Measurement of spatial coherence of electron beams by using a small selected-area aperture

被引:11
|
作者
Morishita, Shigeyuki [1 ]
Yamasaki, Jun [2 ]
Tanaka, Nobuo [2 ]
机构
[1] Nagoya Univ, Dept Crystalline Mat Sci, Nagoya, Aichi 4648603, Japan
[2] Nagoya Univ, EcoTopia Sci Inst, Nagoya, Aichi 4648603, Japan
基金
日本学术振兴会;
关键词
Spatial coherence; Transmission electron microscope; Selected-area diffraction; Electron beam; Coherence length; Airy pattern; DIFFRACTION; TRANSPORT;
D O I
10.1016/j.ultramic.2013.02.019
中图分类号
TH742 [显微镜];
学科分类号
摘要
A new method for measuring the spatial coherence of an electron beam in a transmission electron microscope is proposed. In this method, an Airy pattern produced by a circular selected-area (SA) aperture with an effective diameter of several nanometers is analyzed to obtain the degree of coherence as a function of separation in the specimen plane. Using typical TEM illumination conditions, demonstrative measurements were carried out to determine the spatial coherence length, angular size of the electron source and shape of the coherence function. Based on the results, it was shown that the ratio of the spatial coherence length to the beam radius is about 5% for a condenser aperture with a diameter of 100 mu m. This means that perfectly coherent illumination exists within the small SA aperture for beam diameters larger than 560 nm. As an example application of these results, the advantage of SA diffraction over nano-beam diffraction in electron diffractive imaging is discussed. The proposed method is unaffected by temporal coherence or geometric aberrations of the lenses. The possibility of carrying out future measurements using SA apertures with conventional sizes is also discussed. (C) 2013 Elsevier B.V. All rights reserved.
引用
收藏
页码:10 / 17
页数:8
相关论文
共 50 条
  • [21] EXTENDED ELECTRON ENERGY-LOSS FINE-STRUCTURE AND SELECTED-AREA ELECTRON-DIFFRACTION STUDIES OF SMALL PALLADIUM CLUSTERS
    DIOCIAIUTI, M
    PICOZZI, P
    SANTUCCI, S
    LOZZI, L
    DECRESCENZI, M
    JOURNAL OF MICROSCOPY-OXFORD, 1992, 166 : 231 - 245
  • [22] MEASUREMENT OF THE DEGREE OF COHERENCE OF ELECTRON-BEAMS
    OHSHITA, A
    KANOU, M
    TOMITA, H
    JOURNAL OF ELECTRON MICROSCOPY, 1980, 29 (03): : 319 - 319
  • [23] Method to measure spatial coherence of subangstrom electron beams
    Dwyer, Christian
    Erni, Rolf
    Etheridge, Joanne
    APPLIED PHYSICS LETTERS, 2008, 93 (02)
  • [24] Retrieving the complex degree of spatial coherence of electron beams
    Carrasquilla-Alvarez, J.
    Castaneda, R.
    Garcia-Sucerquia, J.
    Schofield, M. A.
    Beleggia, M.
    Zhu, Y.
    Matteucci, G.
    OPTIK, 2008, 119 (03): : 127 - 133
  • [25] A REPORT ON SPECKLE AND COHERENT DIFFRACTION PHENOMENA IN SELECTED-AREA TRANSMISSION ELECTRON-DIFFRACTION PATTERNS
    DAYKIN, AC
    ULTRAMICROSCOPY, 1994, 55 (02) : 121 - 125
  • [26] Measuring Optical Spatial Coherence by Using a Programmable Aperture
    Choi, Jai-Min
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2012, 60 (01) : 177 - 180
  • [27] Measuring optical spatial coherence by using a programmable aperture
    Jai-Min Choi
    Journal of the Korean Physical Society, 2012, 60 : 177 - 180
  • [29] Structures of nanometre-size crystals determined from selected-area electron diffraction data
    Weirich, TE
    Zou, XD
    Ramlau, R
    Simon, A
    Cascarano, GL
    Giacovazzo, C
    Hovmöller, S
    ACTA CRYSTALLOGRAPHICA SECTION A, 2000, 56 : 29 - 35
  • [30] SPATIAL COHERENCE DURING PULSED INJECTION OF ELECTRON-BEAMS
    PRITCHETT, PL
    JOURNAL OF GEOPHYSICAL RESEARCH-SPACE PHYSICS, 1990, 95 (A7) : 10671 - 10677