Modeling and Estimation of Tip Contact Force for Steerable Ablation Catheters

被引:67
|
作者
Khoshnam, Mahta [1 ,2 ]
Skanes, Allan C. [3 ,4 ]
Patel, Rajni V. [1 ,2 ,5 ]
机构
[1] Univ Western Ontario, Dept Elect & Comp Engn, London, ON N6A 5A5, Canada
[2] Canadian Surg Technol & Adv Robot, London, ON N6A 5A5, Canada
[3] Univ Western Ontario, Schulich Sch Med & Dent, London, ON N6A 5A5, Canada
[4] Univ Western Ontario Hosp, London Hlth Sci Ctr, Electrophysiol Lab, London, ON N6A 5A5, Canada
[5] Univ Western Ontario, Dept Surg, London, ON N6A 5A5, Canada
基金
加拿大创新基金会; 加拿大自然科学与工程研究理事会;
关键词
Ablation catheter; contact force; kinematics; modeling;
D O I
10.1109/TBME.2015.2389615
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective: The efficacy of catheter-based cardiac ablation procedures can be significantly improved if real-time information is available concerning contact forces between the catheter tip and cardiac tissue. However, the widely used ablation catheters are not equipped for force sensing. This paper proposes a technique for estimating the contact forces without direct force measurements by studying the changes in the shape of the deflectable distal section of a conventional 7-Fr catheter (henceforth called the "deflectable distal shaft," the "deflectable shaft," or the "shaft" of the catheter) in different loading situations. Method: First, the shaft curvature when the tip is moving in free space is studied and based on that, a kinematic model for the deflectable shaft in free space is proposed. In the next step, the shaft shape is analyzed in the case where the tip is in contact with the environment, and it is shown that the curvature of the deflectable shaft provides useful information about the loading status of the catheter and can be used to define an index for determining the range of contact forces exerted by the ablation tip. Results: Experiments with two different steerable ablation catheters show that the defined index can detect the range of applied contact forces correctly in more than 80% of the cases. Based on the proposed technique, a framework for obtaining contact force information by using the shaft curvature at a limited number of points along the deflectable shaft is constructed. Conclusion: The proposed kinematic model and the force estimation technique can be implemented together to describe the catheter's behavior before contact, detect tip/tissue contact, and determine the range of contact forces. Significance: This study proves that the flexibility of the catheter's distal shaft provides a means of estimating the force exerted on tissue by the ablation tip.
引用
收藏
页码:1404 / 1415
页数:12
相关论文
共 50 条
  • [21] Model-Based Force Estimation of a Steerable Ablation Catheter in the Presence of the Blood Flow
    Tabatabaei, Seyyede Shahrzad
    Kheiri, Mojtaba
    Dargahi, Javad
    2024 IEEE INTERNATIONAL SYMPOSIUM ON ROBOTIC AND SENSORS ENVIRONMENTS, ROSE 2024, 2024,
  • [22] The Impact of Steerable Sheaths on Stability of the Catheter Ablation for Paroxysmal Atrial Fibrillation, Evaluated by Contact Force
    Xu, Chao
    Xu, Buyun
    Peng, Fang
    Pan, Jie
    Lou, Yuanqing
    Yan, Longhui
    Xing, Yangbo
    IRANIAN RED CRESCENT MEDICAL JOURNAL, 2022, 24 (04)
  • [23] Force Estimation on Steerable Catheters through Learning-from-Simulation with ex-vivo Validation
    Sayadi, Amir
    Nourani, Hamid Reza
    Jolaei, Mohammad
    Dargahi, Javad
    Hooshiar, Amir
    2021 INTERNATIONAL SYMPOSIUM ON MEDICAL ROBOTICS (ISMR), 2021,
  • [24] Comparison of the accuracy of contact force measurement in four commercially available force-sensing ablation catheters
    Kueffer, T.
    Haeberlin, A.
    Knecht, S.
    Baldinger, S. H.
    Servatius, H.
    Madaffari, A.
    Seiler, J.
    Muhl, A.
    Franzeck, F.
    Asatryan, B.
    Noti, F.
    Tanner, H.
    Roten, L.
    Reichlin, T.
    SWISS MEDICAL WEEKLY, 2022, 152 : 16S - 16S
  • [25] INCREASED RATE OF POSTOPERATIVE PERICARDITIS WITH CONTACT FORCE SENSING CATHETERS IN ATRIAL FIBRILLATION ABLATION
    de Oliveira Filho, Cilomar Martins
    Tung, Roderick
    Bradfield, Jason
    Vaseghi, Marmar
    Ajijola, Olujimi
    Macias, Carlos
    Boyle, Noel
    Shivkumar, Kalyanam
    Buch, Eric
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2016, 67 (13) : 758 - 758
  • [26] Interest of irrigated tip catheters compare to 8 mm tip catheters for ablation of typical atrial flutter
    Lagrange, A
    Lagrange, P
    Arentz, T
    Boveda, S
    Defaye, P
    Virot, P
    Blanc, P
    EUROPEAN HEART JOURNAL, 2000, 21 : 351 - 351
  • [27] Similar outcomes with manual contact force ablation catheters and traditional catheters in the treatment of outflow tract premature ventricular complexes
    Abraham, Pal
    Ambrus, Mercedesz
    Herczeg, Szilvia
    Szegedi, Nandor
    Nagy, Klaudia Vivien
    Sallo, Zoltan
    Osztheimer, Istvan
    Szeplaki, Gabor
    Tahin, Tamas
    Merkely, Bela
    Geller, Laszlo
    EUROPACE, 2021, 23 (04): : 596 - 602
  • [28] Contact force-sensing catheters: correlation between contact force ablation algorithms and lesion size in an ex vivo porcine model
    Franco, E.
    Rodriguez-Munoz, D.
    Matia, R.
    Hernandez-Madrid, A.
    Zamorano, J. L.
    Moreno, J.
    EUROPEAN HEART JOURNAL, 2016, 37 : 1091 - 1091
  • [29] Irrigated Tip Catheters for Radiofrequency Ablation in Ventricular Tachycardia
    Muessigbrodt, Andreas
    Grothoff, Matthias
    Dinov, Borislav
    Kosiuk, Jedrzej
    Richter, Sergio
    Sommer, Philipp
    Breithardt, Ole A.
    Rolf, Sascha
    Bollmann, Andreas
    Arya, Arash
    Hindricks, Gerhard
    BIOMED RESEARCH INTERNATIONAL, 2015, 2015