Evaluation of a 2D detector array for patient-specific VMAT QA with different setups

被引:48
|
作者
Boggula, Ramesh [1 ]
Birkner, Mattias [2 ]
Lohr, Frank [1 ]
Steil, Volker [1 ]
Wenz, Frederik [1 ]
Wertz, Hansjoerg [1 ]
机构
[1] Univ Med Ctr Mannheim, Dept Radiat Oncol, Mannheim, Germany
[2] IBA Dosimetry, Schwarzenbruck, Germany
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2011年 / 56卷 / 22期
关键词
VOLUMETRIC MODULATED ARC; PIXEL IONIZATION-CHAMBER; PHOTON-BEAM ATTENUATION; IMRT QUALITY-ASSURANCE; CARBON-FIBER COUCH; TREATMENT DELIVERY; THERAPY; VERIFICATION; RADIOTHERAPY; MATRIXX;
D O I
10.1088/0031-9155/56/22/011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
For pre-treatment plan verification of advanced treatment techniques such as intensity-modulated arc therapy, a fast and reliable dosimetric device is required. In this study, we investigated the suitability of MatriXX in different setups for verification of volumetric modulated arc therapy (VMAT) plans. If MatriXX is used in a stationary phantom (MULTICube), the measured dose is dependent on the beam angle. For the first setup (MatriXX/MULTICube), we developed correction factors (CFs) for each detector element (1020 CFs). We investigated the accuracy of these CFs by verifying 12 VMAT plans. In the second setup, we also assessed the suitability of MatriXX in a dedicated holder. Using this setup (MatriXX/Holder), 30 additional VMAT plans were verified. Deviations of up to similar to 17% and similar to 11% were noted for one of the ion chambers at 90 degrees and 180 degrees gantry positions. The influence of the beam angle dependence (MULTICube) can explicitly be seen when a gamma criterion of 2%/2 mm was chosen. An overall improvement of 4.3% of passing pixels (pp) was noted after applying beam angular-dependent CFs. When the gamma criterion was 3%/3 mm, the %pp was >= 95% without and similar to 100% with correction. With the second setup, MatriXX/holder, we showed excellent agreement between measurements and calculations. The % pp averaged over all plans (30 VMAT treatment plans) was nearly similar to 100%. The combination of MatriXX with MULTICube or with holder proved to be a fast and reliable method for pretreatment verification of arc therapy with sufficient accuracy.
引用
收藏
页码:7163 / 7177
页数:15
相关论文
共 50 条
  • [31] Reduction of Systematic Dosimetric Uncertainties in VMAT Treatments Triggered by Patient-Specific QA
    Mijnheer, B.
    Janssen, T.
    van Mourik, A.
    Kaas, J.
    Damen, E.
    Mans, A.
    MEDICAL PHYSICS, 2019, 46 (06) : E245 - E245
  • [32] Clinical Implementation of Deep Learning-Based VMAT Patient-Specific QA
    Yang, X.
    Wang, L.
    Li, D.
    Guo, Y.
    Li, Y.
    Guan, Y.
    Wu, X.
    Xu, S.
    Zhang, S.
    Chan, M.
    Yang, R.
    Geng, L.
    Sui, J.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [33] Evaluation of QA Dosimetry System for SRS and SBRT Patient-Specific QA
    Hillman, Y.
    Kim, J.
    Chetty, I.
    Wen, N.
    MEDICAL PHYSICS, 2017, 44 (06) : 3273 - 3273
  • [34] Evaluation of the ArcCHECK device for commissioning and patient-specific QA
    Shahine, B.
    Badragan, I.
    Ramaseshan, R.
    MEDICAL PHYSICS, 2012, 39 (07) : 4633 - 4633
  • [35] Evaluation of SRS MapCHECK™ for SABR patient-specific QA
    Miao, Y.
    Kidane, G.
    Ifthaker, A.
    Crees, L.
    Almond, E.
    RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S743 - S743
  • [36] Evaluation of the SRS MapCHECK for CyberKnife Patient-Specific QA
    Davis, S. D.
    Acha, A.
    George, S.
    Romaguera, W.
    Tirpak, L.
    Mishra, V.
    Gutierrez, A.
    MEDICAL PHYSICS, 2019, 46 (06) : E647 - E647
  • [37] Comparative performance analysis of 2D and 3D gamma metrics for patient specific QA in VMAT using Octavius 4D with 2D-Array 1500
    Pal, Bipasha
    Pal, Angshuman
    Bag, Santanu
    Ali, Md Abbas
    Das, Suresh
    Palit, Soura
    Sarkar, Papai
    Mallik, Suman
    Goswami, Jyotirup
    Das, Sayan
    Manir, Kazi Sazzad
    Sen, Arijit
    Mondol, Monidipa
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 91 : 18 - 27
  • [38] Mobius FX Evaluation and Comparison Against a Commercial 4D Detector Array for VMAT Plan QA
    Quino, L. Vazquez
    Hernandez, C. Huerta
    Morrow, A.
    Massingill, B.
    Rangaraj, D.
    MEDICAL PHYSICS, 2016, 43 (06) : 3533 - 3534
  • [39] Assessment of radiobiological metrics applied to patient-specific QA process of VMAT prostate treatments
    Clemente-Gutierrez, Francisco
    Perez-Vara, Consuelo
    Clavo-Herranz, Maria H.
    Lopez-Carrizosa, Concepcion
    Perez-Regadera, Jose
    Ibanez-Villoslada, Carmen
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (02): : 341 - 367
  • [40] The effect of different table top models on patient-specific QA
    Paelinck, L.
    Wuyts, E.
    Vanderstraeten, B.
    De Wagter, C.
    Lievens, Y.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S933 - S934