Novel method to measure the intrinsic spatial resolution in PET detectors based on monolithic crystals

被引:22
|
作者
Gonzalez-Montoro, Andrea [1 ]
Sanchez, Filomeno [1 ]
Bruyndonckx, Peter [2 ]
Canizares, Gabriel [1 ]
Benlloch, Jose M. [1 ]
Gonzalez, Antonio J. [1 ]
机构
[1] Univ Politecn Valencia, I3M, CSIC, E-46022 Valencia, Spain
[2] Bruker BBIO, Preclin Imaging, Kontich, Belgium
基金
欧洲研究理事会;
关键词
Gamma ray detectors; Positron Emission Tomography; Intrinsic spatial resolution; SiPM array; Monolithic blocks; POSITRON RANGE; DESIGN; SIPM; EMISSION; ARRAY; LYSO;
D O I
10.1016/j.nima.2018.12.056
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The main aim of this work is to provide a method to retrieve the intrinsic spatial resolution of a gamma-ray detector block based on monolithic crystals within an assembled scanner. This method consists on a discrimination of the data using a software collimation process. The results are compared with an alternative method of separating two detector blocks far enough to produce a "virtual" source collimation due to the geometric constraints on the allowed coincidence event angles. A theoretical model has been deduced to fit the measured light distribution profiles, allowing estimating the detector intrinsic spatial resolution. The detector intrinsic spatial resolution is expected to follow a Gaussian distribution and the positron-emitter source shape, given the small size of a Na-22 source with 0.25 mm in diameter, can be assumed to follow a Lorentzian profile. However, the collimation of the data modifies the source shape that is no longer a pure Lorentzian distribution. Therefore, the model is based on the convolution of a Gaussian shaped distribution (contribution of the detector) and a modified Lorentzian distribution (contribution of the collimated source profile) that takes into account the collimation effect. Three LYSO crystals geometries have been studied in the present work, namely a 10 mm thick trapezoidal monolithic block, and two rectangular monolithic blocks with thicknesses of 15 mm and 20 mm, respectively. All the blocks have size dimensions of 50 mm x 50 mm. The experimental results yielded an intrinsic detector spatial resolution of 0.64 +/- 0.02 mm, 0.82 +/- 0.02 and 1.07 +/- 0.03 mm, for the 10 mm, 15 mm and 20 mm thick blocks, respectively, when the source was placed at the center of the detector. The detector intrinsic spatial resolution was moreover evaluated across one of the axis of each crystal. These values worsen to an average value of 0.68 +/- 0.04 mm, 0.90 +/- 0.14 and 1.29 +/- 0.19 mm, respectively, when the whole crystal size is considered, as expected. These tests show an accurate method to determine the intrinsic spatial resolution of monolithic-based detector blocks, once assembled in the PET system.
引用
收藏
页码:58 / 67
页数:10
相关论文
共 50 条
  • [41] A PET Design Based on SiPM and Monolithic LYSO Crystals: Performance Evaluation
    Gonzalez, Antonio J.
    Aguilar, Albert
    Conde, Pablo
    Hernandez, Liczandro
    Moliner, Laura
    Vidal, Luis F.
    Sanchez, Filomeno
    Sanchez, Sebastian
    Correcher, Carlos
    Molinos, Cesar
    Barbera, Julio
    Lankes, Konrad
    Junge, Sven
    Bruckbauer, Thomas
    Bruyndonckx, Peter
    Benlloch, Jose M.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2016, 63 (05) : 2471 - 2477
  • [42] Analysis of Time Resolution in Monolithic Crystal PET detectors using different Digital Time Extraction Algorithms
    Monzo, J. M.
    Lerche, C. W.
    Martinez, J. D.
    Esteve, R.
    Toledo, J.
    Herrero, V.
    Ferrando, N.
    Aliaga, R. J.
    Colom, R. J.
    Gadea, R.
    Sanchez, F.
    Mora, F. J.
    Benlloch, J. M.
    Sebastia, A.
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 3370 - +
  • [43] Performance evaluation of the MOLECUBES β-CUBE-a high spatial resolution and high sensitivity small animal PET scanner utilizing monolithic LYSO scintillation detectors
    Krishnamoorthy, Srilalan
    Blankemeyer, Eric
    Mollet, Pieter
    Surti, Suleman
    Van Holen, Roel
    Karp, Joel S.
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (15):
  • [44] A New Method Based on Position Sensitive Detectors to Measure Spatial Angle of Large-scale Gyroidal
    Sun, Zhao
    Song, Qing
    Liu, Chang-qing
    Wu, Han-fei
    FRONTIERS OF MECHANICAL ENGINEERING AND MATERIALS ENGINEERING II, PTS 1 AND 2, 2014, 457-458 : 1352 - 1356
  • [45] Hybrid charge division multiplexing method for SiPM-based high-resolution PET detectors
    Park, Haewook
    Ko, Guen Bae
    Lee, Jae Sung
    JOURNAL OF NUCLEAR MEDICINE, 2016, 57
  • [46] Novel radiation hard microfabricated scintillation detectors with high spatial resolution
    Mapelli, A.
    Gorini, B.
    Haguenauer, M.
    Jiguet, S.
    Trivino, N. Vico
    Renaud, P.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 617 (1-3): : 400 - 401
  • [47] Spatial resolution in PET and the effect of gamma-ray interaction depth in block detectors
    St James, S
    Thompson, C
    MEDICAL PHYSICS, 2005, 32 (07) : 2408 - 2408
  • [48] Monolithic-scintillator TOF-PET detectors based on digital silicon photomultipliers
    Schaart, Dennis
    Borghi, Giacomo
    Van Dam, Herman
    Van der Lei, Gerben
    Seifert, Stefan
    Tabacchini, Valerio
    JOURNAL OF NUCLEAR MEDICINE, 2012, 53
  • [49] Performance of APD-based monolithic-crystal detectors for small animal PET
    Maas, Marnix C.
    van der Laan, D. J.
    Schaart, Dennis R.
    Bruyndonckx, Peter
    Lemaitre, Cedric
    van Eijk, Carel W. E.
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 2017 - 2021
  • [50] Simulated performance of a small-animal PET scanner based on monolithic scintillation detectors
    van der Laan, D. J.
    Maas, M. C.
    de Jong, H. W. A. M.
    Schaart, D. R.
    Bruyndonckx, P.
    Lemaitre, C.
    van Eijk, C. W. E.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 571 (1-2): : 227 - 230