Amperometric L-lactate biosensor based on screen-printed carbon electrode containing cobalt phthalocyanine, coated with lactate oxidase-mesoporous silica conjugate layer

被引:54
|
作者
Shimomura, Takeshi [1 ]
Sumiya, Touru [1 ]
Ono, Masatoshi [1 ]
Ito, Tetsuji [2 ]
Hanaoka, Taka-aki [2 ]
机构
[1] Funai Elect Adv Appl Technol Res Inst Inc FEAT, Tsukuba, Ibaraki 3050047, Japan
[2] Natl Inst Adv Ind Sci & Technol, Miyagino Ku, Sendai, Miyagi 9838551, Japan
基金
日本科学技术振兴机构;
关键词
Screen-printed carbon electrode; Cobalt phthalocyanine; Lactate oxidase; Mesoporous silica; Nafion; Sampling unit; GLUCOSE BIOSENSOR; ENZYME ELECTRODES; HYDROGEN-PEROXIDE; IMMOBILIZATION; DESIGN; MATRIX; SENSORS; STABILIZATION; BIOCOMPOSITE; SYSTEM;
D O I
10.1016/j.aca.2011.11.053
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A novel amperometric biosensor for the measurement of L-lactate has been developed. The device comprises a screen-printed carbon electrode containing cobalt phthalocyanine (CoPC-SPCE), coated with lactate oxidase CLOD) that is immobilized in mesoporous silica (FSM8.0) using a polymer matrix of denatured polyvinyl alcohol; a Nafion layer on the electrode surface acts as a barrier to interferents. The sampling unit attached to the SPCE requires only a small sample volume of 100 mu L for each measurement. The measurement of I.-lactate is based on the signal produced by hydrogen peroxide, the product of the enzymatic reaction. The behavior of the biosensor, LOD-FSM8.0/Naf/CoPC-SPCE, was examined in terms of pH, applied potential, sensitivity and operational range, selectivity, and storage stability. The sensor showed an optimum response at a pH of 7.4 and an applied potential of +450 mV. The determination range and the response time for L-lactate were 18.3 mu M to 1.5 mM and approximately 90 s, respectively. In addition, the sensor exhibited high selectivity for I.-lactate and was quite stable in storage, showing no noticeable change in its initial response after being stored for over 9 months. These results indicate that our method provides a simple, cost-effective, high-performance biosensor for L-lactate. (C) 2011 Elsevier B.V. All rights reserved.
引用
收藏
页码:114 / 120
页数:7
相关论文
共 50 条
  • [21] Amperometric Biosensor for Choline Based on Gold Screen-Printed Electrode Modified with Electrochemically-Deposited Silica Biocomposite
    Mazurenko, I.
    Tananaiko, O.
    Biloivan, O.
    Zhybak, M.
    Pelyak, I.
    Zaitsev, V.
    Etienne, M.
    Walcarius, A.
    ELECTROANALYSIS, 2015, 27 (07) : 1685 - 1692
  • [22] Amperometric Biosensor Based on a Semipermeable Poly-Meta-Phenylenediamine Membrane and Immobilized Lactate Oxidase for Highly Accurate L-Lactate Determination in Blood Serum
    Berketa, Kseniia
    Buzhak, Anhelina
    Vakhovskyi, Yevhen
    Mruga, Daryna
    Sverstiuk, Andrii
    Soldatkina, Olha
    Lyubovych, Olha
    Marchuk, Olga
    Dzyadevych, Serhii
    Soldatkin, Oleksandr
    ELECTROANALYSIS, 2025, 37 (01)
  • [23] MWCNTs/PEDOT: PSS Composite as Guiding Layer on Screen-Printed Carbon Electrode for Linear Range Lactate Detection
    Ali, Muhsin
    Khalid, Muhammad Asad Ullah
    Kim, Young Su
    Soomro, Afaque Manzoor
    Hussain, Saddam
    Doh, Yang Hoi
    Choi, Kyung Hyung
    Journal of the Electrochemical Society, 2021, 168 (03):
  • [24] L-Lactate Electrochemical Biosensor Based on an Integrated Supramolecular Architecture of Multiwalled Carbon Nanotubes Functionalized with Avidin and a Recombinant Biotinylated Lactate Oxidase
    Tamborelli, Alejandro
    Mujica, Michael Lopez
    Amaranto, Marilla
    Barra, Jose Luis
    Rivas, Gustavo
    Godino, Agustina
    Dalmasso, Pablo
    BIOSENSORS-BASEL, 2024, 14 (04):
  • [25] Electrochemical lactate biosensor based upon chitosan/carbon nanotubes modified screen-printed graphite electrodes for the determination of lactate in embryonic cell cultures
    Hernandez-Ibanez, Naiara
    Garcia-Cruz, Leticia
    Montiel, Vicente
    Foster, Christopher W.
    Banks, Craig E.
    Iniesta, Jesus
    BIOSENSORS & BIOELECTRONICS, 2016, 77 : 1168 - 1174
  • [26] VOLTAMMETRIC AND AMPEROMETRIC STUDIES OF THIOCHOLINE AT A SCREEN-PRINTED CARBON ELECTRODE CHEMICALLY-MODIFIED WITH COBALT PHTHALOCYANINE - STUDIES TOWARDS A PESTICIDE SENSOR
    HART, JP
    HARTLEY, IC
    ANALYST, 1994, 119 (02) : 259 - 263
  • [27] Development of a disposable ethanol biosensor based on a chemically modified screen-printed electrode coated with alcohol oxidase for the analysis of beer
    Boujtita, M
    Hart, JP
    Pittson, R
    BIOSENSORS & BIOELECTRONICS, 2000, 15 (5-6): : 257 - 263
  • [28] A novel, disposable, screen-printed amperometric biosensor for ketone 3-β-hydroxybutyrate fabricated using a 3-β-hydroxybutyrate dehydrogenase–mesoporous silica conjugate
    Takeshi Shimomura
    Touru Sumiya
    Masatoshi Ono
    Tetsuji Ito
    Taka-aki Hanaoka
    Analytical and Bioanalytical Chemistry, 2013, 405 : 297 - 305
  • [29] Voltammetric and amperometric studies of selected thiols and dimethyldisulfide using a screen-printed carbon electrode modified with cobalt phthalocyanine: Studies towards a gas sensor
    Napier, A
    Hart, JP
    ELECTROANALYSIS, 1996, 8 (11) : 1006 - 1013
  • [30] An amperometric glucose biosensor constructed by immobilizing glucose oxidase on titanium-containing mesoporous composite material of no. 41 modified screen-printed electrodes
    Dai, Zhihui
    Fang, Min
    Bao, Jianchun
    Wang, Huaisheng
    Lu, Tianhong
    ANALYTICA CHIMICA ACTA, 2007, 591 (02) : 195 - 199