Metal Artifact Reduction by Dual-Energy Computed Tomography Using Energetic Extrapolation A Systematically Optimized Protocol

被引:136
|
作者
Meinel, Felix G. [1 ]
Bischoff, Bernhard [1 ]
Zhang, Qiaowei [2 ]
Bamberg, Fabian [1 ]
Reiser, Maximilian F. [1 ]
Johnson, Thorsten R. C. [1 ]
机构
[1] Univ Munich, Dept Clin Radiol, D-81377 Munich, Germany
[2] Zhejiang Univ, Sch Med, Sir Run Run Shaw Hosp, Dept Radiol, Hangzhou 310003, Zhejiang, Peoples R China
关键词
metal implants; metal artifacts; dual-energy CT; dual-source CT; energetic extrapolation; CONE-BEAM CT; BONE REMOVAL; RENAL MASSES; IN-VITRO; ANGIOGRAPHY; PROJECTIONS; MULTISLICE; EXPERIENCE; IMPLANTS;
D O I
10.1097/RLI.0b013e31824c86a3
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: Energetic extrapolation is a promising strategy to reduce metal artifacts in dual-source computed tomography (DSCT). We performed this study to systematically optimize image acquisition parameters for this approach in a hip phantom and assess its value in a clinical study. Materials and Methods: Titanium and steel hip prostheses were placed in a standard hip phantom and a water tank and scanned on a DSCT scanner. Tube spectra, tube current ratio, collimation, pitch, and rotation time were optimized in a stepwise process. Artifacts were quantified by measuring the standard deviation of the computed tomography density in a doughnut-shaped region of interest placed around the prosthesis. A total of 22 adult individuals with metallic implants referred for computed tomography for a musculoskeletal indication were scanned using the optimized protocol. Degree of artifacts and diagnostic image quality were rated visually (0-10) and maximum streak intensity was measured. Results: Sn140/100 kVp proved superior to Sn140/80 kVp. There was a benefit for increasing tube current ratio from 1:1 to 3:1, but not beyond, in favor of the Sn140 kVp spectrum. Artifacts were less severe for a collimation of 32 x 0.6 mm as compared with 40 x 0.6 mm. A pitch of 0.5 at a rotation time of 0.5 seconds per rotation was preferable to other combinations with comparable scanning times. In the clinical study, increasing the extrapolated photon energy from 64 to 120 keV decreased the severity of artifacts from 8.0 to 2.0 (P < 0.001) and decreased streak intensity from 871 to 153 HU (P < 0.001). The median diagnostic image quality rating improved from 2.5 to 8.0 (P < 0.001). The median energy level visually perceived as optimal for diagnostic evaluation was 113 keV (range, 100-130 keV). Conclusions: Sn140/100 kVp with a tube current ratio of 3:1, a collimation of 32 x 0.6 mm, and extrapolated energies of 105 to 120 keV are optimal parameters for a dedicated DSCT protocol that effectively reduces metal artifacts by energetic extrapolation. The protocol effectively reduces metal artifacts in all types of metal implants. The optimized reconstructions yielded relevant additional findings.
引用
收藏
页码:406 / 414
页数:9
相关论文
共 50 条
  • [21] Dual-Energy CBCT Imaging for Metal Artifact Reduction and Contrast Enhancement
    Li, H.
    Giles, W.
    Ren, L.
    Bowsher, J.
    Yin, F.
    MEDICAL PHYSICS, 2012, 39 (06) : 4017 - 4017
  • [22] Complementary contrast media for metal artifact reduction in Dual-Energy CT
    Lambert, Jack W.
    Edic, Peter M.
    FitzGerald, Paul
    Torres, Andrew S.
    Yeh, Benjamin M.
    MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING, 2015, 9412
  • [23] Metal artifact reduction by dual-layer computed tomography using virtual monoenergetic images
    Neuhaus, Victor
    Hokamp, Nils Grosse
    Abdullayev, Nuran
    Rau, Robert
    Mpotsaris, Anastasios
    Maintz, David
    Borggrefe, Jan
    EUROPEAN JOURNAL OF RADIOLOGY, 2017, 93 : 143 - 148
  • [24] Quantitative Assessment of Lipiodol-Related Artifact Reduction for Dual-Energy Computed Tomography After Transcatheter Arterial Chemoembolization: A Phantom Study Evaluating the Use of Metal Artifact Reduction Algorithms
    Ting, Yun-Lung
    Lin, Hsiang-Yi
    Huang, Tzu-Ting
    Hwang, Yi-Shuan
    Chui, Chen-Shou
    Li, Mei-Rung
    Wu, Tung-Hsin
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2023, 47 (05) : 704 - 712
  • [25] Introduction to Dual-Energy Computed Tomography
    Elster, Allen D.
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2018, 42 (06) : 823 - 823
  • [26] Comparison of dual-energy computed tomography (DECT) polychromatic and monochromatic images with and without iterative metal artifact reduction algorithm in patients with dental implants
    Huang, Shiqi
    Liang, Yuhong
    Yao, Xinqun
    Qin, Xiangyun
    He, Changguang
    Luo, Li'an
    Huang, Lveqin
    Lv, Yaping
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2024, 14 (07) : 4688 - 4702
  • [27] Heterogeneous Pulmonary Perfusion on Dual-Energy Computed Tomography: Faulty Postprocessing or New Artifact?
    Borse, Riddhi
    Homayounieh, Fatemeh
    Singh, Ramandeep
    Digumarthy, Subba R.
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2018, 42 (06) : 885 - 886
  • [28] Metal Artefact Reduction by Dual-energy Computed Tomography Using Monoenergetic Extrapolation: In-vitro Determination of Optimal Monoenergetic Level with Different Metallic Implants Using a Phantom Body
    Chan, W. C. S.
    Tsang, J. P. K.
    Wong, W. Y.
    Chu, P. Y.
    To, V. Y. K.
    Lee, C. Y.
    Yeung, T. W.
    Leung, O. C.
    Sin, N. Y.
    Li, O. C.
    Yuen, M. K.
    HONG KONG JOURNAL OF RADIOLOGY, 2016, 19 (01): : 35 - 42
  • [29] Optimal Contrast of Computed Tomography Portal Venography Using Dual-Energy Computed Tomography
    Wang, Qi
    Shi, Gaofeng
    Liu, Xiang
    Wu, Runze
    Wang, Shijie
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2013, 37 (02) : 142 - 148
  • [30] Noise reduction in dual-energy computed tomography virtual monoenergetic imaging
    Liu, Chi-Kuang
    Huang, Hsuan-Ming
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (09): : 104 - 113