Development of a flow standard to enable highly reproducible measurements of deformability of stored red blood cells in a microfluidic device

被引:15
|
作者
Robidoux, Jonathan [1 ]
Laforce-Lavoie, Audrey [1 ]
Charette, Steve J. [2 ]
Shevkoplyas, Sergey S. [3 ]
Yoshida, Tatsuro [4 ]
Brouard, Danny [1 ]
Lewin, Antoine [1 ]
机构
[1] Hema Quebec, Med Affairs & Innovat, Montreal, PQ, Canada
[2] Univ Laval, Biochem Microbiol & Bioinformat Dept, Montreal, PQ, Canada
[3] Univ Houston, Dept Biomed Engn, Houston, TX USA
[4] Hemanext Inc, Lexington, MA USA
关键词
ERYTHROCYTE DEFORMABILITY; MICROVASCULAR NETWORK; STORAGE LESION; TRANSFUSION; RHEOLOGY; PERFUSION; TOOL;
D O I
10.1111/trf.15770
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND Great deformability allows red blood cells (RBCs) to flow through narrow capillaries in tissues. A number of microfluidic devices with capillary-like microchannels have been developed to monitor storage-related impairment of RBC deformability during blood banking operations. This proof-of-concept study describes a new method to standardize and improve reproducibility of the RBC deformability measurements using one of these devices. STUDY DESIGN AND METHODS The rate of RBC flow through the microfluidic capillary network of the microvascular analyzer (MVA) device made of polydimethylsiloxane was measured to assess RBC deformability. A suspension of microbeads in a solution of glycerol in phosphate-buffered saline was developed to be used as an internal flow rate reference alongside RBC samples in the same device. RBC deformability and other in vitro quality markers were assessed weekly in six leukoreduced RBC concentrates (RCCs) dispersed in saline-adenine-glucose-mannitol additive solution and stored over 42 days at 4 degrees C. RESULTS The use of flow reference reduced device-to-device measurement variability from 10% to 2%. Repeated-measure analysis using the generalized estimating equation (GEE) method showed a significant monotonic decrease in relative RBC flow rate with storage from Week 0. By the end of storage, relative RBC flow rate decreased by 22 +/- 6% on average. CONCLUSIONS The suspension of microbeads was successfully used as a flow reference to increase reproducibility of RBC deformability measurements using the MVA. Deformability results suggest an early and late aging phase for stored RCCs, with significant decreases between successive weeks suggesting a highly sensitive measurement method.
引用
收藏
页码:1032 / 1041
页数:10
相关论文
共 50 条
  • [1] Deformability of Stored Red Blood Cells
    Barshtein, Gregory
    Pajic-Lijakovic, Ivana
    Gural, Alexander
    FRONTIERS IN PHYSIOLOGY, 2021, 12
  • [2] Extensional flow-based microfluidic device: deformability assessment of red blood cells in contact with tumor cells
    Vera Faustino
    Diana Pinho
    Tomoko Yaginuma
    Ricardo C. Calhelha
    Isabel C.F.R. Ferreira
    Rui Lima
    BioChip Journal, 2014, 8 : 42 - 47
  • [3] Extensional flow-based microfluidic device: deformability assessment of red blood cells in contact with tumor cells
    Faustino, Vera
    Pinho, Diana
    Yaginuma, Tomoko
    Calhelha, Ricardo C.
    Ferreira, Isabel C. F. R.
    Lima, Rui
    BIOCHIP JOURNAL, 2014, 8 (01) : 42 - 47
  • [4] A simple microfluidic device for the deformability assessment of blood cells in a continuous flow
    Rodrigues, Raquel O.
    Pinho, Diana
    Faustino, Vera
    Lima, Rui
    BIOMEDICAL MICRODEVICES, 2015, 17 (06)
  • [5] A simple microfluidic device for the deformability assessment of blood cells in a continuous flow
    Raquel O. Rodrigues
    Diana Pinho
    Vera Faustino
    Rui Lima
    Biomedical Microdevices, 2015, 17
  • [6] Electrical measurement of red blood cell deformability on a microfluidic device
    Zheng, Yi
    Nguyen, John
    Wang, Chen
    Sun, Yu
    LAB ON A CHIP, 2013, 13 (16) : 3275 - 3283
  • [7] A Comparative Study of Image Processing Methods for the Assessment of the Red Blood Cells Deformability in a Microfluidic Device
    Faustino, Vera
    Catarino, Susana O.
    Pinho, Diana
    Minas, Graca
    Lima, Rui
    VIPIMAGE 2017, 2018, 27 : 923 - 929
  • [8] Measuring Red Blood Cell Deformability Using a Microfluidic Device: Development of an External Standard Based Approach to Reduce Chip-to-Chip Variations
    Robidoux, Jonathan
    Ducas, Eric
    Laforce-Lavoie, Audrey
    Brouard, Danny
    TRANSFUSION, 2018, 58 : 205A - 205A
  • [9] DEFORMABILITY OF STORED RED-BLOOD-CELLS - RELATIONSHIP TO DEGREE OF PACKING
    CARD, RT
    MOHANDAS, N
    PERKINS, HA
    SHOHET, SB
    TRANSFUSION, 1982, 22 (02) : 96 - 101
  • [10] A DEVICE FOR THE ELECTRICAL MEASUREMENT OF THE DEFORMABILITY OF RED-BLOOD-CELLS
    ROGGENKAMP, HG
    JUNG, F
    KIESEWETTER, H
    BIOMEDIZINISCHE TECHNIK, 1983, 28 (05): : 100 - 104