Primary stability of fixation methods for periprosthetic fractures of the humerus: a biomechanical investigation

被引:2
|
作者
Ott, Nadine [1 ]
Harland, Arne [1 ]
Hackl, Michael [1 ]
Prescher, Andreas [2 ]
Leschinger, Tim [1 ]
Mu, Lars Peter [1 ]
Wegmann, Kilian [1 ]
机构
[1] Univ Hosp Cologne, Dept Orthoped & Trauma Surg, Kerpener Str 62, D-50937 Cologne, Germany
[2] Rhein Westfal TH Aachen, Inst Mol & Cellular Anat Prosektur, Aachen, Germany
关键词
Periprosthetic fracture; revision total shoulder arthroplasty; long stem; biomechanical; locking plates; strain gauges; TOTAL SHOULDER ARTHROPLASTY; BONE-MINERAL DENSITY; HOUNSFIELD UNITS; COMPLICATIONS; MANAGEMENT; REVISION; ADJACENT;
D O I
10.1016/j.jse.2020.12.027
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: The incidence of periprosthetic fractures of the proximal humerus is gradually increasing, following an increase in reverse shoulder arthroplasties in recent years. Locking plate fixation and revision arthroplasty are both valuable treatment methods. However, the primary stability of fixation methods for periprosthetic fractures has not been investigated in detail. The aim of this study was to analyze and compare the primary stability of the common treatment measures. Materials and methods: Cemented reverse total shoulder arthroplasty (Delta Xtend; DePuy Synthes, Warsaw, IN, USA) was performed in 5 shoulders, and a distal, mid-diaphysis humeral fracture (Wright and Cofield type B) was induced. The implant was left in place, and 3 distinct fixation scenarios were tested: osteosynthesis using 4.5-mm locking plate fixation (subgroup A), 4.5-mm locking plate fixation with an additional 3.5-mm locking plate (subgroup B), and 4.5-mm locking plate fixation with an additional K-wire cerclage (subgroup C). The specimens were tested in a biomechanical setup simulating activities of daily living including rotation. Strain gauges (4-wire strain at 120 U; Vishay Measurements Group, Chartres, France) mounted on the 4.5-mm locking plates were used to evaluate the strain of the fixation and to give an estimate of primary stability. Results: Regarding the simulation of activities of daily living, no statistically significant differences were found in the measured strains on the locking plate between subgroups A, B, and C. A maximum measured strain of 216.85 mm/m in subgroup A resulted in bending of the locking plate (length, 134 mm) of 0.03 mm. In subgroup B (277.01 mm/m), the plate strained 0.04 mm compared with a strain measurement of 0.01 mm in subgroup C (75.93 mm/m). Conclusion: Additional K-wire cerclages or additional 3.5-mm locked plating did not increase primary stability. With a stable prosthetic implant in place, 4.5-mm locked plating is sufficient to address periprosthetic humeral shaft fractures in the present in vitro setup. Level of evidence: Basic Science Study; Biomechanics (c) 2021 Journal of Shoulder and Elbow Surgery Board of Trustees. All rights reserved.
引用
收藏
页码:2184 / 2190
页数:7
相关论文
共 50 条
  • [41] Biomechanical evaluation of a new technique for external fixation of unstable supracondylar humerus fractures in children
    Hohloch, Lisa
    Konstantinidis, Lukas
    Wagner, Ferdinand C.
    Strohm, Peter C.
    Suedkamp, Norbert P.
    Reising, Kilian
    TECHNOLOGY AND HEALTH CARE, 2015, 23 (04) : 453 - 461
  • [42] Internal fixation of proximal humerus fractures. Influence of implant design on biomechanical properties
    Hessmann, MH
    Sternstein, W
    Krummenauer, F
    Hofmann, A
    Rommens, PM
    CHIRURG, 2005, 76 (02): : 167 - 174
  • [43] A BIOMECHANICAL ANALYSIS OF PROPHYLACTIC FIXATION FOR PATHOLOGICAL FRACTURES OF THE DISTAL 3RD OF THE HUMERUS
    DAMRON, TA
    HEINER, JP
    FREUND, EMN
    DAMRON, LA
    MCCABE, R
    VANDERBY, R
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1994, 76A (06): : 839 - 847
  • [44] Hybrid blade and locking plate fixation for proximal humerus fractures: a comparative biomechanical analysis
    Ali Jabran
    Chris Peach
    Zhenmin Zou
    Lei Ren
    BioMedical Engineering OnLine, 17
  • [45] The application of a dual-lead locking screw could enhance the reduction and fixation stability of the proximal humerus fractures: a biomechanical evaluation
    Lee, Eunju
    Jeong, Hyeon Jang
    Lee, Yeon Soo
    Oh, Joo Han
    FRONTIERS IN SURGERY, 2024, 11
  • [46] PROXIMAL HUMERUS FRACTURES: COMPARATIVE STUDY OF TWO DIFFERENT FIXATION METHODS
    Morelli, Ricardo Souza e Silva
    de Souza Travizanuto, Rodrigo Eisler
    ACTA ORTOPEDICA BRASILEIRA, 2010, 18 (02): : 79 - 84
  • [47] Primary Stability of Kyphoplasty in Incomplete Vertebral Body Burst Fractures in Osteoporosis: A Biomechanical Investigation
    Riesenbeck, Oliver
    Czarnowski, Niklas
    Raschke, Michael Johannes
    Oeckenpoehler, Simon
    Hartensuer, Rene
    BIOENGINEERING-BASEL, 2024, 11 (08):
  • [48] Displaced olecranon fractures in children - A biomechanical analysis of fixation methods
    Parent, Stefan
    Kedemeyer, Michelle
    Mahar, Andrew T.
    Anderson, Megan
    Taro, Frances
    Steinman, Suzanne
    Lalonde, Franqois
    Newton, Peter
    JOURNAL OF PEDIATRIC ORTHOPAEDICS, 2008, 28 (02) : 147 - 151
  • [49] Intraarticular calcaneus fractures: A biomechanical comparison of two fixation methods
    Bailey, EJ
    Waggoner, SM
    Albert, MJ
    Hutton, WC
    JOURNAL OF ORTHOPAEDIC TRAUMA, 1997, 11 (01) : 34 - 37
  • [50] Fixation for comminuted phalangeal fractures - A biomechanical study of five methods
    Lu, WW
    Furumachi, K
    Ip, WY
    Chow, SP
    JOURNAL OF HAND SURGERY-BRITISH AND EUROPEAN VOLUME, 1996, 21B (06): : 765 - 767