A test of silicon photomultipliers as readout for PET

被引:134
|
作者
Otte, AN
Barral, J
Dolgoshein, B
Hose, J
Klemin, S
Lorenz, E
Mirzoyan, R
Popova, E
Teshima, M
机构
[1] Max Planck Inst Phys & Astrophys, D-80805 Munich, Germany
[2] Ecole Polytech, F-91128 Palaiseau, France
[3] Moscow Engn Phys Inst, Moscow 115409, Russia
[4] Pulsar Enterprise, Moscow, Russia
关键词
positron emission tomography (PET); silicon photomultiplier (SiPM); lutetium yttrium orthosilicate (LYSO); avalanche photo diode (APD);
D O I
10.1016/j.nima.2005.02.014
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The silicon photomultiplier (SiPM) is a novel photon detector based on Geiger mode operating avalanche photodiodes. In this paper, we present results from a test, demonstrating the feasibility of SiPM as readout elements in scintillator-based positron emission tomography (PET). As scintillator we use the newly developed LYSO crystals having similar characteristics as LSO. With our setup we measure an energy resolution of about 22% and a time resolution of a single crystal element of (1.51 +/- 0.07) ns, both full-width at half-maximum. A significant improvement in time resolution could be achieved by triggering on the first photoelectron in the signal. We also present the coincidence rate of two detector channels vs. the position of a small point-like Na-22 positron source. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:705 / 715
页数:11
相关论文
共 50 条
  • [41] Evaluation of HPK and FBK Silicon Photomultipliers for TOF and PET Applications
    Ambrosi, G.
    Azzarello, P.
    Battiston, R.
    Ionica, M.
    Pignatel, G. U.
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 1566 - 1571
  • [42] Performance of Digital Silicon Photomultipliers for Time of Flight PET Scanners
    Marcinkowski, R.
    Espana, S.
    Thoen, H.
    Vandenberghe, S.
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 2825 - 2829
  • [43] Multi-channel readout electronics of silicon photomultipliers for plastic scintillating fiber detector
    Wu, Zibing
    Hu, Kun
    Li, Huiling
    Ren, Xiangxiang
    Wang, Hongbo
    Xu, Weiwei
    JOURNAL OF INSTRUMENTATION, 2024, 19 (11):
  • [44] PRELIMINARY RESULTS OF PLASTIC SCINTILLATORS DETECTOR READOUT WITH SILICON PHOTOMULTIPLIERS FOR COSMIC RAYS STUDIES
    Stanca, D.
    ROMANIAN REPORTS IN PHYSICS, 2012, 64 (03) : 831 - 840
  • [45] A 4D-PET block detector based on Silicon Photomultipliers
    Marcatili, Sara
    Belcari, Nicola
    Bisogni, Maria G.
    Collazuol, Gianmaria
    Pedreschi, Elena
    Spinella, Franco
    Corsi, Francesco
    Foresta, Maurizio
    Marzocca, Cristoforo
    Matarrese, Gianvito
    Sportelli, Giancarlo
    Guerra, Pedro
    Santos, Andres
    Del Guerra, Alberto
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 1949 - 1953
  • [46] Test procedures and characterization of silicon photomultipliers for photometric detection in astrophysics
    Ambrosi, G.
    Ambrosio, M.
    Aramo, C.
    Bertucci, B.
    Bissaldi, E.
    Bitossi, M.
    Boiano, A.
    Bonavolonta, C.
    Caprai, M.
    Consiglio, L.
    Fiandrini, E.
    Giglietto, N.
    Giordano, F.
    Lonica, M.
    Licciulli, F.
    Loporchio, S.
    Pantaleo, F. R.
    Paoletti, R.
    Rugliancich, A.
    Tosti, L.
    Vagelli, V.
    di Venere, L.
    NUOVO CIMENTO C-COLLOQUIA AND COMMUNICATIONS IN PHYSICS, 2020, 43 (2-3):
  • [47] Evaluation of silicon photomultipliers: A promising new detector for MR compatible PET
    McElroy, David P.
    Saveliev, Valeri
    Reznik, Alla
    Rowlands, John A.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 571 (1-2): : 106 - 109
  • [48] Delay grid multiplexing: simple time-based multiplexing and readout method for silicon photomultipliers
    Won, Jun Yeon
    Ko, Guen Bae
    Lee, Jae Sung
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (19): : 7113 - 7135
  • [49] Development of a light readout based on silicon photomultipliers for a wavelength-shifting fibre gamma camera
    Castro, I. F.
    Soares, A. J.
    Moutinho, L. M.
    Veloso, J. F. C. A.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2012, 695 : 298 - 302
  • [50] Performance of FBK low-afterpulse NUV silicon photomultipliers for PET application
    Ferri, A.
    Acerbi, F.
    Gola, A.
    Paternoster, G.
    Piemonte, C.
    Zorzi, N.
    JOURNAL OF INSTRUMENTATION, 2016, 11