LabPET II, an APD-based Detector Module with PET and Counting CT Imaging Capabilities

被引:27
|
作者
Bergeron, Melanie [1 ,2 ]
Thibaudeau, Christian [1 ,2 ,3 ]
Cadorette, Jules [1 ,2 ]
Tetrault, Marc-Andre [3 ]
Pepin, Catherine M. [1 ,2 ]
Clerk-Lamalice, Julien [1 ,2 ]
Loignon-Houle, Francis [1 ,2 ]
Davies, Murray [4 ]
Dautet, Henri [4 ]
Deschamps, Pierre [4 ]
Fontaine, Rejean [3 ]
Lecomte, Roger [1 ,2 ]
机构
[1] Univ Sherbrooke, Sherbrooke Mol Imaging Ctr, Sherbrooke, PQ J1H 5N4, Canada
[2] Univ Sherbrooke, Dept Nucl Med & Radiobiol, Sherbrooke, PQ J1H 5N4, Canada
[3] Univ Sherbrooke, Dept Elect Engn & Comp Engn, Sherbrooke, PQ J1G 0A2, Canada
[4] Excelitas Technol, Vaudreuil, PQ J7V 8P7, Canada
基金
加拿大健康研究院; 加拿大自然科学与工程研究理事会;
关键词
Avalanche photodiodes (APD); computed tomography (CT); dual modality; positron emission tomography/computed tomography (PET/CT); positron emission tomography (PET); scintillators; ITERATIVE RECONSTRUCTION; COMPUTED-TOMOGRAPHY; ANIMAL PET; MICRO-CT; SIMULATION TOOLKIT; QUALITY; SYSTEMS; RESOLUTION; PERFORMANCE; ELECTRONICS;
D O I
10.1109/TNS.2015.2420796
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Computed tomography (CT) is currently the standard modality to provide anatomical reference for positron emission tomography (PET) in molecular imaging applications. Since both PET and CT rely on detecting radiation to generate images, using the same detection system for data acquisition is a compelling idea even though merging PET and CT hardware imposes stringent requirements on detectors. These requirements include large signal dynamic range with high signal-to-noise ratio for good energy resolution in PET and energy-resolved photon-counting CT, high pixelization for suitable spatial resolution in CT, and high count rate capability for reasonable CT acquisition time. To meet these criteria, the avalanche photodiode (APD)-based LabPET II module is proposed as the building block for a truly combined PET/CT scanner. The module is made of two monolithic 4 x 8 APD pixel arrays mounted side-by-side on a custom ceramic holder. Individual APD pixels have an active area of 1.1 x 1.1 mm(2) at a 1.2 mm pitch. The APD arrays are coupled to a 12-mm high, 8 x 8 LYSO scintillator array made of 1.12 x 1.12 mm(2) pixels also at a pitch of 1.2 mm to ensure direct one-to-one coupling to individual APD pixels. The scintillator array was designed with unbound specular reflective material between pixels to maximize the difference between refractive indices and enhance total internal reflection at the crystal side surfaces for better light collection, and the APD quantum efficiency was improved to similar to 60% at 420 nm to optimize intrinsic detector performance. Mean energy resolution was 20 +/- 1% at 511 keV and 40 +/- 4% at 60 keV. The measured intrinsic spatial and time resolution for PET were respectively 0.81 +/- 0.04 mm FWTM/1.57 +/- 0.04 and 3.6 +/- 0.3 ns FWHM with an energy threshold of 400 keV. Initial phantom images obtained using a CT test bench demonstrated excellent contrast linearity as a function of material density. With a magnification factor of 2, a CT spatial resolution of 0.66 mm FWHM/1.2 mm FWTM, corresponding to 1.18 lp/mm at MTF10% / 0.67 lp/mm at MTF50%, was measured, allowing 0.75 mm air holes in an Ultra-Micro Hot Spot resolution phantom to be clearly distinguished.
引用
收藏
页码:756 / 765
页数:10
相关论文
共 50 条
  • [21] Versatile APD-based PET Modules for High Resolution, Fast Medical Imaging
    Kataoka, J.
    Matsuda, H.
    Yoshino, M.
    Miura, T.
    Nishikido, F.
    Koizumi, M.
    Tanaka, T.
    Ikeda, H.
    Ishikawa, Y.
    Kawabata, N.
    Matsunaga, Y.
    Kishimoto, S.
    Kubo, H.
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 3542 - +
  • [22] Development of an ASIC for APD-based small animal PET
    Woodring, ML
    Christian, JF
    Shah, KS
    Squillante, MR
    Kogan, AI
    Cherry, S
    Shao, YP
    Augustine, FL
    2001 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORDS, VOLS 1-4, 2002, : 737 - 741
  • [23] Electronics design of a PET detector module with APD array
    Wang, Yong-Gang
    Hedianzixue Yu Tance Jishu/Nuclear Electronics & Detection Technology, 2002, 22 (05):
  • [24] A prototype detector module for combined PET/CT or combined photon counting/standard CT based on SiPM technology
    Persson, Andreas
    Khaplanov, Anton
    Cederwall, Bo
    Bohm, Christian
    2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, : 3503 - +
  • [25] Study of spatial resolution and depth of interaction of APD-based PET detector modules using light sharing schemes
    Bruyndonckx, P
    Léonard, S
    Liu, JG
    Tavernier, S
    Szupryczynski, P
    Fedorov, A
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2003, 50 (05) : 1415 - 1419
  • [26] A data acquisition sub-system for distributed, digital, computational, APD-based, bimodal PET/CT architecture for small animal imaging
    Fontaine, R
    Bélanger, F
    Cadorette, J
    Leroux, JD
    Martin, JP
    Michaud, JB
    Pratte, JF
    Robert, S
    Viscogliosi, N
    Lecomte, R
    2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, : 1733 - 1737
  • [27] Characteristics of Lu1.8Gd0.2SiO5:Ce (LGSO) for APD-Based PET Detector
    Shimizu, Shigenori
    Pepin, Catherine M.
    Bergeron, Melanie
    Lecomte, Roger
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2010, 57 (01) : 55 - 62
  • [28] Characteristics of Lu1.8Gd0.2SiO5:Ce (LGSO) for APD-based PET Detector
    Shimizu, Shigenori
    Pepin, Catherine M.
    Bergeron, Melanie
    Lecomte, Roger
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 4204 - 4210
  • [29] APD-based Detection Module of Single-photons in Chemiluminescence Immunoassay System
    Meng, Hao
    Guo, Yangkuan
    Wen, Chao
    MECHANICAL DESIGN AND POWER ENGINEERING, PTS 1 AND 2, 2014, 490-491 : 1636 - 1640
  • [30] Initial results of a truly pixelated APD-based PET scanner for high-resolution preclinical imaging
    Gaudin, Emilie
    Thibaudeau, Christian
    Arpin, Louis
    Leroux, Jean-Daniel
    Beaudoin, Jean-Franois
    Cadorette, Jules
    Njejimana, Larissa
    Samson, Arnaud
    Bouchard, Jonathan
    Koua, Konin
    Fontaine, jean
    Lecomte, Roger
    JOURNAL OF NUCLEAR MEDICINE, 2017, 58