Validation of Ankle Strength Measurements by Means of a Hand-Held Dynamometer in Adult Healthy Subjects

被引:8
|
作者
Ancillao, Andrea [1 ]
Palermo, Eduardo [1 ]
Rossi, Stefano [2 ]
机构
[1] Sapienza Univ Rome, Dept Mech & Aerosp Engn, Via Eudossiana 18, I-00184 Rome, Italy
[2] Univ Tuscia, Dept Econ & Management Ind Engn DEIM, Via Paradiso 47, I-01100 Viterbo, Italy
关键词
MUSCLE STRENGTH; GAIT PATTERN; RELIABILITY; KNEE; VALIDITY; QUALITY; CHILDREN; POSTURE; TORQUE; MOTION;
D O I
10.1155/2017/5426031
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Uniaxial Hand-Held Dynamometer (HHD) is a low-cost device widely adopted in clinical practice to measure muscle force. HHD measurements depend on operator's ability and joint movements. The aim of the work is to validate the use of a commercial HHD in both dorsiflexion and plantarflexion ankle strength measurements quantifying the effects of HHD misplacements and unwanted foot's movements on the measurements. We used an optoelectronic system and a multicomponent load cell to quantify the sources of error in the manual assessment of the ankle strength due to both the operator's ability to hold still the HHD and the transversal components of the exerted force that are usually neglected in clinical routine. Results showed that foot's movements and angular misplacements of HHD on sagittal and horizontal planes were relevant sources of inaccuracy on the strength assessment. Moreover, ankle dorsiflexion and plantarflexion force measurements presented an inaccuracy less than 2% and higher than 10%, respectively. In conclusion, the manual use of a uniaxial HHD is not recommended for the assessment of ankle plantarflexion strength; on the contrary, it can be allowed asking the operator to pay strong attention to the HHD positioning in ankle dorsiflexion strength measurements.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Reliability of Isometric Knee Extension Muscle Strength Measurements of Healthy Elderly Subjects Made with a Hand-held Dynamometer and a Belt
    Katoh, Munenori
    Isozaki, Koji
    JOURNAL OF PHYSICAL THERAPY SCIENCE, 2014, 26 (12) : 1855 - 1859
  • [2] Reliability of measurements of hip abduction strength obtained with a hand-held dynamometer
    Ieiri, Akira
    Tushima, Eiki
    Ishida, Kazuhiro
    Inoue, Masahiro
    Kanno, Taiki
    Masuda, Takeshi
    PHYSIOTHERAPY THEORY AND PRACTICE, 2015, 31 (02) : 146 - 152
  • [3] Development and reliability of a hand-held dynamometer device to measure the ankle dorsiflexor muscle strength
    Kimura, Masanobu
    Ogata, Yuto
    Akebi, Toru
    Ochi, Mitsuhiro
    Hachisuka, Kenji
    Saeki, Satoru
    JOURNAL OF REHABILITATION AND ASSISTIVE TECHNOLOGIES ENGINEERING, 2018, 5
  • [4] Reliability and validity of measurements of cervical retraction strength obtained with a hand-held dynamometer
    Tudini, Frank
    Myers, Bradley
    Bohannon, Richard
    JOURNAL OF MANUAL & MANIPULATIVE THERAPY, 2019, 27 (04) : 222 - 228
  • [5] Inter- and Intratester Reliability of Strength Measurement of Knee Extension and Flexion with a Modified Hand-Held Dynamometer in Healthy Subjects
    Schaefer, A.
    Dressel, M.
    Kaemmerer, S.
    Peschke, L.
    Braun, M. Ramos
    Schoettker-Koeniger, T.
    MANUELLE THERAPIE, 2015, 19 (01) : 38 - 44
  • [6] Analysis of Knee Strength Measurements Performed by a Hand-Held Multicomponent Dynamometer and Optoelectronic System
    Ancillao, Andrea
    Rossi, Stefano
    Cappa, Paolo
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2017, 66 (01) : 85 - 92
  • [7] Reliability of hand-held dynamometer in measuring gluteus medius isometric muscle strength in healthy population
    Soomro, Rabail Rani
    Karimi, Hossein
    Gillani, Syed Amir
    PAKISTAN JOURNAL OF MEDICAL SCIENCES, 2022, 38 (05) : 1238 - 1242
  • [8] Reliability of Hand-Held Dynamometer for assessing Isometric Lumbar Muscles Strength in Asymptomatic Healthy Population
    Tanveer, Fahad
    Arslan, Syed Asadullah
    Darain, Haider
    Ahmad, Ashfaq
    PAKISTAN JOURNAL OF MEDICAL SCIENCES, 2021, 37 (02) : 461 - 465
  • [9] RELIABILITY, COMPARABILITY, AND VALIDITY OF FOOT INVERSION AND EVERSION STRENGTH MEASUREMENTS USING A HAND-HELD DYNAMOMETER
    Alfuth, Martin
    Murielle, Magali
    INTERNATIONAL JOURNAL OF SPORTS PHYSICAL THERAPY, 2016, 11 (01): : 72 - 84
  • [10] Validity of Isometric Muscle Strength Measurements of the Lower Limbs Using a Hand-held Dynamometer and Belt: a Comparison with an Isokinetic Dynamometer
    Katoh, Munenori
    Hiiragi, Yukinobu
    Uchida, Manabu
    JOURNAL OF PHYSICAL THERAPY SCIENCE, 2011, 23 (04) : 553 - 557