A study of atypical APTT derivative curves on the ACL TOP coagulation analyser

被引:40
|
作者
Solano, C. [1 ]
Zerafa, P. [1 ]
Bird, R. [1 ]
机构
[1] Princess Alexandra Hosp, Dept Haematol, Clin & Statewide Serv Div, Queensland Hlth, Brisbane, Qld 4102, Australia
关键词
Activated partial thromboplastin time; clot reaction curve; second derivative; first derivative; WAVE-FORM ANALYSIS; DISSEMINATED INTRAVASCULAR COAGULATION; HEMOPHILIA-A; DISORDERS; DIAGNOSIS;
D O I
10.1111/j.1751-553X.2010.01248.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
P>Introduction: The graphical representation of clotting data can provide useful information. A novel feature of the ACL TOP software allows display of clot reaction curves with superimposed first and second derivative curves. We noted atypical derivative plots associated with normal 'S'-shaped clot reaction curves in some abnormal activated partial thromboplastin time (APTT). Methods: We tested 179 citrate plasmas with four different APTT reagents on the ACL TOP. We documented the prevalence and utility of these APTT atypical derivative curves in these patients. Results: Ellagic acid APTT reagents commonly produce atypical derivative curves despite normal 'S'-shaped clot reaction curves. Occasionally, these atypical second derivative curves may cause incorrect APTT results. With silica activator APTT reagents, atypical derivative curves were associated with genuine coagulation abnormalities such as single factor deficiencies and lupus anticoagulants (LAs). To our knowledge, this novel finding has not been reported. Conclusion: Atypical derivative curves seen in ellagic acid APTTs are of limited diagnostic use because of the frequency with which they occur. This may be related to the need to optimize the data reduction utilized on the ACL TOP for these reagents. With silica activator APTTs, the presence of atypical derivative curves proved to be a very simple tool when troubleshooting unexpected abnormal APTT results, commonly predicting a factor deficiency or LA that would warrant further investigation. The cause of these aberrant derivative curves is probably related to abnormal thrombin generation in the APTT test and warrants further study.
引用
收藏
页码:67 / 78
页数:12
相关论文
共 43 条
  • [31] Effect of Selected Crystalloid and Colloid Solutions on Coagulation Status Evaluated by Rotational Thromboelastometry and Platelet Function Analyser: An in vitro Study
    Jonas, Jakub
    Durila, Miroslav
    Vymazal, Tomas
    TRANSFUSION MEDICINE AND HEMOTHERAPY, 2024,
  • [32] Modified Anti-FXa Assays for Measuring the Residual Activity of Apixaban and Rivaroxaban in Andexanet Alfa-Containing Samples on the ACL TOP Family® Coagulation Analyzers
    Khan, Anuja
    Shishko, Krisjela
    Arcidiacono, Sam
    Bui, Khanh
    Lu, Genmin
    Conley, Pamela B.
    Winkler, Anne
    BLOOD, 2019, 134
  • [33] ExC810全自动凝血分析仪检测APTT、PT、TT及FIB性能验证及与ACL-TOP700比对分析
    邓婷婷
    庄丽颖
    刘爱胜
    生物医学工程与临床, 2025, 29 (01) : 110 - 114
  • [34] Electron optical study of the Venus Express ASPERA-4 Electron Spectrometer (ELS) top-hat electrostatic analyser
    Collinson, Glyn A.
    Kataria, D. O.
    Coates, Andrew J.
    Tsang, Sharon M. E.
    Arridge, Christopher S.
    Lewis, Gethyn R.
    Frahm, Rudy A.
    Winningham, J. David
    Barabash, Stas
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2009, 20 (05)
  • [35] Study of stability curves of a top tension riser under the effect of response spectrum of platform motion
    Sun L.
    Wang J.
    Dai S.
    Dai, Shaoshi (daishaoshi@163.com), 1600, Editorial Board of Journal of Harbin Engineering (37): : 326 - 331
  • [36] A study of differential polarization curves and thermodynamic properties for mild steel in acidic solution with nitrophenyltriazole derivative
    Tao, Z.
    He, W.
    Wang, S.
    Zhang, S.
    Zhou, G.
    CORROSION SCIENCE, 2012, 60 : 205 - 213
  • [37] USE OF DERIVED DERIVATIVE THERMOGRAVIMETRIC CURVES IN COMPOSITIONAL STUDY OF 2-COMPONENT OXIDE SYSTEMS
    ARORA, BR
    SINGH, BN
    BANERJEE, RK
    JOURNAL OF THERMAL ANALYSIS, 1978, 13 (03): : 499 - 508
  • [38] Theoretical study on modification of top profile based on arc shaped curves with equal beta angle for scroll fluid machinery
    Gao, X.
    Liu, W.
    Feng, S.
    Gu, Z.
    Yu, Y.
    2001, Xi'an Jiaotong University (35):
  • [39] PROTHROMBIN TIME, ACTIVATED PARTIAL THROMBOPLASTIN TIME, FIBRINOGENEMIA - A COMPARATIVE-STUDY BETWEEN 2 AUTOMATES OF COAGULATION - KOAGULAB-40A AND ACL-100
    BARBE, D
    BAUDUER, F
    TANGUY, A
    ANNALES DE BIOLOGIE CLINIQUE, 1991, 49 (10) : 507 - 513
  • [40] In vitro effect of the direct Xa inhibitor, apixaban, on global coagulation laboratory tests and coagulation factor assays: Comparison of 10 aPTT and 10 PT reagents, and two coagulometers. A "Groupe D'Etude sur l'Hemostase et la Thrombose" study
    Isabelle, Gouin-Thibault
    Claire, Flaujac
    Marie-Helene, Horellou
    Nicole, Lacarriere
    Thomas, Lecompte
    THROMBOSIS RESEARCH, 2012, 130 : S108 - S108