Auxiliary device for calibration of roller pumps: development and experimental results

被引:1
|
作者
Vieira, F. U., Jr. [1 ,2 ]
Dantas de Medeiros, J., Jr. [3 ]
Antunes, N. [2 ]
Vieira, R. W. [2 ]
Costa, E. T. [1 ,3 ]
机构
[1] State Univ Campinas UNICAMP, Ctr Biomed Engn, Sao Paulo, Brazil
[2] State Univ Campinas UNICAMP, Sch Med Sci, Sao Paulo, Brazil
[3] State Univ Campinas UNICAMP, Sch Elect & Comp Engn, Dept Biomed Engn, Sao Paulo, Brazil
来源
PERFUSION-UK | 2012年 / 27卷 / 06期
关键词
roller pump; hemolysis; calibration device; CARDIOPULMONARY BYPASS; CENTRIFUGAL PUMPS; RETROGRADE FLOW; DYNAMIC METHOD; HEMOLYSIS; SURGERY; BLOOD;
D O I
10.1177/0267659112453077
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: Roller pumps are commonly used in surgeries involving extracorporeal circulation and its regulation may influence the hemolysis rate. Objective: To describe an auxiliary device for adjusting roller pumps using the dynamic calibration method and show preliminary results of its application in surgical procedures. Method:The device was tested in the laboratory environment with three disposable pressure transducers normally used in surgeries. In the operation room, tests were carried out with two groups of patients. Free hemoglobin plasma rates were measured,5 minutes after perfusion started and immediately before the passage of the residual volume of the pump. For Group 1 (n = 22), occlusion measurements were carried out using the drop rate method and for Group 2 (n = 18), measurements were carried out using the auxiliary device and 300 mmHg calibration (mean pressure for the dynamic calibration method). Results:The auxiliary calibration device has shown calibration results statistically equal to those obtained with a reference device in the laboratory environment. We have found less variability of hemolysis rates in the operating room, using the auxiliary device and we did not observe any influence of unbalanced rollers in the hemolysis rates.
引用
收藏
页码:512 / 519
页数:8
相关论文
共 50 条
  • [1] Preliminary Tests Using an Auxiliary Device to Adjust Roller Pumps Used in Extracorporeal Circulation
    Costa, E. T.
    Vieira Junior, F. U.
    de Medeiros Junior, J. D.
    Antunes, N.
    Vieira, R. W.
    5TH LATIN AMERICAN CONGRESS ON BIOMEDICAL ENGINEERING (CLAIB 2011): SUSTAINABLE TECHNOLOGIES FOR THE HEALTH OF ALL, PTS 1 AND 2, 2013, 33 (1-2): : 754 - 757
  • [2] Development of a simulation device for impact roller
    Chen Zhong-qing
    Xu Chao
    Ye Guan-bao
    Qiang Hai-fei
    Lu Sheng
    ROCK AND SOIL MECHANICS, 2015, 36 (01) : 279 - 285
  • [3] Methodology and Tool Development for Mobile Device Cameras Calibration and Evaluation of the Results
    Patonis, Photis
    SENSORS, 2023, 23 (03)
  • [4] Development of a Cantilever calibration device
    Dannberg, Oliver
    Kuehnel, Michael
    Froehlich, Thomas
    TM-TECHNISCHES MESSEN, 2020, 87 (10) : 622 - 629
  • [5] Development of a Cantilever calibration device
    Dannberg, Oliver
    Kühnel, Michael
    Fröhlich, Thomas
    Technisches Messen, 2020, 87 (10): : 622 - 629
  • [6] Considerations on roller pumps setting methods used in cardiopulmonary bypass: Drop rate and dynamic calibration
    Vieira Jr., Francisco Ubaldo
    Vieira, Reinaldo Wilson
    Antunes, Nilson
    de Oliveira, Pedro Paulo Martins
    Petrucci Jr., Orlando
    do Carmo, Marcio Roberto
    de Oliveira Severino, Elaine Soraya Barbosa
    Revista Brasileira de Engenharia Biomedica, 2010, 26 (01): : 25 - 32
  • [7] The Influence of the Residual Stress in Silicone Tubes in the Calibration Methods of Roller Pumps Used in Cardiopulmonary Bypass
    Vieira, Francisco Ubaldo, Jr.
    Vieira, Reinaldo Wilson
    Antunes, Nilson
    Petrucci, Orlando, Jr.
    de Oliveira, Pedro Paulo
    Silveira Filho, Lindemberg da Mota
    de Sousa Vilarinho, Karlos Alexandre
    Barbosa de Oliveira Severino, Elaine Soraya
    ASAIO JOURNAL, 2010, 56 (01) : 12 - 16
  • [8] Design and Experiment of Spiral Blades Auxiliary Roller of Organic Fertilizer Side Throwing Device
    Liu H.
    Zhao Y.
    Xie Y.
    Zhang Y.
    Shang J.
    Nongye Jixie Xuebao/Transactions of the Chinese Society for Agricultural Machinery, 2023, 54 (04): : 107 - 119
  • [9] Development of the pulsation device for rotary blood pumps
    Yambe, T
    Shiraishi, Y
    Sekine, K
    Shibata, M
    Yamaguchi, T
    Jian, LH
    Yoshizawa, M
    Tanaka, A
    Matsuki, H
    Sato, F
    Haga, Y
    Esashi, M
    Tabayashi, K
    Mitamura, Y
    Sasada, H
    Nitta, S
    ARTIFICIAL ORGANS, 2005, 29 (11) : 912 - 915
  • [10] Experimental results on kinematic calibration of parallel manipulators using a partial pose measurement device
    Rauf, A
    Pervez, A
    Ryu, J
    IEEE TRANSACTIONS ON ROBOTICS, 2006, 22 (02) : 379 - 384