The application of a dual-lead locking screw could enhance the reduction and fixation stability of the proximal humerus fractures: a biomechanical evaluation

被引:1
|
作者
Lee, Eunju [1 ]
Jeong, Hyeon Jang [2 ]
Lee, Yeon Soo [1 ]
Oh, Joo Han [2 ]
机构
[1] Daegu Catholic Univ, Sch BioMed Sci, Dept BioMed Engn, Gyungbuk, South Korea
[2] Seoul Natl Univ, Coll Med, Bundang Hosp, Dept Orthopaed Surg, Seoul, South Korea
来源
FRONTIERS IN SURGERY | 2024年 / 11卷
关键词
proximal humerus; locked plating; dual-lead; lead ratio; compression; bone crush; insertion torque; stability; PLATE; BONE; STRENGTH;
D O I
10.3389/fsurg.2024.1333670
中图分类号
R61 [外科手术学];
学科分类号
摘要
Introduction: Bicortical screw fixation, which penetrates and fixes the near and far cortex of bone, has been conventionally used to achieve compressive fixation for fracture using screws. Open reduction and internal fixation using the locking plate are widely used for treating proximal humerus fractures. However, minimal contact between the bone and the locking plate can lead to an insufficient reduction. Theoretically, a dual-lead locking screw with different leads for the screw head and body could enhance the reduction and fixation stability of fragments in proximal humeral fractures without bicortical fixation, and achieve additional compression at the bone-plate-screw interface. This study assessed the insertion mechanics of the lead ratio of the dual-lead locking screw and its effect on the fixation stability of the proximal humerus fracture. Methods: A Multi-Fix (R) locking plating system composed of & empty;& empty; 3.5 mm locking screws and a locking plate was used to make a locked plating for Sawbone bone blocks and fourth-generation composite humeri. Two different types of Sawbone bone blocks were used to simulate the osteoporotic (10 PCF) and normal cancellous (20 PCF) bones. The lead of the screw head thread (L-head) was 0.8 mm, and that of the screw body (L-body) was 0.8, 1.25, 1.6, 2.0, and 2.4 mm, whose lead ratios (Rlead=L-body/L-head) were 1.0, 1.56, 2.0, 2.5, and 3.0, respectively. Results: The dual-lead locking screw elevated the compression between the locking plate and the bone. The elevation in the compression due to the dual-lead thread became weaker for the cancellous bone when the lead of the screw body was more than twice that of the screw head. The plate/humerus compression with strong bone quality withstood higher dual-lead-driven compression. Discussion: A dual-lead locking screw of L-body=1.25mm (R-lead=1.56) is recommended for maximum rotational stability for the locked humerus plating. The screws with over L-body=1.6mm (R-lead=2) have no advantage in terms of the failure torque and maximum torsional deformation. Any locking dual-lead screw with a body thread lead of <1.6 mm (R-lead=2) can be used without the risk of bone crush when surgeons require additional compression to the locked cancellous bone plating.
引用
收藏
页数:13
相关论文
共 50 条
  • [41] PEEK versus titanium locking plates for proximal humerus fracture fixation: a comparative biomechanical study in two- and three-part fractures
    Benedikt Schliemann
    Robert Seifert
    Christina Theisen
    Dominic Gehweiler
    Dirk Wähnert
    Martin Schulze
    Michael J. Raschke
    Andre Weimann
    Archives of Orthopaedic and Trauma Surgery, 2017, 137 : 63 - 71
  • [42] Biomechanical Analysis of the Proximal Femoral Locking Compression Plate: Do Quality of Reduction and Screw Orientation Influence Construct Stability?
    Zderic, Ivan
    Oh, Jong-Keon
    Stoffel, Karl
    Sommer, Christoph
    Helfen, Tobias
    Camino, Gaston
    Richards, Geoff
    Nork, Sean E.
    Gueorguiev, Boyko
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2018, 32 (02) : 67 - 74
  • [43] Open Reduction and Internal Fixation With Proximal Humerus Locking Plate, Screws, and Fibular Strut Graft: An Alternative to Shoulder Arthroplasty and Solution to Medial Instability and Varus Collapse in Comminuted Proximal Humerus Fractures
    Pop, Horea
    Morcovescu, Andrei
    Mihai, Maria M.
    Constantinescu, Andrei S.
    Bica, Florin
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2024, 16 (01)
  • [44] Biomechanical evaluation of the stability of extra-articular distal radius fractures fixed with volar locking plates according to the length of the distal locking screw
    Oh, Gyung-Hwan
    Kim, Hak-Sung
    Lee, Jung Il
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2021, 24 (08) : 922 - 932
  • [45] Screw-in-screw fixation of fragility sacrum fractures provides high stability without loosening-biomechanical evaluation of a new concept
    Zderic, Ivan
    Wagner, Daniel
    Schopper, Clemens
    Lodde, Moritz
    Richards, Geoff
    Gueorguiev, Boyko
    Rommens, Pol
    Acklin, Yves P.
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2021, 39 (04) : 761 - 770
  • [46] Biomechanical effect of medial cortical support and medial screw support on locking plate fixation in proximal humeral fractures with a medial gap: a finite element analysis
    Yang, Pan
    Zhang, Ying
    Liu, Jian
    Xiao, Jin
    Ma, Li Min
    Zhu, Chang Rong
    ACTA ORTHOPAEDICA ET TRAUMATOLOGICA TURCICA, 2015, 49 (02) : 203 - 209
  • [47] Biomechanical evaluation of a modified proximal humeral locking plate application for distal extra-articular diaphyseal humeral fractures
    Lim, Joon-Ryul
    Yoon, Tae-Hwan
    Choi, Yun-Rak
    Lee, Hwan-Mo
    Chun, Yong-Min
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2021, 39 (09) : 1877 - 1883
  • [48] Primary stability in locking plate fixation for proximal humeral fractures may be increased by using an additional nail osteosynthesis in combination with locking plate osteosynthesis-A biomechanical comparison
    Schoebel, Tobias
    Gemkow, Maximilian
    Wendler, Toni
    Schleifenbaum, Stefan
    Loeffler, Sabine
    Theopold, Jan
    Hepp, Pierre
    CLINICAL BIOMECHANICS, 2024, 114
  • [49] Open reduction and internal fixation versus minimally invasive plate osteosynthesis of unstable proximal humerus fractures treated with locking plate and intramedullary allograft: A retrospective study
    Rusimov, Lyubomir
    Baltov, Asen
    Enchev, Dian
    Gueorguiev, Boyko
    Prodanova, Krasimira
    Hadzhinikolova, Mariya
    Rusimov, Vladimir
    Rashkov, Mihail
    SHOULDER & ELBOW, 2024,
  • [50] Early locking plate removal following open reduction and internal fixation of proximal humeral fractures could prevent secondary implant-related complications
    Dimitriou, Dimitris
    Waldmann, Soeren
    Antoniadis, Alexander
    Liebhauser, Martin
    Helmy, Naeder
    Riede, Ulf
    JOURNAL OF ORTHOPAEDICS, 2020, 17 : 106 - 109