Distal interphalangeal joint arthrodesis with a headless compression screw: morphometric and functional analyses

被引:23
|
作者
Song, Joo-Hyoun [2 ]
Lee, Joo-Yup [2 ]
Chung, Yang-Guk [3 ]
Park, Il-Jung [1 ]
机构
[1] Catholic Univ Korea, Bucheon St Marys Hosp, Dept Orthoped Surg, Puchon 420717, South Korea
[2] Catholic Univ Korea, St Vincent Hosp, Dept Orthoped Surg, Suwon 442723, South Korea
[3] Catholic Univ Korea, Seoul St Marys Hosp, Dept Orthoped Surg, Seoul, South Korea
关键词
Distal interphalangeal joint; Arthrodesis; Headless compression screw; ARTHROPLASTY; COMPLICATIONS; FIXATION; HAND;
D O I
10.1007/s00402-011-1413-3
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We performed radiologic measurement of the distal and middle phalanges in volunteers to determine the size of a headless compression screw suitable for distal interphalangeal (DIP) joint arthrodesis in Korean subjects and report on clinical results using an acutrak fusion screw. Radiologic measurements on the distal and middle phalanx were performed to determine the optimal size of screw. Five hundred fingers from 50 adult Koreans without any abnormality on plain radiographs were selected and anteroposterior and lateral radiographs were obtained for measurements. For the distal phalanx, the narrowest diameter of the cortical bone was measured to determine the minimal diameter of the screw that would not penetrate the cortex. For the middle phalanx, the narrowest diameter of the medullary canal was measured to determine the appropriate size of the screw for fixation. Between May 2004 and December 2007, there were 23 fingers in 22 patients (6 male, 16 female) that had finger DIP joint or thumb IP joint arthrodesis performed with the acutrak fusion screws. At the final follow up, time to union, complications, clinical fusion angle, pinch power, visual analogue score (VAS) for pain and the Korean version of the disabilities of the arm, shoulder and hand (DASH) questionnaire were assessed. In the distal phalanx, the narrowest diameter of the cortex was 2.64 +/- A 0.51 mm for the little finger. In the middle phalanx, the narrowest diameter of the medullary canal was 1.83 +/- A 0.50 mm for the little finger and 4.17 +/- A 0.68 for the thumb. The mean time to union was 10 weeks (range 8-12). The mean clinical fusion angle of the DIP joint was 11.9A degrees (range 0-20). The VAS pain score was 0.4 (range 0-3). Pinch power was 75% of the normal side. The average Korean DASH score was 5 points (range 0-8). We experienced one intraoperative fixation failure for thumb IP joint arthrodesis caused by a wide medullary canal of the proximal phalanx. The acutrak fusion screw was a feasible and adequate tool for DIP arthrodesis, particularly in Koreans. However, meticulous attention to technique was important to avoid complications in some little fingers. If preoperative radiographs suggest the thumb has a wide medullary canal, alternate methods of fixation should be considered.
引用
收藏
页码:663 / 669
页数:7
相关论文
共 50 条
  • [31] Arthrodesis of the distal interphalangeal Joint - a literature Review
    Vonderlind, Hans Christoph
    Eisenschenk, Andreas
    Struwer, Johannes
    Millrose, Michael
    HANDCHIRURGIE MIKROCHIRURGIE PLASTISCHE CHIRURGIE, 2022, 54 (01) : 5 - 20
  • [32] A Systematic Review of Distal Interphalangeal Joint Arthrodesis
    Dickson, D. R.
    Mehta, S. S.
    Nuttall, D.
    Ng, C. Y.
    JOURNAL OF HAND AND MICROSURGERY, 2014, 6 (02) : 74 - 84
  • [33] Compressive Force With 2-Screw and 3-Screw Subtalar Joint Arthrodesis With Headless Compression Screws
    Matsumoto, Takumi
    Glisson, Richard R.
    Reidl, Markus
    Easley, Mark E.
    FOOT & ANKLE INTERNATIONAL, 2016, 37 (12) : 1357 - 1363
  • [34] DIGITAL ARTHRODESIS BY SCREWING .1. ARTHRODESIS OF DISTAL INTERPHALANGEAL JOINT
    ISELIN, F
    PRADET, G
    NOUVELLE PRESSE MEDICALE, 1972, 1 (22): : 1520 - &
  • [35] Metacarpophalangeal and interphalangeal joint arthrodesis: a comparative study between tension band and compression screw fixation
    Breyer, J. M.
    Vergara, P.
    Parra, L.
    Sotelo, P.
    Bifani, A.
    Andrade, F.
    JOURNAL OF HAND SURGERY-EUROPEAN VOLUME, 2015, 40 (04) : 374 - 378
  • [36] Distal Interphalangeal Joint Arthrodesis With the Reverse Fix Nail
    Matsumoto, Takumi
    Nakamura, Ichiro
    Miura, Ayumi
    Momoyama, Gen
    Ito, Katsumi
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2013, 38A (07): : 1301 - 1306
  • [37] INTERPHALANGEAL JOINT ARTHRODESIS WITH OBLIQUE PLACEMENT OF AN AO LAG SCREW
    TEOH, LC
    YEO, SJ
    SINGH, I
    JOURNAL OF HAND SURGERY-BRITISH AND EUROPEAN VOLUME, 1994, 19B (02): : 208 - 211
  • [38] Arthrodesis of the Distal Interphalangeal Joint of the Finger Using an Intramedullary Implant
    Barabas, P.
    Pavlicny, R.
    ACTA CHIRURGIAE ORTHOPAEDICAE ET TRAUMATOLOGIAE CECHOSLOVACA, 2024, 91 (05) : 282 - 288
  • [39] Percutaneous in situ versus open arthrodesis of the distal interphalangeal joint
    Renfree, K. J.
    JOURNAL OF HAND SURGERY-EUROPEAN VOLUME, 2015, 40 (04) : 379 - 383
  • [40] Does distal interphalangeal joint arthrodesis affect proximal interphalangeal joint arthroplasty outcomes in the same finger?
    Startseva, Xenia
    Marks, Miriam
    Schweizer, Andreas
    Herren, Daniel B.
    Schindele, Stephan
    JOURNAL OF HAND SURGERY-EUROPEAN VOLUME, 2023, 48 (10) : 1056 - 1061