Primary stability of fixation methods for periprosthetic fractures of the humerus: a biomechanical investigation

被引:2
|
作者
Ott, Nadine [1 ]
Harland, Arne [1 ]
Hackl, Michael [1 ]
Prescher, Andreas [2 ]
Leschinger, Tim [1 ]
Mu, Lars Peter [1 ]
Wegmann, Kilian [1 ]
机构
[1] Univ Hosp Cologne, Dept Orthoped & Trauma Surg, Kerpener Str 62, D-50937 Cologne, Germany
[2] Rhein Westfal TH Aachen, Inst Mol & Cellular Anat Prosektur, Aachen, Germany
关键词
Periprosthetic fracture; revision total shoulder arthroplasty; long stem; biomechanical; locking plates; strain gauges; TOTAL SHOULDER ARTHROPLASTY; BONE-MINERAL DENSITY; HOUNSFIELD UNITS; COMPLICATIONS; MANAGEMENT; REVISION; ADJACENT;
D O I
10.1016/j.jse.2020.12.027
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: The incidence of periprosthetic fractures of the proximal humerus is gradually increasing, following an increase in reverse shoulder arthroplasties in recent years. Locking plate fixation and revision arthroplasty are both valuable treatment methods. However, the primary stability of fixation methods for periprosthetic fractures has not been investigated in detail. The aim of this study was to analyze and compare the primary stability of the common treatment measures. Materials and methods: Cemented reverse total shoulder arthroplasty (Delta Xtend; DePuy Synthes, Warsaw, IN, USA) was performed in 5 shoulders, and a distal, mid-diaphysis humeral fracture (Wright and Cofield type B) was induced. The implant was left in place, and 3 distinct fixation scenarios were tested: osteosynthesis using 4.5-mm locking plate fixation (subgroup A), 4.5-mm locking plate fixation with an additional 3.5-mm locking plate (subgroup B), and 4.5-mm locking plate fixation with an additional K-wire cerclage (subgroup C). The specimens were tested in a biomechanical setup simulating activities of daily living including rotation. Strain gauges (4-wire strain at 120 U; Vishay Measurements Group, Chartres, France) mounted on the 4.5-mm locking plates were used to evaluate the strain of the fixation and to give an estimate of primary stability. Results: Regarding the simulation of activities of daily living, no statistically significant differences were found in the measured strains on the locking plate between subgroups A, B, and C. A maximum measured strain of 216.85 mm/m in subgroup A resulted in bending of the locking plate (length, 134 mm) of 0.03 mm. In subgroup B (277.01 mm/m), the plate strained 0.04 mm compared with a strain measurement of 0.01 mm in subgroup C (75.93 mm/m). Conclusion: Additional K-wire cerclages or additional 3.5-mm locked plating did not increase primary stability. With a stable prosthetic implant in place, 4.5-mm locked plating is sufficient to address periprosthetic humeral shaft fractures in the present in vitro setup. Level of evidence: Basic Science Study; Biomechanics (c) 2021 Journal of Shoulder and Elbow Surgery Board of Trustees. All rights reserved.
引用
收藏
页码:2184 / 2190
页数:7
相关论文
共 50 条
  • [31] Biomechanical comparison of fixation methods in transverse patella fractures
    Scilaris, TA
    Grantham, JL
    Prayson, MJ
    Marshall, MP
    Hamilton, JJ
    Williams, JL
    JOURNAL OF ORTHOPAEDIC TRAUMA, 1998, 12 (05) : 356 - 359
  • [32] Capitellum fractures: A biomechanical evaluation of three fixation methods
    Elkowitz, SJ
    Polatsch, DB
    Egol, KA
    Kummer, FJ
    Koval, KJ
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2002, 16 (07) : 503 - 506
  • [33] A BIOMECHANICAL COMPARISON OF 2 METHODS OF FIXATION OF FRACTURES OF THE FOREARM
    JONES, DJ
    HENLEY, MB
    SCHEMITSCH, EH
    TENCER, AF
    JOURNAL OF ORTHOPAEDIC TRAUMA, 1995, 9 (03) : 198 - 206
  • [34] Biomechanical analysis of fixation methods for acetabular fractures: A review
    Ferrante, Hannah
    Schemitsch, Emil H.
    Zdero, Radovan
    Bagheri, Z. Shaghayegh
    MEDICAL ENGINEERING & PHYSICS, 2021, 89 : 51 - 62
  • [35] Locking Plates With Computationally Enhanced Screw Trajectories Provide Superior Biomechanical Fixation Stability of Complex Proximal Humerus Fractures
    Mischler, Dominic
    Schader, Jana Felicitas
    Dauwe, Jan
    Tenisch, Lara
    Gueorguiev, Boyko
    Windolf, Markus
    Varga, Peter
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2022, 10
  • [36] Biomechanical Comparison of 2 Different Locking Plate Fixation Methods in Vancouver B1 Periprosthetic Femur Fractures
    Pletka, Joshua D.
    Marsland, Daniel
    Belkoff, Stephen M.
    Mears, Simon C.
    Kates, Stephen L.
    GERIATRIC ORTHOPAEDIC SURGERY & REHABILITATION, 2011, 2 (02) : 51 - 55
  • [37] Fixation or Revision for Periprosthetic Fractures
    Minutillo, Gregory T.
    Karnuta, Jaret M.
    Koressel, Joseph
    Dehghani, Bijan
    Deangelis, Ryan D.
    Donegan, Derek J.
    Mehta, Samir
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2024, 106 (12): : 1054 - 1061
  • [38] Periprosthetic fractures of the humerus and scapula - Management and prevention
    Cameron, B
    Iannotti, JP
    ORTHOPEDIC CLINICS OF NORTH AMERICA, 1999, 30 (02) : 305 - +
  • [39] Hybrid blade and locking plate fixation for proximal humerus fractures: a comparative biomechanical analysis
    Jabran, Ali
    Peach, Chris
    Zou, Zhenmin
    Ren, Lei
    BIOMEDICAL ENGINEERING ONLINE, 2018, 17
  • [40] Biomechanical comparison of three internal fixation configurations for low transcondylar fractures of the distal humerus
    Zha, Yejun
    Hua, Kehan
    Huan, Yong
    Chen, Chen
    Sun, Weitong
    Ji, Shangwei
    Xiao, Dan
    Gong, Maoqi
    Jiang, Xieyuan
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2023, 54 (02): : 362 - 369