The effect of MV cone-beam CT cupping artifacts on dose calculation accuracy

被引:2
|
作者
Chen, J [1 ]
Morin, O [1 ]
Chen, H [1 ]
Aubin, M [1 ]
Pouliot, J [1 ]
机构
[1] Univ Calif San Francisco, Ctr Comprehens Canc, San Francisco, CA 94143 USA
关键词
D O I
10.1118/1.1997618
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: MV cone‐beam (CB) CT imaging can be used to position the patient prior to external‐beam radiotherapy. If calibrated, these images may also be used in dose calculations, opening the possibility to reconstruct the delivered dose or re‐optimize the treatment plan before delivery. This study investigates the effect of imaging artifacts on dosimetric accuracy. Method and Materials: Two water cylinders of 13 cm and 22.5 cm diameters were imaged using A) a MV CBCT system integrated with a Siemens accelerator and B) a conventional kV CT imager. Cupping trends observed in the MV CBCT were imposed on the kV CT images to create artificial data with simulated artifacts. Using the Phillips Pinnacle planning system, a 6 MV 10 × 16 cm2 field was applied to 1) the original kV CT of the smaller cylinder and 2) the kV CT with simulated artifacts. Similarly, a 18 × 18 cm2 field was applied to the original and artificial kV CT of the larger cylinder. Results: For the smaller cylinder, the MV CBCT water signal relative to air at the cylinder center differs from the radial and axial extremities by approximately 14% and 17% respectively. For the larger cylinder, the differences are 38% and 20% in radial and axial directions. The dose distributions for the artificial images show a systematic deviation which increases with depth. In the high‐dose, low‐gradient regions, the largest deviation for the smaller cylinder is 1% of the maximum delivered dose. For the larger cylinder, the largest deviation is 4%. Conclusion: Cupping artifacts in water phantoms produce dosimetric errors much smaller in magnitude than the cupping trend itself but as large as 4% for large phantoms. These results suggest that rough correction of MV CBCT artifacts may be sufficient for dosimetric accuracy. Conflict of Interest: Research supported by Siemens OCS. © 2005, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:1936 / 1936
页数:1
相关论文
共 50 条
  • [31] Investigation of the accuracy of MV radiation isocentre calculations in the Elekta cone-beam CT software XVI
    Zimmermann, S. J.
    Rowshanfarzad, P.
    Ebert, M. A.
    Riis, H. L.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S842 - S843
  • [32] Dose calculation in presence of a metallic object using megavoltage cone-beam CT
    Aubin, M.
    Morin, O.
    Aubry, J.
    Chen, J.
    Descovich, M.
    Pouliot, J.
    MEDICAL PHYSICS, 2006, 33 (06) : 2222 - 2222
  • [33] MV and kV cone-beam CT on a medical linear accelerator
    Groh, BA
    Siewerdsen, JH
    Drake, DG
    Wong, JW
    Jaffray, DA
    USE OF COMPUTERS IN RADIATION THERAPY, 2000, : 561 - 563
  • [34] Evaluation of In-Room Cone-Beam CT Imaging Techniques Compared to CT for Dose Calculation
    Leeper, S. A.
    Yin, F.
    Yoo, S.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (03): : S670 - S671
  • [35] Metal artifacts correction in cone-beam CT bone imaging
    Zhang, Yan
    Ning, Ruola
    Conover, David
    MEDICAL IMAGING 2007: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2007, 6510
  • [36] Raystation KV cone-beam CT radiotherapy dose calculation: toward clinical dose tracking
    Snir, Jonatan
    Murrell, Donna
    Fakir, Hatim
    MEDICAL PHYSICS, 2017, 44 (08) : 4380 - 4380
  • [37] Reduction of artifacts induced by misaligned geometry in cone-beam CT
    Sun, Yi
    Hou, Ying
    Hu, Jiasheng
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2007, 54 (08) : 1461 - 1471
  • [38] Comparison of verification accuracy and radiation dose between megavoltage CT and kilovoltage cone-beam CT
    Wu, Vincent W. C.
    Ho, M. L.
    Yuen, L. Y.
    Li, K. L.
    Lai, N. F.
    Tsang, Y. C.
    Lee, W. Y.
    Chiu, G.
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2011, 10 (01) : 35 - 44
  • [39] Multi-pass approach to reduce cone-beam artifacts in a circular orbit cone-beam CT system
    Han, Chulhee
    Baek, Jongduk
    OPTICS EXPRESS, 2019, 27 (07) : 10108 - 10126
  • [40] Correction of megavoltage cone-beam CT images for dose calculation in the head and neck region
    Aubrya, Jean-Francois
    Pouliot, Jean
    Beaulieu, Luc
    MEDICAL PHYSICS, 2008, 35 (03) : 900 - 907