Evaluation of a 2D detector array for patient-specific VMAT QA with different setups

被引:48
|
作者
Boggula, Ramesh [1 ]
Birkner, Mattias [2 ]
Lohr, Frank [1 ]
Steil, Volker [1 ]
Wenz, Frederik [1 ]
Wertz, Hansjoerg [1 ]
机构
[1] Univ Med Ctr Mannheim, Dept Radiat Oncol, Mannheim, Germany
[2] IBA Dosimetry, Schwarzenbruck, Germany
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2011年 / 56卷 / 22期
关键词
VOLUMETRIC MODULATED ARC; PIXEL IONIZATION-CHAMBER; PHOTON-BEAM ATTENUATION; IMRT QUALITY-ASSURANCE; CARBON-FIBER COUCH; TREATMENT DELIVERY; THERAPY; VERIFICATION; RADIOTHERAPY; MATRIXX;
D O I
10.1088/0031-9155/56/22/011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
For pre-treatment plan verification of advanced treatment techniques such as intensity-modulated arc therapy, a fast and reliable dosimetric device is required. In this study, we investigated the suitability of MatriXX in different setups for verification of volumetric modulated arc therapy (VMAT) plans. If MatriXX is used in a stationary phantom (MULTICube), the measured dose is dependent on the beam angle. For the first setup (MatriXX/MULTICube), we developed correction factors (CFs) for each detector element (1020 CFs). We investigated the accuracy of these CFs by verifying 12 VMAT plans. In the second setup, we also assessed the suitability of MatriXX in a dedicated holder. Using this setup (MatriXX/Holder), 30 additional VMAT plans were verified. Deviations of up to similar to 17% and similar to 11% were noted for one of the ion chambers at 90 degrees and 180 degrees gantry positions. The influence of the beam angle dependence (MULTICube) can explicitly be seen when a gamma criterion of 2%/2 mm was chosen. An overall improvement of 4.3% of passing pixels (pp) was noted after applying beam angular-dependent CFs. When the gamma criterion was 3%/3 mm, the %pp was >= 95% without and similar to 100% with correction. With the second setup, MatriXX/holder, we showed excellent agreement between measurements and calculations. The % pp averaged over all plans (30 VMAT treatment plans) was nearly similar to 100%. The combination of MatriXX with MULTICube or with holder proved to be a fast and reliable method for pretreatment verification of arc therapy with sufficient accuracy.
引用
收藏
页码:7163 / 7177
页数:15
相关论文
共 50 条
  • [1] Comparison of Two Different Setups for VMAT Patient-Specific QA
    Clemente, F.
    Perez-Vara, C.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [2] Evaluation and improvement of angular response for a commercial 2D detector array for patient-specific QA
    Wang, Sha
    Zheng, Yuanshui
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (09):
  • [3] Evaluation and Improvement of Angular Response for a Commercial 2D Detector Array for Patient Specific QA
    Wang, Sha
    Zheng, Y.
    MEDICAL PHYSICS, 2022, 49 (06) : E896 - E896
  • [4] Validation of a digital method for patient-specific verification of VMAT treatment using a 2D ionisation detector array
    U'wais, Fatima Adel
    Radzi, Yasmin
    Rizan, Naslinda Noor
    Zin, Hafiz M.
    RADIATION PHYSICS AND CHEMISTRY, 2023, 202
  • [5] Development of Correction Factors for a 2D Ion Chamber Array Inserted in a Cubic and Homogeneous Phantom for VMAT Patient-Specific QA
    Clemente, F.
    Perez-Vara, C.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [6] EXPERIENCE OF PATIENT-SPECIFIC QA FOR VMAT WITH DIFFERENT INSTRUMENTS.
    Declich, F.
    Bonsignore, F.
    Butti, I.
    Frigerio, C.
    Sangalli, G.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S511 - S511
  • [7] Patient-specific QA for CyberKnife plans using 2D ion chamber array
    Adamczyk, S.
    Skrobala, A.
    Adamczyk, M.
    Pawalowski, B.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S171 - S171
  • [8] Comparison of 2D and 3D Gamma Analysis in Patient-Specific QA for Prostate VMAT Plans
    Clemente, F.
    Perez, C.
    MEDICAL PHYSICS, 2014, 41 (06) : 239 - 239
  • [9] Implementation of 2D Arrays for TomoTherapy Patient-Specific QA
    Winningham, T.
    Staton, R.
    Meeks, S.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [10] Dosimetric commissioning of a high-resolution CMOS 2D detector array for patient-specific QA of single-isocenter multi-target VMAT stereotactic radiosurgery
    Teng, Ching-Ling
    Lin, Shih-Chi
    Lovelock, Dale Michael
    Lim, Seng Boh
    JOURNAL OF RADIATION RESEARCH, 2024, 65 (06) : 787 - 797