Metal Artifact Reduction by Dual-Energy Computed Tomography Using Energetic Extrapolation A Systematically Optimized Protocol

被引:136
|
作者
Meinel, Felix G. [1 ]
Bischoff, Bernhard [1 ]
Zhang, Qiaowei [2 ]
Bamberg, Fabian [1 ]
Reiser, Maximilian F. [1 ]
Johnson, Thorsten R. C. [1 ]
机构
[1] Univ Munich, Dept Clin Radiol, D-81377 Munich, Germany
[2] Zhejiang Univ, Sch Med, Sir Run Run Shaw Hosp, Dept Radiol, Hangzhou 310003, Zhejiang, Peoples R China
关键词
metal implants; metal artifacts; dual-energy CT; dual-source CT; energetic extrapolation; CONE-BEAM CT; BONE REMOVAL; RENAL MASSES; IN-VITRO; ANGIOGRAPHY; PROJECTIONS; MULTISLICE; EXPERIENCE; IMPLANTS;
D O I
10.1097/RLI.0b013e31824c86a3
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: Energetic extrapolation is a promising strategy to reduce metal artifacts in dual-source computed tomography (DSCT). We performed this study to systematically optimize image acquisition parameters for this approach in a hip phantom and assess its value in a clinical study. Materials and Methods: Titanium and steel hip prostheses were placed in a standard hip phantom and a water tank and scanned on a DSCT scanner. Tube spectra, tube current ratio, collimation, pitch, and rotation time were optimized in a stepwise process. Artifacts were quantified by measuring the standard deviation of the computed tomography density in a doughnut-shaped region of interest placed around the prosthesis. A total of 22 adult individuals with metallic implants referred for computed tomography for a musculoskeletal indication were scanned using the optimized protocol. Degree of artifacts and diagnostic image quality were rated visually (0-10) and maximum streak intensity was measured. Results: Sn140/100 kVp proved superior to Sn140/80 kVp. There was a benefit for increasing tube current ratio from 1:1 to 3:1, but not beyond, in favor of the Sn140 kVp spectrum. Artifacts were less severe for a collimation of 32 x 0.6 mm as compared with 40 x 0.6 mm. A pitch of 0.5 at a rotation time of 0.5 seconds per rotation was preferable to other combinations with comparable scanning times. In the clinical study, increasing the extrapolated photon energy from 64 to 120 keV decreased the severity of artifacts from 8.0 to 2.0 (P < 0.001) and decreased streak intensity from 871 to 153 HU (P < 0.001). The median diagnostic image quality rating improved from 2.5 to 8.0 (P < 0.001). The median energy level visually perceived as optimal for diagnostic evaluation was 113 keV (range, 100-130 keV). Conclusions: Sn140/100 kVp with a tube current ratio of 3:1, a collimation of 32 x 0.6 mm, and extrapolated energies of 105 to 120 keV are optimal parameters for a dedicated DSCT protocol that effectively reduces metal artifacts by energetic extrapolation. The protocol effectively reduces metal artifacts in all types of metal implants. The optimized reconstructions yielded relevant additional findings.
引用
收藏
页码:406 / 414
页数:9
相关论文
共 50 条
  • [41] Imaging of nanoparticles with dual-energy computed tomography
    Ducote, J. L.
    Alivov, Y.
    Molloi, S.
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (07): : 2031 - 2044
  • [42] Dual-energy computed tomography of canine uroliths
    Nykamp, Stephanie G.
    AMERICAN JOURNAL OF VETERINARY RESEARCH, 2017, 78 (10) : 1150 - 1155
  • [43] Beam hardening artifact reduction using dual energy computed tomography: implications for myocardial perfusion studies
    Rodriguez-Granillo, Gaston A.
    Carrascosa, Patricia
    Cipriano, Silvina
    De Zan, Macarena
    Deviggiano, Alejandro
    Capunay, Carlos
    Cury, Ricardo C.
    CARDIOVASCULAR DIAGNOSIS AND THERAPY, 2015, 5 (01) : 79 - 85
  • [44] Dual-energy Computed Tomography Applications in Uroradiology
    Park, Jong
    Chandarana, Hersh
    Macari, Michael
    Megibow, Alec J.
    CURRENT UROLOGY REPORTS, 2012, 13 (01) : 55 - 62
  • [45] Dual-energy computed tomography: what is it useful for?
    Delgado Sanchez-Gracian, C.
    Martinez Rodriguez, C.
    Trinidad Lopez, C.
    RADIOLOGIA, 2013, 55 (04): : 346 - 352
  • [46] Dual-energy computed tomography in pulmonary embolism
    Lu, G-M
    Wu, S-Y
    Yeh, B. M.
    Zhang, L-J
    BRITISH JOURNAL OF RADIOLOGY, 2010, 83 (992): : 707 - 718
  • [47] Dual-Energy Computed Tomography in Genitourinary Imaging
    Mileto, Achille
    Marin, Daniele
    RADIOLOGIC CLINICS OF NORTH AMERICA, 2017, 55 (02) : 373 - +
  • [48] Pediatric Applications of Dual-Energy Computed Tomography
    Pena-Trujillo, Valeria
    Gallo-Bernal, Sebastian
    Tung, Erik L.
    Gee, Michael S.
    RADIOLOGIC CLINICS OF NORTH AMERICA, 2023, 61 (06) : 1069 - 1083
  • [49] Dual-Energy Computed Tomography : Technological Considerations
    Chung, Ryan
    Dane, Bari
    Yeh, Benjamin M.
    Morgan, Desiree E.
    Sahani, Dushyant V.
    Kambadakone, Avinash
    RADIOLOGIC CLINICS OF NORTH AMERICA, 2023, 61 (06) : 945 - 961
  • [50] Dual-Energy Computed Tomography in Cardiac Imaging
    Boettcher, Benjamin
    Zsarnoczay, Emese
    Varga-Szemes, Akos
    Schoepf, Uwe Joseph
    Meinel, Felix G.
    van Assen, Marly
    De Cecco, Carlo N.
    RADIOLOGIC CLINICS OF NORTH AMERICA, 2023, 61 (06) : 995 - 1009