Spatial resolution in PET and the effect of gamma-ray interaction depth in block detectors

被引:0
|
作者
St James, S [1 ]
Thompson, C [1 ]
机构
[1] McGill Univ, Montreal Neurol Inst, Montreal, PQ H3A 2T5, Canada
关键词
D O I
10.1118/1.2030972
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The spatial resolution in PET is poorer than that of CT or MRI. All modern PET scanners use block detectors, i.e. clusters of scintillation crystals coupled to four photomultiplier tubes (PMTs). Some of the loss of spatial resolution in PET is attributed to the use of block detectors, because a photon that interacts with one crystal in the cluster may be incorrectly positioned, resulting in blurring of the reconstructed image, called the “block effect”. We examined the effect of changing gamma‐ray interaction depth in the scintillation crystals in detectors from the CTI HR+ and GE Advance PET scanners. We postulated that the depth at which the gamma‐ray interacts may contribute to the “block effect” blurring. The “block effect” was measured for both detectors, and it was found to be 1.2 mm for the central crystals and negligible for the edge crystals in the CTI HR+ block. However, it was 0.9 mm in all crystals of the GE Advance detector. In the CTI HR+ detector, a depth dependence on the positioning of the event was observed, as was a dependence on the crystal location (edge vs. centre). In the GE Advance detector, no such dependence was observed. These results suggest that the depth of interaction of an annihilation photon may contribute to the block effect in detectors that use crystals cut from a single scintillation crystal (pseudo‐discrete crystals). In detectors that use discrete crystals no additional blurring as a function of gamma‐ray interaction depth in the detectors was observed. © 2005, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:2408 / 2408
页数:1
相关论文
共 50 条
  • [1] Depth of interaction effect on timing resolution in PET block detectors
    Ivan, A
    Burr, KC
    Shao, YP
    LeBlanc, JW
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 4156 - 4158
  • [2] Investigation of the "block effect" on spatial resolution in PET detectors
    Tomic, N
    Thompson, CJ
    Casey, ME
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2005, 52 (03) : 599 - 605
  • [3] Investigation of the "block effect" on spatial resolution in PET detectors
    Tomic, N
    Thompson, CJ
    Casey, ME
    2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, : 2201 - 2205
  • [4] Maximum likelihood positioning for gamma-ray imaging detectors with depth of interaction measurement
    Lerche, Ch. W.
    Ros, A.
    Monzo, J. M.
    Aliaga, R. J.
    Ferrando, N.
    Martinez, J. D.
    Herrero, V.
    Esteve, R.
    Gadea, R.
    Colom, R. J.
    Toledo, J.
    Mateo, F.
    Sebastia, A.
    Sanchez, F.
    Benlloch, J. M.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2009, 604 (1-2): : 359 - 362
  • [5] The spatial response of CdZnTe gamma-ray detectors as measured by gamma-ray mapping
    Brunett, BA
    Van Scyoc, JM
    Schlesinger, TE
    James, RB
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 458 (1-2): : 76 - 84
  • [6] Electrode configuration and energy resolution in gamma-ray detectors
    Luke, PN
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 380 (1-2): : 232 - 237
  • [7] Electrode configuration and energy resolution in gamma-ray detectors
    Luke, P.N.
    Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1996, 380 (1-2): : 232 - 237
  • [8] Spatial pileup considerations for pixellated gamma-ray detectors
    Furenlid, LR
    Clarkson, E
    Marks, DG
    Barrett, HH
    1999 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1-3, 1999, : 713 - 717
  • [9] Spatial pileup considerations for pixellated gamma-ray detectors
    Furenlid, LR
    Clarkson, E
    Marks, DG
    Barrett, HH
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2000, 47 (04) : 1399 - 1403
  • [10] Deconvolution and spatial resolution of airborne gamma-ray surveys
    Billings, SD
    Minty, BR
    Newsam, GN
    GEOPHYSICS, 2003, 68 (04) : 1257 - 1266