Novel method to measure the intrinsic spatial resolution in PET detectors based on monolithic crystals

被引:22
|
作者
Gonzalez-Montoro, Andrea [1 ]
Sanchez, Filomeno [1 ]
Bruyndonckx, Peter [2 ]
Canizares, Gabriel [1 ]
Benlloch, Jose M. [1 ]
Gonzalez, Antonio J. [1 ]
机构
[1] Univ Politecn Valencia, I3M, CSIC, E-46022 Valencia, Spain
[2] Bruker BBIO, Preclin Imaging, Kontich, Belgium
基金
欧洲研究理事会;
关键词
Gamma ray detectors; Positron Emission Tomography; Intrinsic spatial resolution; SiPM array; Monolithic blocks; POSITRON RANGE; DESIGN; SIPM; EMISSION; ARRAY; LYSO;
D O I
10.1016/j.nima.2018.12.056
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The main aim of this work is to provide a method to retrieve the intrinsic spatial resolution of a gamma-ray detector block based on monolithic crystals within an assembled scanner. This method consists on a discrimination of the data using a software collimation process. The results are compared with an alternative method of separating two detector blocks far enough to produce a "virtual" source collimation due to the geometric constraints on the allowed coincidence event angles. A theoretical model has been deduced to fit the measured light distribution profiles, allowing estimating the detector intrinsic spatial resolution. The detector intrinsic spatial resolution is expected to follow a Gaussian distribution and the positron-emitter source shape, given the small size of a Na-22 source with 0.25 mm in diameter, can be assumed to follow a Lorentzian profile. However, the collimation of the data modifies the source shape that is no longer a pure Lorentzian distribution. Therefore, the model is based on the convolution of a Gaussian shaped distribution (contribution of the detector) and a modified Lorentzian distribution (contribution of the collimated source profile) that takes into account the collimation effect. Three LYSO crystals geometries have been studied in the present work, namely a 10 mm thick trapezoidal monolithic block, and two rectangular monolithic blocks with thicknesses of 15 mm and 20 mm, respectively. All the blocks have size dimensions of 50 mm x 50 mm. The experimental results yielded an intrinsic detector spatial resolution of 0.64 +/- 0.02 mm, 0.82 +/- 0.02 and 1.07 +/- 0.03 mm, for the 10 mm, 15 mm and 20 mm thick blocks, respectively, when the source was placed at the center of the detector. The detector intrinsic spatial resolution was moreover evaluated across one of the axis of each crystal. These values worsen to an average value of 0.68 +/- 0.04 mm, 0.90 +/- 0.14 and 1.29 +/- 0.19 mm, respectively, when the whole crystal size is considered, as expected. These tests show an accurate method to determine the intrinsic spatial resolution of monolithic-based detector blocks, once assembled in the PET system.
引用
收藏
页码:58 / 67
页数:10
相关论文
共 50 条
  • [31] Novel high resolution detectors for Positron Emission Tomography (PET)
    Belcari, N
    Camarda, M
    Del Guerra, A
    Herbert, DJ
    Vaiano, A
    Di Domenico, G
    Zavattini, G
    NUCLEAR PHYSICS B-PROCEEDINGS SUPPLEMENTS, 2003, 125 : 48 - 52
  • [32] High-resolution PET detector design: modelling components of intrinsic spatial resolution
    Stickel, JR
    Cherry, SR
    PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (02): : 179 - 195
  • [33] Intrinsic spatial resolution of semiconductor X-ray detectors: a simulation study
    Hoheisel, M
    Giersch, J
    Bernhardt, P
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 531 (1-2): : 75 - 81
  • [34] Performance of a PET/MR Insert for Mice Based on Monolithic Crystals
    Gonzalez, Antonio
    Gonzalez-Montoro, Andrea
    Freire, Marta
    Vidal, Luis
    Hernandez, Liczandro
    Correcher, Carlos
    Polo, Ramiro
    Orero, Abel
    Pastor, Gustavo
    Carrilero, Vicente
    Greeb, Thorsten
    Junge, Sven
    Molinos, Cesar
    Bruyndonckx, Peter
    Heidenreich, Michael
    JOURNAL OF NUCLEAR MEDICINE, 2020, 61
  • [35] Reducing Calibration Time in PET Systems Based on Monolithic Crystals
    Freire, Marta
    Canizares, Gabriel
    Echegoyen, Sara
    Gonzalez-Montoro, Andrea
    Gonzalez, Antonio J.
    FRONTIERS IN MEDICINE, 2021, 8
  • [36] A Novel Method for Estimating PET & SPECT Spatial Resolution Using Measured Recovery Coefficients
    Marquis, Harry
    Schmidtlein, Charles
    Bailey, Dale
    Beattie, Bradley
    Ramos, Juan Ocampo
    Carter, Lukas
    Willowson, Kathy
    Kesner, Adam
    JOURNAL OF NUCLEAR MEDICINE, 2024, 65
  • [37] Spatial resolution of the detectors based on Gas Electron Multipliers
    Kudryavtsev, V. N.
    Maltsev, T. V.
    Shekhtman, L. I.
    JOURNAL OF INSTRUMENTATION, 2020, 15 (05):
  • [38] Limits of a spatial resolution of the cascaded GEM based detectors
    Kudryavtsev, V. N.
    Maltsev, T. V.
    Shekhtman, L. I.
    JOURNAL OF INSTRUMENTATION, 2017, 12
  • [39] PET Detectors Based on Multi-Resolution SiPM Arrays
    Xie, Jiahao
    Wang, Haibo
    Cherry, Simon R.
    Du, Junwei
    IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES, 2024, 8 (05) : 493 - 500
  • [40] A Direct Image Reconstruction Algorithm for PET Scanners Based on Monolithic Crystals
    Iborra, A.
    Conde, P.
    Sanchez, S.
    Gonzalez, A. J.
    Rodriguez-Alvarez, M. J.
    Aguilar, A.
    Bellido, P.
    Diaz-Caballero, E.
    Garcia-Garrigos, J. J.
    Gonzalez-Montoro, A.
    Grau-Ruiz, D.
    Hernandez, L.
    Martos, E.
    Moliner, L.
    Rigla, J. P.
    Sanchez, E.
    Seimetz, M.
    Soriano, A.
    Valderas, J. C.
    Vidal, L. E.
    Benlloch, J. M.
    2016 IEEE NUCLEAR SCIENCE SYMPOSIUM, MEDICAL IMAGING CONFERENCE AND ROOM-TEMPERATURE SEMICONDUCTOR DETECTOR WORKSHOP (NSS/MIC/RTSD), 2016,