Evaluation of a 2D detector array for patient-specific VMAT QA with different setups

被引:48
|
作者
Boggula, Ramesh [1 ]
Birkner, Mattias [2 ]
Lohr, Frank [1 ]
Steil, Volker [1 ]
Wenz, Frederik [1 ]
Wertz, Hansjoerg [1 ]
机构
[1] Univ Med Ctr Mannheim, Dept Radiat Oncol, Mannheim, Germany
[2] IBA Dosimetry, Schwarzenbruck, Germany
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2011年 / 56卷 / 22期
关键词
VOLUMETRIC MODULATED ARC; PIXEL IONIZATION-CHAMBER; PHOTON-BEAM ATTENUATION; IMRT QUALITY-ASSURANCE; CARBON-FIBER COUCH; TREATMENT DELIVERY; THERAPY; VERIFICATION; RADIOTHERAPY; MATRIXX;
D O I
10.1088/0031-9155/56/22/011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
For pre-treatment plan verification of advanced treatment techniques such as intensity-modulated arc therapy, a fast and reliable dosimetric device is required. In this study, we investigated the suitability of MatriXX in different setups for verification of volumetric modulated arc therapy (VMAT) plans. If MatriXX is used in a stationary phantom (MULTICube), the measured dose is dependent on the beam angle. For the first setup (MatriXX/MULTICube), we developed correction factors (CFs) for each detector element (1020 CFs). We investigated the accuracy of these CFs by verifying 12 VMAT plans. In the second setup, we also assessed the suitability of MatriXX in a dedicated holder. Using this setup (MatriXX/Holder), 30 additional VMAT plans were verified. Deviations of up to similar to 17% and similar to 11% were noted for one of the ion chambers at 90 degrees and 180 degrees gantry positions. The influence of the beam angle dependence (MULTICube) can explicitly be seen when a gamma criterion of 2%/2 mm was chosen. An overall improvement of 4.3% of passing pixels (pp) was noted after applying beam angular-dependent CFs. When the gamma criterion was 3%/3 mm, the %pp was >= 95% without and similar to 100% with correction. With the second setup, MatriXX/holder, we showed excellent agreement between measurements and calculations. The % pp averaged over all plans (30 VMAT treatment plans) was nearly similar to 100%. The combination of MatriXX with MULTICube or with holder proved to be a fast and reliable method for pretreatment verification of arc therapy with sufficient accuracy.
引用
收藏
页码:7163 / 7177
页数:15
相关论文
共 50 条
  • [21] Dosimetric characteristics of a 2D silicon diode array for stereotactic radiotherapy end-to-end patient-specific QA
    Bai, Long
    Xiao, Qing
    Wang, Qiang
    Zhao, Jianling
    Li, Guangjun
    Bai, Sen
    RADIATION PHYSICS AND CHEMISTRY, 2020, 173
  • [22] The limitation of patient-specific QA in IMRT with detector arrays
    Huh, S.
    Cho, W.
    Park, Y.
    MEDICAL PHYSICS, 2006, 33 (06) : 2143 - 2143
  • [23] 2D dose reconstruction of IMRT patient-specific QA based on log file
    Mubarok, Sayid
    Wahyu, Wibowo Edy
    Supriyanto, Pawiro Ardjo
    RADIATION PHYSICS AND CHEMISTRY, 2020, 166
  • [24] Statistical process control analysis for patient-specific IMRT and VMAT QA
    Sanghangthum, Taweap
    Suriyapee, Sivalee
    Srisatit, Somyot
    Pawlicki, Todd
    JOURNAL OF RADIATION RESEARCH, 2013, 54 (03) : 546 - 552
  • [25] Influence of dose specification on prostate VMAT patient-specific QA results
    Clemente Gutierrez, F.
    Perez Vara, C.
    Clavo Herranz, M.
    Lopez Carrizosa, C.
    Saez Garrido, J.
    Ibanez Villoslada, C.
    Couselo Paniagua, M.
    Zapatero Ortuno, J.
    Martin de Miguel, M.
    Dominguez Morcillo, M.
    Jerviz Guia, V.
    Calapaqui Teran, A.
    Guijarro Verdu, M.
    Navarro, J.
    RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S726 - S726
  • [26] A study on correlation between 2D and 3D gamma evaluation metrics in patient-specific quality assurance for VMAT
    Rajasekaran, Dhanabalan
    Jeevanandam, Prakash
    Sukumar, Prabakar
    Ranganathan, Arulpandiyan
    Johnjothi, Samdevakumar
    Nagarajan, Vivekanandan
    MEDICAL DOSIMETRY, 2014, 39 (04) : 300 - 308
  • [27] Evaluation of a 2D diode array for IMRT QA
    Letourneau, D
    MEDICAL PHYSICS, 2003, 30 (06) : 1396 - 1396
  • [28] Curtailing patient-specific IMRT QA procedures from 2D dose error distribution
    Kurosu, Keita
    Sumida, Iori
    Mizuno, Hirokazu
    Otani, Yuki
    Oda, Michio
    Isohashi, Fumiaki
    Seo, Yuji
    Suzuki, Osamu
    Ogawa, Kazuhiko
    JOURNAL OF RADIATION RESEARCH, 2016, 57 (03) : 258 - 264
  • [29] Comparison of two commercial 2D-array phantoms for patient-specific QA of stereotactic treatments
    Nevelsky, A.
    Borzov, E.
    Daniel, S.
    Bar-Deroma, R.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S929 - S930
  • [30] Complexity metric as a complement to measurement based IMRT/VMAT patient-specific QA
    Gotstedt, J.
    Hauer, A. Karlsson
    Back, A.
    8TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2015, 573