Effects of acetabular abduction on cup wear rates in total hip arthroplasty

被引:112
|
作者
Del Schutte, H [1 ]
Lipman, AJ [1 ]
Bannar, SM [1 ]
Livermore, JT [1 ]
Ilstrup, D [1 ]
Morrey, BF [1 ]
机构
[1] Med Univ S Carolina, Dept Orthopaed Surg, Charleston, SC 29425 USA
来源
JOURNAL OF ARTHROPLASTY | 1998年 / 13卷 / 06期
关键词
hip arthroplasty; polyethylene wear rates; acetabular abduction;
D O I
10.1016/S0883-5403(98)80003-X
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
This study was designed to determine the effect of acetabular abduction on the polyethylene wear rates of the acetabular component. The hypothesis of this study is that acetabular placement, in particular abduction, effects contact forces and therefore polyethylene wear. A total of 364 total hip arthroplasties that were performed between 1974 and 1976 were included in this study. The cemented all-polyethylene acetabular components were the same for each case and came from a single supplier. The polyethylene wear rates were calculated by measuring the shortest radius from the center of the prosthetic femoral head to a point on the outer surface of the acetabular cup. An immediate postoperative radiograph was compared with a follow-up radiograph at least 9.5 years later. A standardized radiograph was used to prevent differences due to magnification. The abduction or inclination of the acetabular cups was measured in all of the patients. The amount of acetabular cup abduction was measured relative to the ischial tuberosity line. The mean abduction was 44.1 degrees with a standard deviation of 9.2 degrees and a median of 44 degrees. The range of abduction was from 0 degrees to 85 degrees. This analysis failed to show a correlation between the amount of acetabular abduction and polyethylene wear rates (Pearson's correlation coefficient = 0.0679; P = .20). With a sample size of 364, there is over a 95% change (statistical power) of detecting an underlying true correlation between wear and abduction that is greater than or equal to 0.20. We were unable to demonstrate a difference in wear that would correlate with the differing degrees of abduction of the acetabular socket. We felt that the radiographic measurements of wear were quite accurate. This method of determining acetabular abduction has also been documented and supported in previous literature and has allowed us to accurately determine acetabular abduction. The results of our study demonstrate that within the normal ranges described by our study, polyethylene wear did not significantly increase with increased abduction of the acetabular component.
引用
收藏
页码:621 / 626
页数:6
相关论文
共 50 条
  • [21] Ceramic acetabular liner fracture in total hip arthroplasty with a ceramic sandwich cup
    Hasegawa, M
    Sudo, A
    Hirata, H
    Uchida, A
    JOURNAL OF ARTHROPLASTY, 2003, 18 (05): : 658 - 661
  • [22] Influence of acetabular cup thickness on seating and primary stability in total hip arthroplasty
    Ruhr, Miriam
    Baetz, Johanna
    Pueschel, Klaus
    Morlock, Michael M.
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2022, 40 (09) : 2139 - 2146
  • [23] Improving Acetabular Cup Orientation in Total Hip Arthroplasty by Using Smartphone Technology
    Peters, Frank M.
    Greeff, Richard
    Goldstein, Neal
    Frey, Chris T.
    JOURNAL OF ARTHROPLASTY, 2012, 27 (07): : 1324 - 1330
  • [24] ENDLERS ACETABULAR CUP IN REVISION ARTHROPLASTY OF THE HIP
    BYE, K
    REIGSTAD, A
    ACTA ORTHOPAEDICA SCANDINAVICA, 1987, 58 (03): : 341 - 342
  • [25] MDCT Image Based Assessment of Acetabular Cup Orientation in Total Hip Arthroplasty
    Kobashi, Syoji
    Iwasa, Kenjiro
    Fujishiro, Takaaki
    Hayashi, Shiya
    Hashimoto, Shingo
    Kuroda, Ryosuke
    Kurosaka, Masahiro
    Kamiura, Naotake
    2014 IEEE INTERNATIONAL CONFERENCE ON SYSTEMS, MAN AND CYBERNETICS (SMC), 2014, : 2557 - 2562
  • [26] Effect of acetabular modularity on polyethylene wear and osteolysis in total hip arthroplasty
    Young, AM
    Sychterz, CJ
    Hopper, RH
    Engh, CA
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2002, 84A (01): : 58 - 63
  • [27] Catastrophic Wear of an Acetabular Component Misdiagnosed as Total Hip Arthroplasty Dislocation
    Papaioannou, Ioannis
    Pantazidou, Georgia
    Chatziperi, Stamatia
    Baikousis, Andreas
    Korovessis, Panagiotis
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2020, 12 (11)
  • [28] The effect of joint loading on acetabular wear measurement in total hip arthroplasty
    Martell, JM
    Leopold, SS
    Liu, XL
    JOURNAL OF ARTHROPLASTY, 2000, 15 (04): : 512 - 518
  • [29] Effect of cup-medialization on flexion and abduction moment arms in total hip arthroplasty
    Rudiger, Hannes
    Florencio, Francesc Levrero
    Terrier, Alexandre
    SWISS MEDICAL WEEKLY, 2013, 143 : 31S - 31S
  • [30] The effect of total hip arthroplasty cup design on polyethylene wear rate
    Hamilton, WG
    Hopper, RH
    Ginn, SD
    Hammell, NP
    Engh, CA
    Engh, CA
    JOURNAL OF ARTHROPLASTY, 2005, 20 (07): : 63 - 72