Audit on source strength determination for HDR and PDR 192Ir brachytherapy in Sweden

被引:21
|
作者
Tedgren, Asa Carlsson [1 ]
Grindborg, Jan-Erik [2 ]
机构
[1] Linkoping Univ, Div Med Radiat Phys, Dept Med & Hlth Sci, Fac Hlth Sci, SE-58183 Linkoping, Sweden
[2] Swedish Radiat Protect Author, Stockholm, Sweden
关键词
(192)Ir brachytherapy; dosimetry audit; RAKR;
D O I
10.1016/j.radonc.2007.12.008
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: To investigate the status of source strength determination in terms of reference air kerma rate (RAKR) for HDR and PDR (192)Ir brachytherapy in Sweden. Materials and methods: RAKR was determined in each of the 14 Swedish afterloaders, using calibrated equipment from the Swedish Secondary Standard Dosimetry Laboratory. Results: Values of RAKR from the external audit, the hospitals and vendors agreed within the uncertainty limits guaranteed by the vendors. Conclusions: The accuracy in RAKR determination has increased over the last years as a result of increased availability of interpolation standards for HDR (192)Ir and the increased use of robust well-type ion chambers designed for brachytherapy. It is recommended to establish a ratio between the RAKR value from own measurements at the hospital and that of the vendor since such a ratio embeds constant systematic differences due to e.g. varying traceability and therefore has the potential of being less uncertain than the RAKR alone. Traceability to primary standards for HDR (192)Ir sources will in the future significantly decrease the uncertainty in RAKR of (192)Ir brachytherapy. (C) 2007 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:126 / 130
页数:5
相关论文
共 50 条
  • [31] Radiochromic film dosimetry of HDR 192Ir source radiation fields
    Aldelaijan, Saad
    Mohammed, Huriyyah
    Tomic, Nada
    Liang, Li-Heng
    DeBlois, Francois
    Sarfehnia, Arman
    Abdel-Rahman, Wamied
    Seuntjens, Jan
    Devic, Slobodan
    MEDICAL PHYSICS, 2011, 38 (11) : 6074 - 6083
  • [32] Positional and angular tracking of HDR 192Ir source for brachytherapy quality assurance using radiochromic film dosimetry
    Aldelaijan, Saad
    Devic, Slobodan
    Bekerat, Hamed
    Papaconstadopoulos, Pavlos
    Schneider, James
    Seuntjens, Jan
    Cormack, Robert A.
    Buzurovic, Ivan M.
    MEDICAL PHYSICS, 2020, 47 (12) : 6122 - 6139
  • [33] Determination of air-kerma strength for the 192Ir GammaMedplus iX pulsed-dose-rate brachytherapy source
    Riley, A. D.
    Pike, T. L.
    Micka, J. A.
    Fulkerson, R. K.
    DeWerd, L. A.
    MEDICAL PHYSICS, 2013, 40 (07)
  • [34] Comparative study of simulation and experiment in determining 192Ir source strength for brachytherapy quality assurance
    Ngoc, Tran Thi Bao
    Uy, Phan Quoc
    Vuong, Le Quang
    Thanh, Tran Thien
    Tao, Chau Van
    RADIATION PHYSICS AND CHEMISTRY, 2025, 233
  • [35] Dose perturbation effects by metal hip prosthesis in gynaecological 192Ir HDR brachytherapy
    Ab Shukor, N. S.
    Abdullah, R.
    Aziz, M. Z. Abdul
    Samson, D. O.
    Musarudin, M.
    APPLIED RADIATION AND ISOTOPES, 2023, 196
  • [36] On the Significance of Model Based Dosimetry for Breast and Head and Neck 192Ir HDR Brachytherapy
    Peppa, V.
    Pappas, E.
    Pantelis, E.
    Major, T.
    Polgar, C.
    Papagiannis, P.
    MEDICAL PHYSICS, 2015, 42 (06) : 3469 - 3469
  • [37] On the accuracy of techniques for obtaining the calibration coefficient NK of 192Ir HDR brachytherapy sources
    Mainegra-Hing, Ernesto
    Rogers, D. W. O.
    MEDICAL PHYSICS, 2006, 33 (09) : 3340 - 3347
  • [38] MAASTRO applicator, a novel rectal applicator for contact brachytherapy with 192Ir HDR sources
    Bellezzo, M.
    Fonseca, G. P.
    Voncken, R.
    Verrijssen, A.
    Van Beveren, C.
    Roelofs, E.
    Yoriyaz, H.
    Reniers, B.
    Van Limbergen, E. J.
    Barbee, M.
    Verhaegen, F.
    RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S1102 - S1103
  • [39] A CT-based analytical dose calculation method for HDR 192Ir brachytherapy
    Poon, Emily
    Verhaegen, Frank
    MEDICAL PHYSICS, 2009, 36 (09) : 3982 - 3994
  • [40] A dosimetric comparison of 169Yb versus 192Ir for HDR prostate brachytherapy
    Lymperopoulou, G
    Papagiannis, P
    Sakelliou, L
    Milickovic, N
    Giannouli, S
    Baltas, D
    MEDICAL PHYSICS, 2005, 32 (12) : 3832 - 3842