Pulse transit time: validation of blood pressure measurement under positive airway pressure ventilation

被引:19
|
作者
Schmalgemeier, Heidi [1 ]
Bitter, Thomas [1 ]
Bartsch, Stephan [1 ]
Bullert, Kevin [1 ]
Fischbach, Thomas [1 ]
Eckert, Siegfried [1 ]
Horstkotte, Dieter [1 ]
Oldenburg, Olaf [1 ]
机构
[1] Ruhr Univ Bochum, Univ Hosp, Heart & Diabet Ctr N Rhine Westphalia, Dept Cardiol, D-32545 Bad Oeynhausen, Germany
关键词
Pulse transit time; Arterial blood pressure measurement; Continuous positive airway pressure (CPAP); Sleep disordered breathing; OBSTRUCTIVE SLEEP-APNEA; HEART-FAILURE; RISK; ECG;
D O I
10.1007/s11325-011-0609-7
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
The objective of the present study was to validate noninvasive, continuous blood pressure (BP) measurement using pulse transit time (PTT) to represent absolute values and detect BP changes under continuous positive airway pressure (CPAP) treatment. We applied CPAP to 78 patients of a cardiological sleep lab using 0 (baseline), 4, 8, and 12 cmH(2)O for 10 min at every level and measured BP simultaneous via PTT-based method and standard oscillometric method (OM). Quality of signal perception was acceptable to convert PTT into BP values in 64 patients (82%). When comparing both methods, we found a strong linear correlation of systolic and diastolic BP (baseline, r = 0.94 for systolic BP; r = 0.95 for diastolic BP, p < 0.001) while no significant differences between absolute values obtained with OM and PTT measurement. Mean bias at baseline was 4.1 +/- 3.2 mmHg for systolic BP and 2.3 +/- 2.2 mmHg for diastolic BP. With increasing CPAP levels, PTT and OM measurements differed continuously up to a systolic difference of 6.6 +/- 4.9 mmHg and a diastolic difference of 4.4 +/- 3.5 mmHg. There was no definite trend of PTT method to either over- or underestimate BP. We found that PTT- and OM-based BP results are closely correlated while applying CPAP over a period of 40 min. With higher CPAP level, bias between both methods increased.
引用
收藏
页码:1105 / 1112
页数:8
相关论文
共 50 条
  • [21] Smartphone-based Continuous Blood Pressure Measurement Using Pulse Transit Time
    Gholamhosseini, Hamid
    Meintjes, Andries
    Baig, Mirza
    Linden, Maria
    PHEALTH 2016, 2016, 224 : 84 - 89
  • [22] A novel calibration method for noninvasive blood pressure measurement using pulse transit time
    Yan, Y. S.
    Zhang, Y. T.
    2007 4TH IEEE/EMBS INTERNATIONAL SUMMER SCHOOL AND SYMPOSIUM ON MEDICAL DEVICES AND BIOSENSORS, 2007, : 1 - +
  • [23] Continuous blood pressure measurement using the pulse transit time: Comparison to intra-arterial measurement
    Patzak, Andreas
    Mendoza, Yuri
    Gesche, Heiko
    Konermann, Martin
    BLOOD PRESSURE, 2015, 24 (04) : 217 - 221
  • [24] Effects of continuous positive airway pressure on very short-term blood pressure variability associated with sleep-disordered breathing by pulse transit time-based blood pressure measurements
    Misaka, Tomofumi
    Yoshihisa, Akiomi
    Yokokawa, Tetsuro
    Takeishi, Yasuchika
    JOURNAL OF HYPERTENSION, 2023, 41 (05) : 733 - 740
  • [25] Evaluation of the Correlation Between Blood Pressure and Pulse Transit Time
    He, Xiaochuan
    Goubran, Rafik A.
    Liu, Xiaoping P.
    2013 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS PROCEEDINGS (MEMEA), 2013, : 17 - 20
  • [26] Blood Pressure Estimation Based on Pulse Transit Time with LANDMARC
    Wang, Yang
    Liu, Zhiwen
    2017 IEEE LIFE SCIENCES CONFERENCE (LSC), 2017, : 141 - 144
  • [27] PULSE TRANSIT-TIME - RELATIONSHIP TO BLOOD-PRESSURE
    OBRIST, PA
    LIGHT, KC
    MCCUBBIN, JA
    HUTCHESON, JS
    HOFFER, JL
    BEHAVIOR RESEARCH METHODS & INSTRUMENTATION, 1978, 10 (05): : 623 - 626
  • [28] A Compensation Method for Blood Pressure Estimation by Pulse Transit Time
    Liu, Songsong
    Ma, Heather T.
    Chen, Long
    Wang, Xiaopeng
    Yuan, Dezhang
    2013 IEEE INTERNATIONAL CONFERENCE OF IEEE REGION 10 (TENCON), 2013,
  • [29] Correlation Study Between Blood Pressure And Pulse Transit Time
    Pereira, T.
    Sanches, R.
    Reis, P.
    Pego, J.
    Simoes, R.
    2015 IEEE 4TH PORTUGUESE MEETING ON BIOENGINEERING (ENBENG), 2015,
  • [30] Blood Pressure - Pulse Transit Time Relationships: Comparative Studies
    Polinski, Artur
    Bujnowski, Adam
    Kocejko, Tomasz
    Wtorek, Jerzy
    2020 13TH INTERNATIONAL CONFERENCE ON HUMAN SYSTEM INTERACTION (HSI), 2020, : 289 - 294