The application of a dual-lead locking screw could enhance the reduction and fixation stability of the proximal humerus fractures: a biomechanical evaluation

被引:1
|
作者
Lee, Eunju [1 ]
Jeong, Hyeon Jang [2 ]
Lee, Yeon Soo [1 ]
Oh, Joo Han [2 ]
机构
[1] Daegu Catholic Univ, Sch BioMed Sci, Dept BioMed Engn, Gyungbuk, South Korea
[2] Seoul Natl Univ, Coll Med, Bundang Hosp, Dept Orthopaed Surg, Seoul, South Korea
来源
FRONTIERS IN SURGERY | 2024年 / 11卷
关键词
proximal humerus; locked plating; dual-lead; lead ratio; compression; bone crush; insertion torque; stability; PLATE; BONE; STRENGTH;
D O I
10.3389/fsurg.2024.1333670
中图分类号
R61 [外科手术学];
学科分类号
摘要
Introduction: Bicortical screw fixation, which penetrates and fixes the near and far cortex of bone, has been conventionally used to achieve compressive fixation for fracture using screws. Open reduction and internal fixation using the locking plate are widely used for treating proximal humerus fractures. However, minimal contact between the bone and the locking plate can lead to an insufficient reduction. Theoretically, a dual-lead locking screw with different leads for the screw head and body could enhance the reduction and fixation stability of fragments in proximal humeral fractures without bicortical fixation, and achieve additional compression at the bone-plate-screw interface. This study assessed the insertion mechanics of the lead ratio of the dual-lead locking screw and its effect on the fixation stability of the proximal humerus fracture. Methods: A Multi-Fix (R) locking plating system composed of & empty;& empty; 3.5 mm locking screws and a locking plate was used to make a locked plating for Sawbone bone blocks and fourth-generation composite humeri. Two different types of Sawbone bone blocks were used to simulate the osteoporotic (10 PCF) and normal cancellous (20 PCF) bones. The lead of the screw head thread (L-head) was 0.8 mm, and that of the screw body (L-body) was 0.8, 1.25, 1.6, 2.0, and 2.4 mm, whose lead ratios (Rlead=L-body/L-head) were 1.0, 1.56, 2.0, 2.5, and 3.0, respectively. Results: The dual-lead locking screw elevated the compression between the locking plate and the bone. The elevation in the compression due to the dual-lead thread became weaker for the cancellous bone when the lead of the screw body was more than twice that of the screw head. The plate/humerus compression with strong bone quality withstood higher dual-lead-driven compression. Discussion: A dual-lead locking screw of L-body=1.25mm (R-lead=1.56) is recommended for maximum rotational stability for the locked humerus plating. The screws with over L-body=1.6mm (R-lead=2) have no advantage in terms of the failure torque and maximum torsional deformation. Any locking dual-lead screw with a body thread lead of <1.6 mm (R-lead=2) can be used without the risk of bone crush when surgeons require additional compression to the locked cancellous bone plating.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] How to enhance the stability of locking plate fixation of proximal humerus fractures? An overview of current biomechanical and clinical data
    Schliemann, Benedikt
    Waehnert, Dirk
    Theisen, Christina
    Herbort, Mirco
    Koesters, Clemens
    Raschke, Michael J.
    Weimann, Andre
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2015, 46 (07): : 1207 - 1214
  • [2] Locking Plates With Computationally Enhanced Screw Trajectories Provide Superior Biomechanical Fixation Stability of Complex Proximal Humerus Fractures
    Mischler, Dominic
    Schader, Jana Felicitas
    Dauwe, Jan
    Tenisch, Lara
    Gueorguiev, Boyko
    Windolf, Markus
    Varga, Peter
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2022, 10
  • [3] Open reduction and locking plate fixation of displaced proximal humerus fractures
    Kumar, Chandan
    Gupta, Anil Kumar
    Nath, Rohit
    Ahmad, Javed
    INDIAN JOURNAL OF ORTHOPAEDICS, 2013, 47 (02) : 156 - 160
  • [4] Open reduction and locking plate fixation of displaced proximal humerus fractures
    Chandan Kumar
    Anil Kumar Gupta
    Rohit Nath
    Javed Ahmad
    Indian Journal of Orthopaedics, 2013, 47 : 156 - 160
  • [5] Hybrid blade and locking plate fixation for proximal humerus fractures: a comparative biomechanical analysis
    Jabran, Ali
    Peach, Chris
    Zou, Zhenmin
    Ren, Lei
    BIOMEDICAL ENGINEERING ONLINE, 2018, 17
  • [6] Computational optimisation of screw orientations for improved locking plate fixation of proximal humerus fractures
    Mischler, Dominic
    Windolf, Markus
    Gueorguiev, Boyko
    Nijs, Stefaan
    Varga, Peter
    JOURNAL OF ORTHOPAEDIC TRANSLATION, 2020, 25 : 96 - 104
  • [7] Hybrid blade and locking plate fixation for proximal humerus fractures: a comparative biomechanical analysis
    Ali Jabran
    Chris Peach
    Zhenmin Zou
    Lei Ren
    BioMedical Engineering OnLine, 17
  • [8] Open Reduction and Internal Fixation of Proximal Humeral Fractures with Use of the Locking Proximal Humerus Plate
    Court-Brown, Charles M.
    McQueen, Margaret
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2009, 91A (11): : 2771 - 2771
  • [9] Open Reduction and Internal Fixation of Proximal Humeral Fractures with Use of the Locking Proximal Humerus Plate
    Konrad, G.
    Bayer, J.
    Hepp, P.
    Voigt, C.
    Oestern, H.
    Kaeaeb, N.
    Luo, C.
    Plecko, M.
    Wendt, K.
    Koestler, W.
    Suedkamp, N.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2010, 92A : 85 - 95
  • [10] IMPORTANCE OF CALCAR COMMINUTION AND SCREW LENGTH IN PROXIMAL HUMERUS FRACTURES TREATED WITH LOCKING PLATE FIXATION
    Raghava, Parthasarathy
    Thompson, Kevin J.
    Ponce, Brent A.
    Eberhardt, Alan W.
    PROCEEDINGS OF THE ASME SUMMER BIOENGINEERING CONFERENCE 2008, PTS A AND B, 2009, : 651 - 652