Comparative biomechanical compression testing with a new implant for vertebral body replacement

被引:0
|
作者
Knop, C [1 ]
Lange, U [1 ]
Bastian, L [1 ]
Oeser, M [1 ]
Blauth, M [1 ]
机构
[1] Hannover Med Sch, Unfallchirurg Klin, D-30625 Hannover, Germany
来源
UNFALLCHIRURG | 2001年 / 104卷 / 01期
关键词
thoracolumbar spine; vertebral body replacement; titanium implant; compression test; biomechanics; bone mineral density;
D O I
10.1007/s001130050684
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
The authors present a new titanium implant for replacement of the vertebral body (Synex(TM)). Possible indications would be fractures or dislocations with destruction of the anterior column, posttraumatic kyphosis as well as tumors in the throracolumbar spine. The construction has to be completed by a stabilizing implant. For best fit and contact to adjacent end-plates Synex(TM) is distractable in situ. The possibility of secondary dislocation or loss of correction should thereby be minimised. Objectives. We performed comparative compression tests with Synex(TM) and MOSS(TM) ("Harms mesh cage") on human cadaveric specimens of intact vertebrae (L1). The aim of the study was to measure the compressive strength of the vertebral body end-plate in uniaxial loading via both implants to exclude a caving of Synex(TM) in vivo. Methods. 12 human cadaveric specimens of intact vertebrae (L1) were divided in 2 similar groups (matched pairs) according to bone mineral density (BMD), determined using DE-QCT. The specimens were loaded with axial compression force at a constant speed of 5 mm/min to failure and the displacement was recorded with a continuous load-displacement curve. Results. The mean ultimate compression force (F-max) showed a tendency towards a higher result testing Synex(TM) with 3396 N versus 2719 N (non significant). The displacement until F-max was 2.9 mm in group 5 (Synex(TM)), which was half as long as in group M (5.8 mm). The difference was significant (p<0.001). The compression force was twice as high and significantly (p<0.05) higher with Synex(TM) at a displacement of 1 mm, 1.5 mm and 2 mm.A significant (p<0.001) correlation (R=0.89) between F-max and BMD was found. Conclusions. Synex(TM) was found to be at least comparable to MOSS(TM) for suspensory replacement of the vertebral body at the thoracolumbar spine. A possible consequence of the significantly higher mean compression forces between 1 and 2 mm displacement might be a decreased segmental deformation or loss of correction.
引用
收藏
页码:25 / 33
页数:9
相关论文
共 50 条
  • [31] The effect of implant size and device keel on vertebral compression properties in lumbar total disc replacement
    Auerbach, Joshua D.
    Ballester, Carrie M.
    Hammond, Frank
    Carine, Ehren T.
    Balderston, Richard A.
    Elliott, Dawn M.
    SPINE JOURNAL, 2010, 10 (04): : 333 - 340
  • [32] Anterior vertebral body replacement with a titanium implant of adjustable height: a prospective clinical study
    Lange, Uta
    Edeling, Sebastian
    Knop, Christian
    Bastian, Leonard
    Oeser, Michael
    Krettek, Christian
    Blauth, Michael
    EUROPEAN SPINE JOURNAL, 2007, 16 (02) : 161 - 172
  • [33] Anterior vertebral body replacement with a titanium implant of adjustable height: a prospective clinical study
    Uta Lange
    Sebastian Edeling
    Christian Knop
    Leonard Bastian
    Michael Oeser
    Christian Krettek
    Michael Blauth
    European Spine Journal, 2007, 16 : 161 - 172
  • [34] PARTIAL REPLACEMENT OF THE VERTEBRAL BODY
    TSIVYAN, YL
    KHIRURGIYA, 1987, (10): : 48 - 53
  • [35] History of vertebral body replacement
    Tarhan, T.
    Froemel, D.
    Rickert, M.
    Rauschmann, M.
    Fleege, C.
    UNFALLCHIRURG, 2015, 118 : S73 - S79
  • [36] Biomechanical effects of the body augmenter for reconstruction of the vertebral body
    Chen, HH
    Wang, WK
    Li, KC
    Chen, TH
    SPINE, 2004, 29 (18) : E382 - E387
  • [37] Comparison of Allograft Bone and Titanium Cages for Vertebral Body Replacement in the Thoracolumbar Spine: A Biomechanical Study
    Cardenas, Raul J.
    Javalkar, Vijayakumar
    Patil, Shashikant
    Gonzalez-Cruz, Jorge
    Ogden, Alan
    Mukherjee, Debi
    Nanda, Anil
    NEUROSURGERY, 2010, 66 (06) : ONS314 - ONS318
  • [38] Osteoporotic spinal deformity - A biomechanical rationale for the clinical consequences and treatment of vertebral body compression fractures
    Yuan, HA
    Brown, CW
    Phillips, FM
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2004, 17 (03): : 236 - 242
  • [39] Augmentation with cortoss improves vertebral body compression biomechanics for lumbar total disc replacement
    Yoder, Jonathon H.
    Auerbach, Joshua D.
    Maurer, Philip M.
    Erbe, Erik M.
    Entrekin, Dean
    Balderston, Richard A.
    Bertagnoli, Rudolf
    Elliott, Dawn M.
    PROCEEDING OF THE ASME SUMMER BIOENGINEERING CONFERENCE - 2007, 2007, : 673 - 674