Comparative study of simulation and experiment in determining 192Ir source strength for brachytherapy quality assurance

被引:0
|
作者
Ngoc, Tran Thi Bao [1 ,2 ]
Uy, Phan Quoc [1 ,2 ,3 ]
Vuong, Le Quang [1 ,2 ,4 ]
Thanh, Tran Thien [1 ,2 ]
Tao, Chau Van [1 ,2 ]
机构
[1] Univ Sci, Fac Phys & Engn Phys, Dept Nucl Phys, 227 Nguyen Van Cu St,Dist 5, Ho Chi Minh City, Vietnam
[2] Vietnam Natl Univ, Ho Chi Minh City, Vietnam
[3] Ho Chi Minh Oncol Hosp, Dept Radiat Technol, Ho Chi Minh City, Vietnam
[4] Univ Educ, Fac Phys, 280 An Duong Vuong St,Dist 5, Ho Chi Minh City, Vietnam
关键词
Brachytherapy; Air kerma strength; Monte Carlo; MCNP6; MONTE-CARLO; CO-60; BRACHYTHERAPY; DOSE DISTRIBUTION; IR-192; DOSIMETRY; CHAMBER; PHOTON;
D O I
10.1016/j.radphyschem.2025.112729
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
This research is of paramount importance as it combines experimental measurements and Monte Carlo simulation to determine the air kerma strength of the 192Ir source, which plays a crucial role in dose calculations for brachytherapy treatment planning. The air kerma strength represents the source strength value, directly influencing the accuracy of dose calculations in cancer treatments. Verifying the source strength during each source exchange is essential to the quality assurance process. In this study, the 192Ir Gammamed Plus source, provided by Varian, was used to calculate the air kerma strength through experimental measurements and Monte Carlo simulation. The measurements were conducted on a brachytherapy system with an HDR 1000 Plus well-type ionization chamber and source tube 70010 from Standard Imaging. Using the MCNP6 code, we constructed a model equivalent to the experimental system, consisting of the source geometry, source tube, and well-type chamber, with source self-absorption also considered. The air kerma strength measurements differed from the certified value by less than 2 %, within the allowable range according to the AAPM TG40 recommendations. The experimental and simulation methods showed strong agreement, with less than 4.1 % discrepancies. Considering source self-absorption, the air kerma strength calculation for the 192Ir source was consistent across both methods. The findings of this study validate that we successfully built an experimental system model using the Monte Carlo simulation with the MCNP6 code. This model can be further developed for future research.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Suitability Assessment of CsPbI3 Dosimeter with 192Ir Source for Brachytherapy Quality Assurance
    Yang, S. W.
    Park, S. K.
    Moon, C. W.
    Kim, K. H.
    JOURNAL OF INSTRUMENTATION, 2023, 18 (07)
  • [2] Audit on source strength determination for HDR and PDR 192Ir brachytherapy in Sweden
    Tedgren, Asa Carlsson
    Grindborg, Jan-Erik
    RADIOTHERAPY AND ONCOLOGY, 2008, 86 (01) : 126 - 130
  • [3] Positional and angular tracking of HDR 192Ir source for brachytherapy quality assurance using radiochromic film dosimetry
    Aldelaijan, Saad
    Devic, Slobodan
    Bekerat, Hamed
    Papaconstadopoulos, Pavlos
    Schneider, James
    Seuntjens, Jan
    Cormack, Robert A.
    Buzurovic, Ivan M.
    MEDICAL PHYSICS, 2020, 47 (12) : 6122 - 6139
  • [4] Assessment of a source position checking tool for the quality assurance of transfer tubes used in HDR 192Ir brachytherapy treatments
    Awunor, O. A.
    BRACHYTHERAPY, 2018, 17 (03) : 628 - 633
  • [5] Dose Distributions of an 192Ir Brachytherapy Source in Different Media
    Wu, C. H.
    Liao, Y. J.
    Liu, Y. W. Hsueh
    Hung, S. K.
    Lee, M. S.
    Hsu, S. M.
    BIOMED RESEARCH INTERNATIONAL, 2014, 2014
  • [6] Size and positioning reproducibility of an 192Ir brachytherapy stepping source
    Berndt, A
    Rickey, DW
    Rathee, S
    Bews, J
    MEDICAL PHYSICS, 2000, 27 (01) : 129 - 131
  • [7] Comparison of brachytherapy sources of endometrial cancer: Electronic brachytherapy source and 192Ir
    Lozares, S.
    Font Gomez, J. A.
    Gandia Martinez, A.
    Miranda Burgos, A.
    Mendez Villamon, A.
    RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S519 - S520
  • [8] The Application of Elliptic Cylindrical Phantom in Brachytherapy Dosimetric Study of HDR 192Ir Source
    Ahn, Woo Sang
    Park, Sung Ho
    Jung, Sang Hoon
    Choi, Wonsik
    Do Ahn, Seung
    Shin, Seong Soo
    SNA + MC 2013 - JOINT INTERNATIONAL CONFERENCE ON SUPERCOMPUTING IN NUCLEAR APPLICATIONS + MONTE CARLO, 2014,
  • [9] Design of a phantom for the quality control of high dose rate 192Ir source used in brachytherapy
    Ochoa, Ricardo
    Gomez, Fredy
    Ferreira, Ivaldo H.
    Gutt, Federico
    de Almeida, Carlos E.
    RADIOTHERAPY AND ONCOLOGY, 2007, 82 (02) : 222 - 228
  • [10] A technique for calibrating a high dose rate 192Ir brachytherapy source
    Fourie, O. L.
    Crabtree, T. G.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2012, 35 (01) : 85 - 92