Coronary computed tomography angiography using ultra-low-dose contrast media: radiation dose and image quality

被引:34
|
作者
Komatsu, Sei [1 ]
Kamata, Teruaki [2 ]
Imai, Atsuko [1 ,3 ]
Ohara, Tomoki [1 ]
Takewa, Mitsuhiko [1 ]
Ohe, Ryoko [1 ]
Miyaji, Kazuaki [2 ]
Yoshida, Junichi [1 ]
Kodama, Kazuhisa [1 ]
机构
[1] Amagasaki Cent Hosp, Ctr Cardiovasc, Amagasaki, Hyogo 6610976, Japan
[2] Amagasaki Cent Hosp, Dept Diagnost Imaging, Amagasaki, Hyogo 6610976, Japan
[3] Osaka Univ, Sch Med, Dept Cardiol, Osaka, Japan
来源
关键词
Coronary computed tomography angiography; Contrast media; Tube potential; CT; PHANTOM; SINGLE;
D O I
10.1007/s10554-013-0201-2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To analyze the invasiveness and image quality of coronary CT angiography (CCTA) with 80 kV. We enrolled 181 patients with low body weight and low calcium level. Of these, 154 patients were randomly assigned to 1 of 3 groups: 280 HU/80 kV (n = 51); 350 HU/80 kV (n = 51); or 350 HU/120 kV (n = 52). The amount of contrast media (CM) was decided with a CT number-controlling system. Twenty-seven patients were excluded because of an invalid time density curve by timing bolus. The predicted amount of CM, volume CT dose index, dose-length product, effective dose, image noise, and 5-point image quality were measured. The amounts of CM for the 80 kV/280 HU, 80 kV/350 HU, and 120 kV/350 HU groups were 10 +/- A 4 mL, 15 +/- A 7 mL, and 30 +/- A 6 mL, respectively. Although image noise was greater at 80 than 120 kV, there was no significant difference in image quality between 80 kV/350 HU and 120 kV/350 HU (p = 0.390). There was no significant difference in image quality between 80 kV/280 HU and 80 kV/350 HU (4.4 +/- A 0.7 vs. 4.7 +/- A 0.4, p = 0.056). The amount of CM and effective dose was lower for 80 kV CCTA than for 120 kV CCTA. CCTA at 80 kV/280 HU may decrease the amount of CM and radiation dose necessary while maintaining image quality.
引用
收藏
页码:1335 / 1340
页数:6
相关论文
共 50 条
  • [1] Coronary computed tomography angiography using ultra-low-dose contrast media: radiation dose and image quality
    Sei Komatsu
    Teruaki Kamata
    Atsuko Imai
    Tomoki Ohara
    Mitsuhiko Takewa
    Ryoko Ohe
    Kazuaki Miyaji
    Junichi Yoshida
    Kazuhisa Kodama
    The International Journal of Cardiovascular Imaging, 2013, 29 : 1335 - 1340
  • [2] CORONARY CT ANGIOGRAPHY USING ULTRA-LOW DOSE CONTRAST MEDIA: RADIATION DOSE AND IMAGE QUALITY
    Komatsu, Sei
    Imai, Atsuko
    Kodama, Kazuhisa
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2012, 59 (13) : E1357 - E1357
  • [3] ULTRA-LOW-DOSE 70-KV CORONARY COMPUTED TOMOGRAPHY ANGIOGRAPHY USING DUALSOURCE COMPUTED TOMOGRAPHY
    Komatsu, Sei
    Takahashi, Satoru
    Takewa, Mitsuhiko
    Ohara, Tomoki
    Yutani, Chikao
    Kodama, Kazuhisa
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2019, 73 (09) : 1468 - 1468
  • [4] Adaptive Statistical Iterative Reconstruction V: Impact on Image Quality in Ultra-low-dose Coronary Computed Tomography Angiography
    Benz, Dominik C.
    Grani, Christoph
    Hirt-Moch, Beatrice
    Mikulicic, Fran
    Vontobel, Jan
    Fuchs, Tobias A.
    Gaemperli, Oliver
    Kaufmann, Philipp A.
    Buechel, Ronny R.
    CIRCULATION, 2015, 132
  • [5] Image Quality in Computed Tomography Coronary Angiography and Radiation Dose Reduction
    Golubickas, Domas
    Lukosevicius, Saulius
    Tamakauskas, Vytenis
    Dobrovolskiene, Laima
    Baseviciene, Ingrida
    Grib, Liviu
    Ragaisyte, Nijole
    Leonavicius, Rytis
    Medvedev, Viktor
    Veikutis, Vincentas
    INFORMATICA, 2021, 32 (04) : 741 - 757
  • [6] ULTRA-LOW-DOSE COMPUTED TOMOGRAPHY IN UROLITHIASIS-EFFECT OF AN ADDITIONAL TIN FILTER ON IMAGE QUALITY AND RADIATION DOSE
    Haubold, Johannes
    Zensen, Sebastian
    Erfanian, Youssef
    Guberina, Nika
    Opitz, Marcel
    Sawicki, Lino Morris
    Forsting, Michael
    Umutlu, Lale
    Theysohn, Jens M.
    RADIATION PROTECTION DOSIMETRY, 2021, 197 (3-4) : 146 - 153
  • [7] Ultra-low radiation 70kv-coronary computed tomography angiography with low-dose contrast media using dual source computed tomography
    Komatsu, S.
    Takahashi, S.
    Yutani, C.
    Takewa, M.
    Ohara, T.
    Kodama, K.
    EUROPEAN HEART JOURNAL, 2019, 40 : 3781 - 3781
  • [8] Evaluation of image quality and radiation dose in low tube voltage coronary computed tomography angiography
    Mousavi-Gazafroudi, Seyedeh Shokoofeh
    Sajjadieh-Khajouei, Amirreza
    Moradi, Maryam
    Mousavi-Gazafroudi, Seyedeh Shabnam
    Yadegarfar, Ghasem
    Tavakoli, Mohammad Bagher
    ARYA ATHEROSCLEROSIS, 2019, 15 (05) : 205 - 210
  • [9] Application of Aortic Computed Tomography Angiography with Low Radiation Dose: Assessment of Image Quality and Radiation Dose
    Liu, Jianhua
    Li, Long
    Zhang, Yan
    Yuan, Qinghai
    Sun, Yong
    2012 IEEE FIFTH INTERNATIONAL CONFERENCE ON ADVANCED COMPUTATIONAL INTELLIGENCE (ICACI), 2012, : 148 - 151
  • [10] Image quality and radiation exposure with low-contrast-dose computed tomography angiography of the lower extremities
    Assi, Abed Al Nasser
    POLISH JOURNAL OF RADIOLOGY, 2020, 85 : E169 - E173