Coronary computed tomography angiography using ultra-low-dose contrast media: radiation dose and image quality

被引:34
|
作者
Komatsu, Sei [1 ]
Kamata, Teruaki [2 ]
Imai, Atsuko [1 ,3 ]
Ohara, Tomoki [1 ]
Takewa, Mitsuhiko [1 ]
Ohe, Ryoko [1 ]
Miyaji, Kazuaki [2 ]
Yoshida, Junichi [1 ]
Kodama, Kazuhisa [1 ]
机构
[1] Amagasaki Cent Hosp, Ctr Cardiovasc, Amagasaki, Hyogo 6610976, Japan
[2] Amagasaki Cent Hosp, Dept Diagnost Imaging, Amagasaki, Hyogo 6610976, Japan
[3] Osaka Univ, Sch Med, Dept Cardiol, Osaka, Japan
来源
关键词
Coronary computed tomography angiography; Contrast media; Tube potential; CT; PHANTOM; SINGLE;
D O I
10.1007/s10554-013-0201-2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To analyze the invasiveness and image quality of coronary CT angiography (CCTA) with 80 kV. We enrolled 181 patients with low body weight and low calcium level. Of these, 154 patients were randomly assigned to 1 of 3 groups: 280 HU/80 kV (n = 51); 350 HU/80 kV (n = 51); or 350 HU/120 kV (n = 52). The amount of contrast media (CM) was decided with a CT number-controlling system. Twenty-seven patients were excluded because of an invalid time density curve by timing bolus. The predicted amount of CM, volume CT dose index, dose-length product, effective dose, image noise, and 5-point image quality were measured. The amounts of CM for the 80 kV/280 HU, 80 kV/350 HU, and 120 kV/350 HU groups were 10 +/- A 4 mL, 15 +/- A 7 mL, and 30 +/- A 6 mL, respectively. Although image noise was greater at 80 than 120 kV, there was no significant difference in image quality between 80 kV/350 HU and 120 kV/350 HU (p = 0.390). There was no significant difference in image quality between 80 kV/280 HU and 80 kV/350 HU (4.4 +/- A 0.7 vs. 4.7 +/- A 0.4, p = 0.056). The amount of CM and effective dose was lower for 80 kV CCTA than for 120 kV CCTA. CCTA at 80 kV/280 HU may decrease the amount of CM and radiation dose necessary while maintaining image quality.
引用
收藏
页码:1335 / 1340
页数:6
相关论文
共 50 条
  • [41] Patient-tailored Contrast Delivery Protocols for Computed Tomography Coronary Angiography Lower Contrast Dose and Better Image Quality
    van den Boogert, Thomas P. W.
    Lopes, Ricardo R.
    Lobe, Nick H. J.
    Verwest, Tim A.
    Stoker, Jaap
    Henriques, Jose P.
    Marquering, Henk A.
    Planken, R. Nils
    JOURNAL OF THORACIC IMAGING, 2021, 36 (06) : 353 - 359
  • [42] AI Denoising Improves Image Quality and Radiological Workflows in Pediatric Ultra-Low-Dose Thorax Computed Tomography Scans
    Brendlin, Andreas S.
    Schmid, Ulrich
    Plajer, David
    Chaika, Maryanna
    Mader, Markus
    Wrazidlo, Robin
    Mannlin, Simon
    Spogis, Jakob
    Estler, Arne
    Esser, Michael
    Schaefer, Juergen
    Afat, Saif
    Tsiflikas, Ilias
    TOMOGRAPHY, 2022, 8 (04) : 1678 - 1689
  • [43] Feasibility of an ultra-low dose contrast media protocol for coronary CT angiography
    Luan, X.
    Gao, Z.
    Sun, J.
    Chen, G.
    Song, H.
    Yao, J.
    Yan, S.
    Yu, H.
    Song, P.
    CLINICAL RADIOLOGY, 2022, 77 (09) : E705 - E710
  • [44] Effect of dose reduction on image quality and diagnostic performance in coronary computed tomography angiography
    Noortje van der Bijl
    Raoul M. S. Joemai
    Bart J. A. Mertens
    Albert de Roos
    Wouter J. H. Veldkamp
    Jeroen J. Bax
    Joanne D. Schuijf
    Jacob Geleijns
    Lucia J. M. Kroft
    The International Journal of Cardiovascular Imaging, 2013, 29 : 453 - 461
  • [45] Effect of dose reduction on image quality and diagnostic performance in coronary computed tomography angiography
    van der Bijl, Noortje
    Joemai, Raoul M. S.
    Mertens, Bart J. A.
    de Roos, Albert
    Veldkamp, Wouter J. H.
    Bax, Jeroen J.
    Schuijf, Joanne D.
    Geleijns, Jacob
    Kroft, Lucia J. M.
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2013, 29 (02): : 453 - 461
  • [46] A low-dose and an ultra-low-dose contrast agent protocol for coronary CT angiography in a clinical setting: quantitative and qualitative comparison to a standard dose protocol
    Benz, Dominik C.
    Grani, Christoph
    Hirt, Beatrice
    Moch, Beatrice Hirt
    Mikulicic, Fran
    Vontobel, Jan
    Fuchs, Tobias A.
    Stehli, Julia
    Clerc, Olivier F.
    Possner, Mathias
    Pazhenkottil, Aju P.
    Gaemperli, Oliver
    Buechel, Ronny R.
    Kaufmann, Philipp A.
    BRITISH JOURNAL OF RADIOLOGY, 2017, 90 (1074):
  • [47] Radiation dose and image quality with new protocol in lower extremity computed tomography angiography
    Schicchi, Nicolo
    Fogante, Marco
    Oliva, Matteo
    Pirani, Paolo Esposto
    Agliata, Giacomo
    Giuseppetti, Gian Marco
    Giovagnoni, Andrea
    RADIOLOGIA MEDICA, 2019, 124 (03): : 184 - 190
  • [48] Radiation dose and image quality with new protocol in lower extremity computed tomography angiography
    Nicolò Schicchi
    Marco Fogante
    Matteo Oliva
    Paolo Esposto Pirani
    Giacomo Agliata
    Gian Marco Giuseppetti
    Andrea Giovagnoni
    La radiologia medica, 2019, 124 : 184 - 190
  • [49] Diagnostic Performance of Low-Radiation-Dose Coronary Computed Tomography Angiography RESPONSE
    Alkadhi, Hatem
    von Ballmoos, Moritz Wyler
    ANNALS OF INTERNAL MEDICINE, 2011, 155 (04) : 278 - +
  • [50] Diagnostic accuracy of ultra-low-dose chest computed tomography in an emergency department
    Taekker, Maria
    Kristjansdottir, Bjorg
    Andersen, Michael B.
    Fransen, Maja L.
    Greisen, Pernille W.
    Laursen, Christian B.
    Mussmann, Bo
    Posth, Stefan
    Graumann, Ole
    ACTA RADIOLOGICA, 2022, 63 (03) : 336 - 344