Evaluation of a 2D detector array for patient-specific VMAT QA with different setups

被引:48
|
作者
Boggula, Ramesh [1 ]
Birkner, Mattias [2 ]
Lohr, Frank [1 ]
Steil, Volker [1 ]
Wenz, Frederik [1 ]
Wertz, Hansjoerg [1 ]
机构
[1] Univ Med Ctr Mannheim, Dept Radiat Oncol, Mannheim, Germany
[2] IBA Dosimetry, Schwarzenbruck, Germany
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2011年 / 56卷 / 22期
关键词
VOLUMETRIC MODULATED ARC; PIXEL IONIZATION-CHAMBER; PHOTON-BEAM ATTENUATION; IMRT QUALITY-ASSURANCE; CARBON-FIBER COUCH; TREATMENT DELIVERY; THERAPY; VERIFICATION; RADIOTHERAPY; MATRIXX;
D O I
10.1088/0031-9155/56/22/011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
For pre-treatment plan verification of advanced treatment techniques such as intensity-modulated arc therapy, a fast and reliable dosimetric device is required. In this study, we investigated the suitability of MatriXX in different setups for verification of volumetric modulated arc therapy (VMAT) plans. If MatriXX is used in a stationary phantom (MULTICube), the measured dose is dependent on the beam angle. For the first setup (MatriXX/MULTICube), we developed correction factors (CFs) for each detector element (1020 CFs). We investigated the accuracy of these CFs by verifying 12 VMAT plans. In the second setup, we also assessed the suitability of MatriXX in a dedicated holder. Using this setup (MatriXX/Holder), 30 additional VMAT plans were verified. Deviations of up to similar to 17% and similar to 11% were noted for one of the ion chambers at 90 degrees and 180 degrees gantry positions. The influence of the beam angle dependence (MULTICube) can explicitly be seen when a gamma criterion of 2%/2 mm was chosen. An overall improvement of 4.3% of passing pixels (pp) was noted after applying beam angular-dependent CFs. When the gamma criterion was 3%/3 mm, the %pp was >= 95% without and similar to 100% with correction. With the second setup, MatriXX/holder, we showed excellent agreement between measurements and calculations. The % pp averaged over all plans (30 VMAT treatment plans) was nearly similar to 100%. The combination of MatriXX with MULTICube or with holder proved to be a fast and reliable method for pretreatment verification of arc therapy with sufficient accuracy.
引用
收藏
页码:7163 / 7177
页数:15
相关论文
共 50 条
  • [41] A Study On the Response of 2D Array Detector for VMAT Delivery for a Period of Two Years
    Kumar, Syam
    George, Anu
    MEDICAL PHYSICS, 2014, 41 (06) : 375 - +
  • [42] 2D pixelated diamond detector for patient QA in advanced radiotherapy treatments
    Bartoli, A.
    Scaringella, M.
    Baldi, A.
    Greto, D.
    Scoccianti, S.
    Masi, L.
    Pallotta, S.
    Bruzzi, M.
    Talamonti, C.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S459 - S460
  • [43] Use of a plastic scintillator detector for patient-specific quality assurance of VMAT SRS
    Snyder, Jesse D.
    Sullivan, Rodney J.
    Wu, Xingen
    Covington, Elizabeth L.
    Popple, Richard A.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (09): : 143 - 148
  • [44] Impact of plan parameters and modulation indices on patient-specific QA results for standard and stereotactic VMAT
    Chun, Minsoo
    An, Hyun Joon
    Kwon, Ohyun
    Oh, Do Hoon
    Park, Jong Min
    Kim, Jung-in
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 62 : 83 - 94
  • [45] Feasibility of dose delivery error detection by a transmission detector for patient-specific QA
    Honda, H.
    Kubo, K.
    Yamamoto, R.
    Ishii, Y.
    Kanzaki, H.
    Hamamoto, Y.
    Mochizuki, T.
    Oita, M.
    Sasaki, M.
    Tominaga, M.
    Uto, Y.
    RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S787 - S788
  • [46] Correlation of phantom-based and log file patient-specific QA with complexity scores for VMAT
    Agnew, Christina E.
    Irvine, Denise M.
    McGarry, Conor K.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (06): : 204 - 216
  • [47] Correlation between tolerances for EPID metrics and CI dose deviation for patient-specific VMAT QA
    Martin, G.
    RADIOTHERAPY AND ONCOLOGY, 2023, 182 : S1470 - S1472
  • [48] AAPM Task Group Report 307: Use of EPIDs for Patient-Specific IMRT and VMAT QA
    Dogan N.
    Mijnheer B.J.
    Padgett K.
    Nalichowski A.
    Wu C.
    Nyflot M.J.
    Olch A.J.
    Papanikolaou N.
    Shi J.
    Holmes S.M.
    Moran J.
    Greer P.B.
    Medical Physics, 2023, 50 (08) : e865 - e903
  • [49] Anatomy-based, patient-specific VMAT QA using EPID or MLC log files
    Defoor, Dewayne L.
    Vazquez-Quino, Luis A.
    Mavroidis, Panayiotis
    Papanikolaou, Nikos
    Stathakis, Sotirios
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (03): : 206 - 215
  • [50] IMRT QA with a 2D diode array
    Ding, L
    Ahluwalia, W
    Liu, C
    Li, J
    Palta, J
    MEDICAL PHYSICS, 2003, 30 (06) : 1396 - 1396