In vitro study of a modified sagittal split osteotomy fixation technique of the mandible: a mechanical test

被引:3
|
作者
Sonego, C. L. [1 ]
Scheffer, M. A. R. [1 ]
Chagas Junior, O. L. [2 ]
Vetromilla, B. M. [3 ]
Fernandes, L. P. [1 ]
Ozkomur, A. [1 ]
Silva Junior, A. N. [1 ]
Miguens Junior, S. A. Q. [1 ]
Hernandez, P. A. G. [1 ]
机构
[1] Univ Luterana Brasil, Grad Program Dent Oral & Maxillofacial Surg, Canoas, RS, Brazil
[2] Univ Fed Pelotas, Univ Hosp, Oral & Maxillofacial Surg Residency Program, Pelotas, RS, Brazil
[3] Univ Fed Pelotas, Grad Program Dent Prosthodont, Sch Dent, Pelotas, RS, Brazil
关键词
sagittal split osteotomy of the mandibular ramus; fracture fixation; material resistance; RIGID INTERNAL-FIXATION; RAMUS OSTEOTOMY; BIOMECHANICAL EVALUATION; PLATE FIXATION; ORTHOGNATHIC SURGERY; ADVANCEMENT SURGERY; SCREW FIXATION; STABILITY; SETBACK; MINIPLATE;
D O I
10.1016/j.ijom.2018.03.023
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
This study was performed to evaluate the compressive mechanical strength of rigid internal fixation (RIF) using 1.5-mm L-shaped plates fixed with monocortical screws in sagittal split osteotomy (SSO). Thirty synthetic hemimandibles, which had all undergone a 5-mm advancement, were divided into three groups: three 12-mm bicortical titanium screws were placed in an inverted L pattern in group A; one straight 2.0-mm system spaced titanium plate fixed with four 5-mm monocortical screws was used in group B; two 1.5-mm system L-shaped titanium plates, each fixed with four 5-mm monocortical screws, were used in group C. The models were subjected to compressive and progressive mechanical tests with forces applied in the area between the second premolar and first molar to verify resistance in Newtons (N). A displacement speed of 1 mm/min was applied, with a maximum 10 mm displacement of the distal segment or until disruption of the fixation. The deformity and/or eventual rupture of the plates were evaluated, and consequently their technical stability was determined. The results showed that the modified fixation technique tested in this study on synthetic mandibles resulted in adequate stability and superior mechanical behaviour compared to simulated osteosynthesis with the use of a straight 2.0-mm titanium plate.
引用
收藏
页码:1330 / 1335
页数:6
相关论文
共 50 条
  • [21] A modified sagittal split ramus osteotomy
    Park, JU
    3RD ASIAN CONGRESS ON ORAL AND MAXILLOFACIAL SURGERY, 1996, : 303 - 305
  • [22] Modified Mandibular Sagittal Split Osteotomy
    Ferri, Joel
    Schlund, Matthias
    Roland-Billecart, Thomas
    Nicot, Romain
    JOURNAL OF CRANIOFACIAL SURGERY, 2019, 30 (03) : 897 - 899
  • [23] Skeletal Stability after Bilateral Sagittal Split Advancement and Setback Osteotomy of the Mandible with Miniplate Fixation
    Rao, Srinivasan Hanumantha
    Selvaraj, Loganathan
    Lankupalli, Arathy S.
    CRANIOMAXILLOFACIAL TRAUMA & RECONSTRUCTION, 2014, 7 (01) : 9 - 15
  • [24] Improving control of lingual split propagation in sagittal split osteotomy of the mandible
    Tebbutt, J. E.
    Markose, G.
    Graham, R. M.
    ANNALS OF THE ROYAL COLLEGE OF SURGEONS OF ENGLAND, 2019, 101 (08) : 619 - 620
  • [25] In vitro biomechanical evaluation of sagittal split osteotomy fixation with a specifically designed miniplate
    Pereira Filho, V. A.
    Iamashita, H. Y.
    Monnazzi, M. S.
    Gabrielli, M. F. R.
    Vaz, L. G.
    Passeri, L. A.
    INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2013, 42 (03) : 316 - 320
  • [26] Infralingular Versus Supralingular Medial Osteotomy in Sagittal Split Osteotomy of the Mandible: A Randomized Control Study
    John, Bobby
    Poorna, T. Anish
    Joshna, E. K.
    Philip, George
    Bharathi, Arivarasan
    JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2023, 81 (03) : 280 - 286
  • [27] Assessment of Resorbable and Non-resorbable Fixation Systems in Sagittal Split Ramus Osteotomy: An In vitro Study
    Costa, Davani Latarullo
    Torres, Alexandre Machado
    Bergamaschi, Isabela Polesi
    Kluppel, Leandro Eduardo
    de Oliveira, Rogerio Belle
    Blessmann Weber, Joao Batista
    JOURNAL OF MAXILLOFACIAL & ORAL SURGERY, 2022, 21 (03): : 779 - 784
  • [28] Does the type of sagittal split ramus osteotomy influence fixation strength? Evaluation of the mechanical behavior of two types of fixation used in three types of sagittal split ramus osteotomy
    Oliveira, Soraya da Silva
    Mattos de Carvalho, Pedro Henrique
    Sverzut, Cassio Edvard
    Trivellato, Alexandre Elias
    ORAL AND MAXILLOFACIAL SURGERY-HEIDELBERG, 2022, 26 (04): : 633 - 639
  • [29] Does the type of sagittal split ramus osteotomy influence fixation strength? Evaluation of the mechanical behavior of two types of fixation used in three types of sagittal split ramus osteotomy
    Soraya da Silva Oliveira
    Pedro Henrique Mattos de Carvalho
    Cássio Edvard Sverzut
    Alexandre Elias Trivellato
    Oral and Maxillofacial Surgery, 2022, 26 : 633 - 639
  • [30] HORIZONTAL ROTATION OF THE CONDYLE AFTER SAGITTAL SPLIT OSTEOTOMY OF THE MANDIBLE
    CARTER, J
    LEONARD, M
    CAVANAUGH, G
    BRAND, J
    AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 1991, 99 (04) : 319 - 327