Beam characteristics of the first clinical 360° rotational single gantry room scanning pencil beam proton treatment system and comparisons against a multi-room system

被引:5
|
作者
Shang, Charles [1 ]
Evans, Grant [1 ]
Rahman, Mushfiqur [1 ]
Lin, Liyong [2 ]
机构
[1] South Florida Proton Therapy Inst, Delray Beach, FL 33484 USA
[2] Emory Proton Therapy Ctr, Atlanta, GA USA
来源
关键词
commissioning; pencil beam scanning; protons; treatment planning system; DOSE CALCULATION; ALGORITHM;
D O I
10.1002/acm2.12984
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose The purpose of this study was to present the proton beam characteristics of the first clinical single-room ProBeam Compact (TM) proton therapy system (SRPT) and comparison against multi-room ProBeam (TM) system (MRPT). Materials and Methods A newly designed SRPT with proton beam energies ranging from 70 to 220 MeV was commissioned in late 2019. Integrated depth doses (IDDs) were scanned using 81.6 mm diameter Bragg peak chambers and normalized by outputs at 15 mm WET and 1.1 RBE offset, following the methodology of TRS 398. The in-air beam spot profiles were acquired by a planar scintillation device, respectively, at ISO, upper and down streams, fitted with single Gaussian distribution for beam modeling in Eclipse v15.6. The field size effect was adjusted for the best overall accuracy of clinically relevant field QAs. The halo effects at near surface were quantified by a pinpoint ionization chamber. Its major dosimetric characteristics were compared against MRPT comparable beam dataset. Results Contrast to MRPT, an increased proton straggling in the Bragg peak region was found with widened beam distal falloffs and elevated proximal transmission dose values. Integrated depth doses showed 0.105-0.221 MeV (energy sigma) or 0.30-0.94 mm broader Bragg peak widths (R-b80-R-a80) for 130 MeV or higher energy beams and up to 0.48-0.79 mm extended distal falloffs (R-b20-R-b80). Minor differences were identified in beam spot sizes, spot divergences, proton particles/MU, and field size output effects. High passing scores are reported for independent end-to-end dosimetry checks by IROC and for initial 108 field-specific QAs at 3%/3 mm Gamma index with fields regardless with or without range shifters. Conclusions The author highlighted the dosimetry differences in IDDs mainly caused by the shortened beam transport system of SRPT, for which new acceptance criteria were adapted. This report offers a unique reference for future commissioning, beam modeling, planning, and analysis of QA and clinical studies.
引用
收藏
页码:266 / 271
页数:6
相关论文
共 50 条
  • [41] Development of a Daily-Treatment Beam-Monitoring System Based Gafchromic EBT3 Film for Pencil-Beam Scanning Proton Therapy
    Seonghoon Jeong
    Myonggeun Yoon
    Kwangzoo Chung
    Journal of the Korean Physical Society, 2020, 76 : 769 - 773
  • [42] Evaluation of the range shifter model for proton pencil-beam scanning for the Eclipse v.11 treatment planning system
    Matysiak, Witold
    Yeung, Daniel
    Slopsema, Roelf
    Li, Zuofeng
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (02): : 391 - 404
  • [43] Geant4 simulation of clinical proton and carbon ion beams for the treatment of ocular melanomas with the full 3-D pencil beam scanning system
    Farina, Edoardo
    Piersimoni, Pierluigi
    Riccardi, Cristina
    Rimoldi, Adele
    Tamborini, Aurora
    Ciocca, Mario
    2015 4TH INTERNATIONAL CONFERENCE ON ADVANCEMENTS IN NUCLEAR INSTRUMENTATION MEASUREMENT METHODS AND THEIR APPLICATIONS (ANIMMA), 2015,
  • [44] First Steps toward Implementing Pencil Beam Scanning Pristine Bragg Peak FLASH Planning into a Commercial Treatment Planning System
    Bookbinder, A.
    Krieger, M.
    Lansonneur, P.
    Magliari, A.
    Zhao, X.
    Choi, I. J.
    Simone, C. B., II
    Lin, H.
    Folkerts, M. M.
    Kang, M.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2024, 120 (02): : E103 - E103
  • [45] Validation of a Commercial Treatment Planning System for Small-Field Patient-Specific Apertures in Pencil Beam Scanning Proton Therapy
    Holmes, J.
    Shen, J.
    Shan, J.
    Wong, W.
    Foote, R.
    Patel, S.
    Bues, M.
    Liu, W.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [46] A Preliminary Treatment Planning Sensitivity Study for Dynamically Collimated Pencil-Beam Scanning Proton Therapy Treatments Delivered with a Dynamic Collimation System
    Smith, B.
    Hyer, D.
    Hill, P.
    Culberson, W.
    MEDICAL PHYSICS, 2018, 45 (06) : E488 - E488
  • [47] Evaluation and Commissioning of the SDX Spirometer Based System for Use in Proton Pencil Beam Scanning Treatment Under Deep Inspiration Breath-Hold
    Perles, L.
    Hobson, M.
    Gutierrez, A.
    MEDICAL PHYSICS, 2020, 47 (06) : E768 - E769
  • [48] Commissioning a clinical proton pencil beam scanning beamline for pre-clinical ultra-high dose rate irradiations on a cyclotron-based system
    Saini, Jatinder
    Erickson, Danielle P. Johnson
    Stappen, Francois Vander
    Ruth, Matt
    Cui, Sunan
    Gorman, Vanessa
    Rossomme, Severine
    Cao, Ning
    Ford, Eric C.
    Meyer, Juergen
    Bloch, Charles
    Wong, Tony
    Grassberger, Clemens
    Rengan, Ramesh
    Zeng, Jing
    Schwarz, Marco
    FRONTIERS IN ONCOLOGY, 2024, 14
  • [49] Commissioning of a fluoroscopic-based real-time markerless tumor tracking system in a superconducting rotating gantry for carbon-ion pencil beam scanning treatment
    Mori, Shinichiro
    Sakata, Yukinobu
    Hirai, Ryusuke
    Furuichi, Wataru
    Shimabukuro, Kazuki
    Kohno, Ryosuke
    Koom, Woong Sub
    Kasai, Shigeru
    Okaya, Keiko
    Iseki, Yasushi
    MEDICAL PHYSICS, 2019, 46 (04) : 1561 - 1574
  • [50] Source Model Parameters for Beam Commissioning of Scanning Proton Therapy Treatment Planning System: Multi-Institutional Survey and Analysis
    Charyyev, S.
    Chang, C. W.
    Lin, L.
    Dong, L.
    Li, H.
    Schuemann, J. P.
    Shen, J.
    Ding, X.
    MEDICAL PHYSICS, 2024, 51 (10) : 7693 - 7693