Anne T.
    Bird, Andrew
    Cowley, Lorraine
    Donnelly, Oliver
    Elhaddad, Mostafa
    Evans, Cheryl
    Kearton, Tracey
    Morrison, Rachel
    Nash, David
    Naylor, Joshua
    Palmer, Joel
    Potterton, Katherine
    Ravindran, Anand M.
    Sandys, Daniel
    Sdrolia, Athina
    de Stefano, Antonio
    Uherek, Maja
    Walker, Zoe
    Palmer, Antony L.
    Nisbet, Andrew
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2025, 131
  • [30] Analysis of the Resolution Power for three Generations of MV-Imaging Systems in Radiotherapy using the Modulation Transfer Function
    Loot, K.
    Waletzko, O.
    Rohn, R.
    Spaan, B.
    Block, A.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2017, 193 : S114 - S114