Carotid intima-media thickness and hemodynamic parameters: reproducibility of manual measurements with Doppler ultrasound

被引:15
|
作者
Plasencia Martinez, Juana M. [1 ]
Garcia Santos, Jose M. [1 ]
Paredes Martinez, Maria L. [1 ]
Moreno Pastor, Ana [1 ]
机构
[1] Univ Gen Hosp Morales Meseguer, Dept Radiol, Murcia, Spain
关键词
ultrasonography; Doppler; carotid intima-media thickness; observer variation; reproducibility; CORONARY-ARTERY-DISEASE; CARDIOVASCULAR-DISEASE; MYOCARDIAL-INFARCTION; DIFFERENT SITES; RISK; PLAQUE; ATHEROSCLEROSIS; EVENTS; VARIABILITY; MEDICINE;
D O I
10.11152/mu.2013.2066.172.ci-m
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Aims: To evaluate the carotid ultrasound intra-and interobserver agreements in a common clinical scenario when making manual measurements of the intima-media thickness (IMT) and peak systolic (PSV) and end diastolic (EDV) velocities in the common (CCA) and the internal carotid (ICA) arteries. Material and methods: Three different experienced operators performed two time-point carotid ultrasounds in 21 patients with cardiovascular risk factors. Each operator measured freehand the CCA IMT three consecutive times in each examination. The CCA and ICA hemodynamic parameters were acquired just once. For our purpose we took the average (IMTmean) and maximum (IMTmax) IMT values. Quantitative variables were analyzed with the t-student, and ANOVA test. Agreements were evaluated with the Intraclass Correlation Coefficient (ICC). Results: IMTmean intraobserver agreement was better on the left (ICC: 0.930-0.851-0.916, operators 1-2-3) than on the right (ICC: 0.789-0.580-0.673, operators 1-2-3). IMTmax agreements (Left ICC: 0.821-0.723-0.853, operators 1-2-3; Right ICC: 0.669-0.421-0.480, operators 1-2-3) were lower and more variable. Interobserver agreements for IMTmean (ICC: 0.852-0.860; first-second ultrasound) and IMTmax (ICC: 0.859-0.835; first-second ultrasound) were excellent on the left, but fair-good and more variable on the right (IMTmean; ICC: 0.680-0.809; first-second ultrasound; IMTmax; 0.694-0.799; first-second ultrasound). Intraobserver agreements were fair-moderate for PSVs and good-excellent for EDVs. Interobserver agreements were good-excellent for both PSVs and EDVs. Overall, 95% confidence intervals were narrower for the left IMTmean and CCA velocities. Conclusions: Intra and interobserver agreements in carotid ultrasound are variable. In order to improve carotid IMT agreements, IMTmean is preferable over IMTmax.
引用
收藏
页码:167 / 174
页数:8
相关论文
共 50 条
  • [1] Reproducibility of Sonographic Measurements of Carotid Intima-Media Thickness
    Velazquez, F.
    Berna, J. D.
    Abellan, J. L.
    Serrano, L.
    Escribano, A.
    Canteras, M.
    ACTA RADIOLOGICA, 2008, 49 (10) : 1162 - 1166
  • [2] Reproducibility of in vivo carotid intima-media thickness measurements - A review
    Kanters, SDJM
    Algra, A
    vanLeeuwen, MS
    Banga, JD
    STROKE, 1997, 28 (03) : 665 - 671
  • [3] Reproducibility of carotid intima-media thickness measurements in young adults
    Gonzalez, Jaime
    Wood, John C.
    Dorey, Frederick J.
    Wren, Tishya A. L.
    Gilsanz, Vicente
    RADIOLOGY, 2008, 247 (02) : 465 - 471
  • [4] Ultrasound semiautomatic versus manual estimation of carotid intima-media thickness: reproducibility and cardiovascular risk stratification
    Monti, Caterina Beatrice
    Ali, Marco
    Capra, Davide
    Wiedenmann, Federico
    Lastella, Giulia
    Secchi, Francesco
    Sardanelli, Francesco
    MEDICAL ULTRASONOGRAPHY, 2020, 22 (04) : 402 - 408
  • [5] Carotid Intima-Media Thickness Measurements Are Not Affected by the Ultrasound Frequency
    Bhagirath, Kapil M.
    Lester, Steven J.
    Humphries, Julie
    Hentz, Joseph G.
    Hurst, R. Todd
    ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 2012, 29 (03): : 354 - 357
  • [6] Reproducibility of measurements of intima-media thickness and distensibility in the common carotid artery
    Kanters, SDJM
    Elgersma, OEH
    Banga, JD
    van Leeuwen, MS
    Algra, A
    EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 1998, 16 (01) : 28 - 35
  • [7] Carotid Artery Intima-Media Thickness Measurements in the Youth: Reproducibility and Technical Considerations
    Tierney, Elif Seda Selamet
    Gauvreau, Kimberlee
    Jaff, Michael R.
    Gal, Dana
    Nourse, Susan E.
    Trevey, Shari
    O'Neill, Stephen
    Baker, Annette
    Newburger, Jane W.
    Colan, Steven D.
    JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2015, 28 (03) : 309 - 316
  • [8] Ultrasound assessment of atherosclerotic vessel wall changes -: Reproducibility of intima-media thickness measurements in carotid and femoral arteries
    Srámek, A
    Bosch, JG
    Reiber, JHC
    Van Oostayen, JA
    Rosendaal, FR
    INVESTIGATIVE RADIOLOGY, 2000, 35 (12) : 699 - 706
  • [9] An update on carotid ultrasound measurement of intima-media thickness
    Barth, JD
    AMERICAN JOURNAL OF CARDIOLOGY, 2002, 89 (4A): : 32B - 38B
  • [10] Ultrasound Evaluation of Carotid Intima-Media Thickness in Children
    Epifanio, Matias
    Baldisserotto, Matteo
    Sarria, Edgar E.
    Lazaretti, Arthur
    Mattiello, Rita